9
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Repeatability and Reproducibility of a Double-Pass Optical Quality Analysis Device

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose

          To evaluate the repeatability and reproducibility of a double-pass instrument (OQASII, Visiomereics SL, Spain), which objectively measures overall optical quality of the human eyes.

          Methods

          The right eye of 119 healthy subjects with best corrected visual acuity of 20/25 or better was included in this prospective, comparative, observational study. Two separate tests with OQASII were conducted sequentially on the same day by two different examiners. A week later, the first examiner conducted the third measurement. All subjects underwent three consecutive tests during each session. The repeatability and reproducibility of the modulation transfer function cut off frequency (MTF cutoff), the Strehl ratio, the OQAS values (OVs) at contrasts of 100%, 20% and 9%, and the objective scatter index (OSI) were analyzed.

          Results

          For MTF cutoff, Strehl ratio, OV100%, OV20%, OV9%, and OSI, the mean values were 39.32±9.75cpd, 0.22±0.06, 1.31±0.33, 1.33±0.39, 1.33±0.41, 0.60±0.42, respectively. Repeatability and reproducibility were good with a very low coefficient of variation and high interclass correlation coefficients (>0.88) for all parameters. Bland-Altman plots showed good correlation with 95% limits of agreement ranged from -6.04 to 6.78cpd, -0.05 to 0.05, -0.20 to 0.23, -0.29 to 0.32, -0.40 to 0.42, -0.23 to 0.21 in inter-observer, and -6.56 to 7.42cpd, -0.06 to 0.06, -0.22 to 0.24, -0.30 to 0.32, -0.35 to 0.34, -0.24 to 0.23 in inter-visit, respectively.

          Conclusion

          The OQASII system yields excellent repeatability and good reproducibility for objective measurements of overall optical quality in clinic.

          Related collections

          Most cited references12

          • Record: found
          • Abstract: found
          • Article: found
          Is Open Access

          An Objective Scatter Index Based on Double-Pass Retinal Images of a Point Source to Classify Cataracts

          Purpose To propose a new objective scatter index (OSI) based in the analysis of double-pass images of a point source to rank and classify cataract patients. This classification scheme is compared with a current subjective system. Methods We selected a population including a group of normal young eyes as control and patients diagnosed with cataract (grades NO2, NO3 and NO4) according to the Lens Opacities Classification System (LOCS III). For each eye, we recorded double-pass retinal images of a point source. In each patient, we determined an objective scatter index (OSI) as the ratio of the intensity at an eccentric location in the image and the central part. This index provides information on the relevant forward scatter affecting vision. Since the double-pass retinal images are affected by both ocular aberrations and intraocular scattering, an analysis was performed to show the ranges of contributions of aberrations to the OSI. Results We used the OSI values to classify each eye according to the degree of scatter. The young normal eyes of the control group had OSI values below 1, while the OSI for subjects in LOCS grade II were around 1 to 2. The use of the objective index showed some of the weakness of subjective classification schemes. In particular, several subjects initially classified independently as grade NO2 or NO3 had similar OSI values, and in some cases even higher than subjects classified as grade NO4. A new classification scheme based in OSI is proposed. Conclusions We introduced an objective index based in the analysis of double-pass retinal images to classify cataract patients. The method is robust and fully based in objective measurements; i.e., not depending on subjective decisions. This procedure could be used in combination with standard current methods to improve cataract patient surgery scheduling.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Comparison of the retinal image quality with a Hartmann-Shack wavefront sensor and a double-pass instrument.

            Wavefront sensors provide quite useful information on the optical quality of the eye. However, in eyes where very high-order aberrations and scattered light are prominent, wavefront sensors may overestimate retinal image quality. This study showed that, in those cases, the double-pass technique is a complementary tool for better estimation of ocular optical quality. A double-pass (DP) instrument was used, based on recording images of a point source in near-infrared light after reflection in the retina and double-pass through the ocular media. The aberrations were also measured with a prototype of near-infrared Hartmann-Shack (HS) wavefront sensor adapted to the clinical environment. From the wave aberrations, the modulation transfer function (MTF) was calculated (MTF_HS). The MTF was also obtained from the double-pass images (MTF_DP). Both techniques were applied in normal young subjects as the control and in three other groups of eyes: older subjects, after LASIK refractive surgery, and after IOL implantation. The MTFs obtained from DP and HS techniques were compared. In the group of normal eyes with low levels of intraocular scattering, these estimates were quite similar, indicating that both techniques captured well most of the optical degradation. However, in eyes where scatter was more predominant (e.g., early cataract, posterior capsular opacification after IOL implantation) the MTF provided by the HS sensor was always higher than the MTF obtained from DP. A single parameter was used to indicate the differences. In eyes with low scattering, DP and HS techniques provided similar estimates of the retinal image quality. However, in a patient's eye with mild to severe amount of scatter, wavefront sensors might overestimate image quality, whereas the DP technique produces a more accurate description of the optical quality, better correlated with the quality of vision.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Optical and visual impact of tear break-up in human eyes.

              The purpose of this study was to examine the optical and visual impact of tear break-up. Optical quality of the eye was assessed during periods of nonblinking by quantifying vessel contrast in the fundus image and by monitoring the psychophysical contrast sensitivity and the spatial distribution of tear thickness changes by retroillumination. All measures were obtained from three eyes either with or without a soft contact lens. A noticeable decrease in retinal vessel contrast and contrast sensitivity were observed soon after a blink. Both of these measures of optical quality of the eye showed a similar pattern of image degradation both with and without a soft contact lens. Although trial-to-trial variability was considerable, sample means show that image contrast in the low spatial frequency range can drop to between 20% and 40% of initial values after 60 seconds of nonblinking. Retroillumination of the tear film showed local intensity fluctuations that progressively spread across the pupil with increasing time after the blink. Optical aberrations created by tear break-up contribute to the decline in image quality observed objectively and psychophysically. The decline in image quality that accompanies tear break-up may be a direct cause of the blurry vision complaints commonly encountered in dry-eye patients.
                Bookmark

                Author and article information

                Contributors
                Role: Academic Editor
                Journal
                PLoS One
                PLoS ONE
                plos
                plosone
                PLoS ONE
                Public Library of Science (San Francisco, CA USA )
                1932-6203
                3 February 2015
                2015
                : 10
                : 2
                : e0117587
                Affiliations
                [001]The Eye Hospital of Wenzhou Medical University, Wenzhou, China
                Sun Yat-sen University, CHINA
                Author notes

                Competing Interests: The authors have declared that no competing interests exist.

                Conceived and designed the experiments: AY QW. Performed the experiments: CX TX YZ JH. Analyzed the data: AY. Wrote the paper: AY TX CX.

                Article
                PONE-D-14-30830
                10.1371/journal.pone.0117587
                4315520
                25646846
                75628e60-c029-4012-b61f-87ea46903723
                Copyright @ 2015

                This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited

                History
                : 13 July 2014
                : 28 December 2014
                Page count
                Figures: 2, Tables: 3, Pages: 8
                Funding
                The authors have no support or funding to report.
                Categories
                Research Article
                Custom metadata
                All relevant data are within the paper.

                Uncategorized
                Uncategorized

                Comments

                Comment on this article