1
Introduction
Approximately 524 cardiac pacemakers per million people are implanted in Europe per
year with an increasing year-on-year trend [1]. It is estimated that up to 75% of
pacemaker recipients will need magnetic resonance imaging (MRI) in their lifetime
[2]. The burden of cardiovascular disease in pacemaker recipients, coupled with the
increasingly prominent role of cardiac MRI in European guidelines for the diagnosis,
management and monitoring of patients with cardiovascular disease has meant providing
cardiac MRI to this population has become a necessity [3,4]. The advent of MRI conditional
pacemakers has facilitated safe scanning of these patients although individual manufacturers'
restrictions remain in place.
The feasibility and safety of performing cardiac MRI in pacemaker patients for acquisition
of cines, late gadolinium imaging and perfusion has previously been established [[5],
[6], [7]]. Furthermore there is increasing evidence that cardiac MRI in patients with
implantable cardiac devices can often aid diagnosis or change clinical management
[5,8]. Four-dimensional flow (4D flow) cardiac MRI is one of the emerging MRI techniques
which has demonstrated high accuracy and precision for intracardiac flow and haemodynamic
assessment [9,10]. Due to its advantages over two-dimensional phase contrast acquisition
s and other Doppler based imaging methods, it is being increasingly advocated for
challenging cases of congenital heart disease, valvular heart disease and haemodynamic
assessment [[11], [12], [13]]. Retrospective valve tracking methods have been shown
to be accurate and reliable for the assessment of valvular flow and regurgitation
quantification [9,14]. However the feasibility, safety and reliability of this technique
remains to be confirmed in patients with pacemakers.
We hypothesised that 4D flow cardiac MRI is feasible in patients with pacemakers and
can accurately quantify valvular flow. Therefore, the main aims of the study were
to (1) assess the feasibility of performing 4D flow in patients with MRI conditional
pacemakers and (2) investigate the consistency and reliability of retrospective valve
tracking in quantification of valvular flow in patients with pacemakers in both atrial
(AOO) and dual chamber (DOO) asynchronous pacing modes.
2
Materials and methods
2.1
Study population
The study was approved by the local Ethics Committee and the study complied with the
Declaration of Helsinki. All patients gave written informed consent before MRI examinations.
Thirteen patients with MRI conditional dual chamber pacemakers were prospectively
recruited from a single centre. Inclusion criteria: Adults (aged over 18), MRI conditional
dual chamber pacemaker system, ventricular pacing burden of less than 5% on most recent
device interrogation. Exclusion criteria: Contraindication to MRI (including non-MRI
conditional pacemakers, intra-orbital debris, severe claustrophobia), pregnant or
breastfeeding, history of prior myocardial infarction, moderate to severe valvular
heart disease and known structural heart disease.
2.2
Device programming
Prior to entering the MRI room, the patients underwent full pacemaker interrogation
which included determination of battery voltage, lead impedance, pacing thresholds
and P- and R-wave sensing amplitude. Devices were then programmed into manufacturer
specific MRI safe mode. Patients were programmed to either AOO or DOO asynchronous
pacing, in an arbitrary fashion, at 10 beats per minute above intrinsic heart rate
to avoid competition. 12 lead electrocardiograms (ECG) were performed prior to MRI
to ensure atrial pacing with intrinsic atrioventricular conduction (AOO mode) and
sequential atrial and ventricular pacing (DOO mode). All patients were scanned in
both AOO and DOO pacing modes during a single visit in order to evaluate feasibility
of 4D flow derived valvular flow quantification in different pacing modes and the
effect of the pacing mode on artefacts. Throughout the MRI examination patients were
monitored using vectrocardiogram (VCG) signal and non-invasive blood pressure measurements.
Following MRI a safety check was performed assessing the device battery voltage, lead
impedance, pacing thresholds and sensing amplitudes and compared to values obtained
prior to the MRI. Patients were then reprogrammed to pre MRI device settings.
2.3
Cardiovascular magnetic resonance
All patients had cardiac MRI imaging at 1.5 Tesla (Ingenia, Philips, Best, The Netherlands)
with a phased array receiver coil (24-channel equipped with Philips dStream digital
broadband MR architecture technology) between November 2017 and October 2018. The
mean time between device implantation and MRI examination was 281 days (range: 88–853 days).
All patients were scanned in normal operating mode (Upper limit of SAR level up to
2 W/kg body weight) with maximised gradient slew rate up to 200 T/m/s and according
to the manufacturer's specific device instructions.
2.4
Image acquisition
The MRI protocol was as follows:
1.
Survey images
2.
Cine imaging: Acquired using balanced steady state free precession (bSSFP) in a single
slice breath-hold sequence. Images obtained included a LV volume contiguous short
axis stack as well as two, three and four chamber views. Typical image parameters
were as follows: Slice thickness 10 mm, echo time (TE) 1.5 milliseconds (ms), repetition
time (TR) 3 ms, flip angle 60°, SENSE factor 2 with 30 phases per cardiac cycle.
3.
Whole heart 4D flow: Field of view (FoV) was planned in the transaxial plane with
changes to FoV and number of slices performed as necessary to ensure whole heart coverage.
Acquisition was performed using a fast field echo (FFE) pulse sequence [EPI based
with sensitivity encoding (SENSE) acceleration, 3D] as previously described with retrospective
ECG triggering [15]. Acquisition voxel size approximately 3x3x3mm. Typical scan parameters
were as follows: TE 3.5 ms, TR 13 ms, flip angle 10°, velocity encoding (VENC) 150 cm/s,
FoV 400 mm, number of signal averages 1, EPI acceleration factor of 5 and SENSE factor
of 2. Images were acquired during free breathing with no respiratory motion correction.
Number of slices was 39 with temporal resolution of 40 ms. Number of reconstructed
phases was set at 30.
4.
Patients were taken out of the MRI room and the device was re-programmed to alternate
pacing mode at the same base rate and steps 1 to 3 were repeated.
2.5
Image analysis
Image analysis was performed offline using MASS software (Version 2018EXP, Leiden
University Medical Centre, Leiden, The Netherlands). All images were analysed by CS
(2 years' experience in advanced cardiac MRI). Endocardial contours were traced on
the LV short-axis (SA) cine stack at end-diastole and end-systole, with exclusion
of papillary muscles and trabeculation, to determine end-diastolic volume, end-systolic
volume, stroke volume and ejection fraction for both left and right ventricles (summation
of disks methodology). Epicardial contours were contoured for the left ventricle at
end-diastole to calculate left ventricular mass.
For each 4D flow data set, visual quality checks on the phase contrast and magnitude
images were performed by CS (2 years experience in advanced cardiac MRI) and doubled
checked by PG (>5 years experience in 4D flow cardiac MRI). 4D flow phase contrast
and magnitude images were visually assessed across each heart valve for the presence
of the following artefacts: signal void, distortions (particularly due to the presence
of pacemaker lead) and phase dispersion. The images were graded according to a 4-point
scale similar to previously published work [15]. 0: excellent quality with no artefacts,
1; good quality with minimal blurring artefacts, 2; moderate quality with moderate
blurring or distortion artefacts, 3; poor quality with severe artefacts in the area
of interest leading to potentially non-evaluable data. Phase unwrapping was performed
on source images if aliasing occurred in the region of interest according to previous
guidelines on phase contrast methods [16]. Spatial misalignment of 4D flow to cine
imaging was corrected prior to flow analysis. This was achieved by visualising velocity
vectors in 4-chamber view in peak systole and repositioning them over the descending
aorta and in 3-chamber view in peak systole and repositioning them over the ascending
aorta. Similar checks were performed in diastole for peak mitral inflow velocity vectors
in 2-, 3- and 4-chamber views.
All 4D flow assessments were performed using validated retrospective valve tracking
techniques with the measurement planes positioned perpendicular to inflow or outflow
direction on two-, three- and four-chamber cines [14]. Background velocity correction
(for correction of through plane motion and phase offset) was used from velocity sampled
in the myocardium as per guidelines on phase contrast methods [16]. Contour segmentation
was performed manually. Flow was determined over the entire cardiac cycle and stroke
volume was calculated by the absolute forward flow minus any regurgitant flow. Susceptibility-related
miscalculations of flow in certain image pixels, due to the presence of pacemaker
leads within the right heart chambers, were expected to be present across the tricuspid
valve on reformatted images (Supplemental fig. 1). Therefore tricuspid valve planes
were manually contoured twice; initially to include the entire tricuspid orifice area
and then subsequently with exclusion of miscalculated pixels that occurred because
of the pacing lead (Fig. 1). To assess inter-observer variability of 4D flow derived
stroke volumes a second observer (AC) repeated the analysis for all three evaluated
heart valves, with exclusion of miscalculated pixels in the tricuspid valve plane,
and was blinded to the previous analysis.
Fig. 1
Example of segmentation of valvular flow contours on the phase contrast multiplanar
reconstruction.
For the tricuspid valvular flow, we just excluded the area with artefact from through
plane valvular flow quantification (orange arrow). The right hand panel demonstrates
flow curves for the same patient in AOO mode with comparable stroke volumes through
the 3 valvular planes.
Fig. 1
2.6
Statistical analysis
Statistical analysis was performed using SPSS 21 (International Business Machines,
Armonk, New York, USA). Normality for quantitative data was established using the
Shapiro-Wilk test. Continuous data measurements are presented as mean ± standard deviation.
For image quality analysis the Wilcoxon signed rank test was performed to establish
significant differences. For investigating agreement between left and right ventricular
stroke volumes from cine imaging and aortic, mitral and tricuspid stroke volumes derived
from 4D flow we used repeated measures analysis of variance (ANOVA) with Bonferroni
adjustment for post-hoc analysis. Bland-Altman plots were used to both visually assess
the agreement between the methods and investigate the bias (in percentage). Association
between aortic and mitral and tricuspid stroke volumes was performed using Pearson
correlation coefficient test. For inter-observer analysis the coefficient of variation
(CV) was calculated using the root mean square method and reliability was assessed
using intraclass correlation coefficient (ICC). For pre and post MRI device parameters
a paired samples t-test was performed for normally distributed variables and the Wilcoxon
signed rank test for not normally distributed variables. A p value <0.05 was considered
significant.
3
Results
3.1
Patient characteristics
All thirteen patients, mean age 66 ± 11 years, seven males, completed the full study
protocol. Five patients were assigned to an initial AOO pacing rhythm and the remainder
to DOO first. A summary of the baseline demographic characteristics of the study participants
and cine volumetric parameters in AOO pacing mode is provided in Table 1. The pacemaker
and lead details for patients can be seen in Table 2.
Table 1
Baseline characteristics of patients recruited to study. Cine volume results obtained
during AOO pacing mode.
Table 1
Parameter
All patients (n = 13)
Female gender
6 (46%)
Age (yr)
66 ± 11
Heart rate (bpm)
81 ± 10
Height (cm)
170 ± 12
Weight (kg)
84 ± 20
Cine volumetric results
LVEDV (ml)
120 ± 30
LVESV (ml)
49 ± 9
LVSV (ml)
71 ± 18
LVEF (%)
59 ± 4
LV Mass (gram)
74 ± 20
RVEDV (ml)
114 ± 28
RVESV (ml)
45 ± 12
RVSV (ml)
69 ± 17
RVEF (%)
60 ± 4
Abbreviations: LVEDV: Left ventricular end diastolic volume, LVESV: Left ventricular
end systolic volume, LVSV: Left ventricular stroke volume, LVEF: Left ventricular
ejection fraction, LV: Left ventricle, RVEDV: Right ventricular end diastolic volume,
RVESV: Right ventricular end systolic volume, RVSV: Right ventricular stroke volume,
RVEF: Right ventricular ejection fraction.
Table 2
Pacemaker and lead models in the study population.
Table 2
Manufacturer
Model
Number
Implantable Pulse Generator
Boston Scientific
Proponent MRI (EL231)
5
Medtronic
Ensura DR MRI (EN1DR01)
2
St Jude Medical
Assurity MRI (PM2272)Endurity MRI (PM2172)
24
Lead
Boston Scientific
Ingevity MRI (7731, 7732, 7735, 7736, 7741, 7742)
10
Medtronic
Capsure Fix (5076)
4
St Jude Medical
Tendril STS (2088TC)Tendril MRI (LPA1200M)Isoflex (1944)
822
3.2
Safety and device parameters
All examinations were completed safely with no adverse clinical events and no unusual
symptoms reported during the scan. All devices were interrogated before and immediately
after MRI (Table 3). No significant differences were noted between battery voltage,
lead impedance, capture threshold or P- and R-wave amplitude. No individual changes
in lead parameters were considered clinically significant.
Table 3
Comparison of device parameters before and immediately after the MRI examination.
Table 3
Parameter
Pre MRI value
Post MRI value
p-value
Pacing lead impedance (Ω)
−
Atrial lead
−
Ventricular lead
527.5 ± 94.1665.6 ± 146.6
514.5 ± 66.9634.8 ± 154.2
0.640.11
Pacing lead capture threshold (V)
−
Atrial lead
−
Ventricular lead
0.6 ± 0.20.9 ± 0.4
0.6 ± 0.20.8 ± 0.2
0.760.92
Battery Voltage (V) (n = 9)*
3.02 ± 0.1
3.02 ± 0.1
NA
P-wave amplitude (mV)
4.0 ± 1.4
4.1 ± 1.4
0.48
R-wave amplitude (mV)
12.3 ± 5.6
12.1 ± 5.3
0.95
*Boston Scientific devices were excluded as the programmer does not given a numerical
value for battery voltage.
3.3
Image quality assessments
Minor banding artefacts secondary to the implantable pulse generator (IPG), predominantly
in apical slices, in SA cine images were observed in 5 patients in both pacing modes.
Artefact scoring for phase and magnitude images across the aortic and mitral valves
was similar with generally no or minimal artefacts observed in both AOO and DOO pacing
modes (Fig. 2). Overall there was no significant difference in the presence of artefacts
on images between pacing modes (nil; p = 1.0, minimal; p = 0.63, moderate; p = 0.06
or severe; p = 0.18). However, due to the presence of the pacing leads, moderate or
severe artefacts, due to susceptibility-related miscalculations of flow, were seen
on phase images across the tricuspid valve in all patients (Supplemental fig. 1).
Fig. 2
Qualitative assessment of flow in the raw data prior to valvular plane reconstruction.
Even though poor quality for tricuspid flow was more often noted, by removing the
miscalculated pixels, we were able to quantify tricuspid stroke volume.
Fig. 2
3.4
Tricuspid flow quantification - with/without inclusion of pacemaker lead artefact
On direct comparison of tricuspid flow with the inclusion of the RV lead artefact
versus exclusion of the lead artefact, we noted that when we included the RV lead
artefact there was significant overestimation of transvalvular stroke volume (SV)
in both AOO (77 ± 18mls vs 69 ± 18mls; p < 0.001) and DOO modes (74 ± 17mls vs 68 ± 17mls;
p < 0.001). Therefore the values that excluded the RV lead artefact were used for
subsequent comparison with stroke volumes of left sided heart valves (Fig. 1). No
significant tricuspid regurgitation was observed, after exclusion of RV pacing lead
artefact, with negligible negative flow seen in either AOO or DOO pacing modes (1.43 ± 1.36
mls vs. 1.91 ± 0.93 mls respectively; p = 0.26).
3.5
Consistency of 4D flow derived flow volume assessment
In AOO pacing mode SV for the aortic valve correlated with both mitral (r = 0.95;
p < 0.001) and tricuspid (r = 0.96; p < 0.001) valvular SVs (Fig. 3). Bias for SV
in AOO pacing mode was highest between the aortic and tricuspid valves (−3.5%, LOA
−17 to 10%; p = 0.09) although was not significant (Fig. 4). In DOO pacing mode, SV
for the aortic valve correlated with both mitral (r = 0.95; p < 0.001) and tricuspid
(r = 0.97; p < 0.001) valvular SVs (Fig. 3). No significant bias for the SV in this
pacing mode was observed between aortic valve and mitral and tricuspid valves (−4.8%,
LOA −26 to 16%; p = 0.13 and − 5.6%, LOA −32 to 20%; p = 0.15 respectively) (Fig.
4). No significant aortic or mitral regurgitation was seen on reformatted images.
There was negligible and non-significant negative flow between AOO and DOO pacing
modes across both the aortic (AOO: 0.39 ± 0.87 mls vs. DOO: 0.79 ± 1.24 mls; p = 0.23)
and mitral (AOO: 0.64 ± 0.44 mls vs. DOO: 0.76 ± 0.33 mls; p = 0.44) valves.
Fig. 3
Scatter plots of aortic stroke volume (SV) against mitral and tricuspid SV for AOO
and DOO pacing modes to investigate consistency between methods.
Excellent correlation was noted for all (r>0.95).
Fig. 3
Fig. 4
Bland-Altman analysis for the assessment of aortic stroke volume (SV) against mitral
and tricuspid SV for AOO and DOO pacing modes.
No significant differences was noted on Bland-Altman analysis.
Fig. 4
3.6
Comparison of cine and 4D flow derived valvular stroke volumes
In both AOO and DOO pacing modes there was no significant difference between the mean
SV obtained from short-axis cine imaging for either the left or right ventricle and
4D flow derived aortic, mitral or tricuspid SV (p > 0.05) (Table 4). Bland-Altman
analysis did not demonstrate any significant bias between the left ventricular cine
SV and 4D flow methods for each valvular SV in either pacing mode (p > 0.05) (Fig.
5).
Fig. 5
Comparison of LV short-axis cine and 4D flow derived stroke volumes (SV).
Bland-Altman plots were used to investigate any significant bias between cine SV and
4D flow derived SV. The Bland-Altman analysis did not demonstrate any significant
bias between cine SV and the 4D flow methods derived SV (P>0.05).
Fig. 5
Table 4
Comparison of mean stroke volume by cine and aortic/mitral/tricuspid valves derived
from 4D flow according to assigned pacing mode. No significant differences were noted.
Table 4
Stroke Volume (ml)
p-value⁎
LV cine
RV cine
Aortic
Mitral
Tricuspid
Atrial pacing mode (AOO)
71 ± 18
69 ± 17
67 ± 15
69 ± 20
69 ± 18
0.15
Ventricular pacing mode (DOO)
67 ± 19
68 ± 19
66 ± 22
68 ± 19
68 ± 17
0.70
Abbreviations: LV: Left ventricle, RV: Right ventricle.
⁎
P-value – from repeated measures ANOVA using Bonferroni post-hoc analysis.
3.7
Inter-observer repeatability
For the aortic and mitral valves in both pacing modes the ICC were strong with a low
CV suggesting good inter-observer agreement (Table 5). For the tricuspid valve, in
both pacing modes, ICC was lower with a higher CV suggesting a more modest inter-observer
agreement.
Table 5
Inter-observer reproducibility for 4D flow derived valvular stroke volumes for both
pacing modes.
Table 5
CV (%)
ICC
Atrial pacing mode (AOO)
Aortic SV
8.7
0.912
Mitral SV
8.1
0.965
Tricuspid SV
14.0
0.762
Ventricular pacing mode (DOO)
Aortic SV
7.6
0.921
Mitral SV
8.8
0.911
Tricuspid SV
10.2
0.861
Abbreviations: CV: coefficient of variation, ICC: intraclass correlation coefficient,
SV: stroke volume.
4
Discussion
The present study investigated the feasibility and consistency of 4D flow derived
valvular flow assessment in patients with MRI conditional pacemakers. The study demonstrates
that: (1) 4D flow cardiac MRI is feasible in patients with MRI conditional pacemakers
in two different pacing modes; (2) Flow across left sided (aortic and mitral) heart
valves is consistent in both AOO and DOO pacing modes; (3) 4D flow derived valvular
stroke volume quantification is comparable with the cine derived stroke volume; (4)
Susceptibility artefacts are commonly present on the tricuspid valve plane due to
the RV pacing lead but can be circumvented to some extent by excluding miscalculated
pixels in close proximity to the lead.
4.1
Safety
All the patients in the study underwent the full protocol with no significant changes
in device parameters noted between the pre and post MRI device interrogation. Therefore
the current study suggests that 4D flow cardiac MRI seems not to pose any additional
risk in patients with MRI conditional pacemakers if scanned in normal operating mode
(SAR level up to 2 W/kg body weight) with a maximised gradient slew rate up to 200 T/m/s.
These findings are in keeping with the previous literature demonstrating the safety
of performing cardiac MRI on patients with MRI conditional pacemakers [6,8].
4.2
Image quality and qualitative assessment of flow
Imaging artefacts in patients with pacemakers occur predominantly due to the presence
of ferromagnetic material within the IPG and pacing leads. However alterations of
patient positioning within the scanner when changing pacing modes and movement of
the pacing lead throughout the cardiac cycle could potentially further effect image
quality.
On cine imaging artefact was predominantly due to the presence of IPG leading to minor
banding artefacts, predominantly in the apical LV segments [6]. These artefacts were
consistent between pacing modes and endocardial/blood pool definition was adequate
to allow determination of stroke volume for both ventricles. The lower ferromagnetic
content in pacing leads meant little to no artefact was seen on cine imaging. No visual
change in artefact was observed between pacing modes.
The presence of an MRI conditional pacemaker has previously been shown not to affect
the image quality or generation of flow curves in 2D aortic phase contrast imaging
[6]. The current study demonstrated the image quality of the phase contrast and magnitude
images for 4D flow acquisition in patients with pacemakers was generally good, particularly
for the left heart. The reconstructed aortic and mitral valve planes generally had
little or no artefact which allowed robust quantification of valvular flow. Furthermore,
no significant artefacts were noted in velocity vector visualisation on cine images
for either the left or right heart (Fig. 6).
Fig. 6
A case example demonstrating two dimensional velocity vectors superimposed over cine
images in a patient with a pacemaker and right ventricular pacing lead (orange arrow).
No significant artefacts were noted.
Fig. 6
Miscalculated pixels secondary to the susceptibility generated by the RV pacing lead
were consistently seen on the phase and magnitude 4D flow data of the tricuspid valve
plane. These were generally limited to a few pixels in close proximity to the RV lead.
Contouring the entire orifice area, including the miscalculated pixels, led to overestimation
of stroke volume relative to the left sided heart valves. In all of our cases the
RV pacing lead was at the edge of the valve orifice area and therefore repeat manual
contouring with exclusion of the miscalculated pixels created by the pacing lead meant
stroke volumes comparable to the aortic and mitral valves could be determined. This
technique clearly requires additional post processing time and the effect of flow
measurements when the pacing lead is positioned in the middle of the valve orifice
is unknown. However, in the latter circumstance we would suggest a second contour
be drawn around the artefact and this value deducted from the total stroke volume
for the entire orifice area. There was no significant difference in artefacts between
the pacing modes.
4.3
Quantitative assessment of transvalvular stroke volume
Current methods of quantifying valvular flow and intra-cardiac shunts are based on
Doppler echocardiography techniques that are often limited by acoustic windows, difficulties
with velocity assessment due to beam alignment and are therefore often dependent on
operator experience meaning measurements often have limited reproducibility [[17],
[18], [19]]. Over recent years 4D flow derived measurements using valvular stroke
volumes obtained by the retrospective valve tracking techniques have been shown to
be accurate, consistent and reproducible across all four heart valves [9,14,20,21].
The present study has shown that stroke volume quantification, particularly for the
left sided heart valves, by retrospective valve tracking is consistent in patients
with pacemakers and is reproducible in two separate pacing modes. These findings are
consistent with a previous study by Garg et al. which used similar undersampling methods
for faster 4D flow whole-heart acquisitions - EPI acceleration with a factor of five
and a SENSE factor of 2 [15]. This study showed robust correlation between aortic
and mitral net forward flow (r = 0.94) in healthy volunteers which is comparable to
our results in both pacing modes (r ≥ 0.95). Importantly inter-observer repeatability
was less robust for the tricuspid valve, which may be a consequence of the artefact
generated by the RV pacing lead, and further work is needed to determine the effect
of the pacing lead on these 4D flow derived stroke volumes. 4D flow derived valvular
stroke volumes were also consistent with stroke volumes determined by cine imaging.
4.4
Clinical applications
The demonstration of feasibility as well as the consistency of 4D flow derived flow
measurements is important as the number of pacemaker implantations in Europe is on
an upward trend due to the ageing population [1]. Given the burden of cardiovascular
disease in pacemaker recipients it seems probable that a significant proportion of
them will require cardiac MRI during their lifespan given cardiac MRI is often recommended
in International guidelines [3,4]. Cardiac MRI has already been shown to provide important
diagnostic and management changing information in patients with pacemakers [5,8].
4D flow MRI can play a vital additive role as it provides accurate and consistent
intra-scan assessment of blood flow with strong rescan reproducibility. Indeed 4D
flow allows sampling and quantification of blood flow in any direction within the
3D volume so may forgo the need for a series of 2D cine breath held phase contrast
sequences and retrospective valve tracking techniques may improve assessment of transvalvular
flow [9,22]. This may be particularly pertinent in the repeated imaging of pacemaker
patients with congenital or valvular heart disease where serial assessment of regurgitant
volumes or shunts is required [10,[23], [24], [25]].
4.5
Possible future applications
Right ventricular apical pacing induces electrical and mechanical dyssynchrony leading
to alterations in cardiac haemodynamics and can lead to adverse cardiac remodelling
and even the development of heart failure in the long-term [26,27]. The mechanisms
underpinning the development of this so called ‘pacing induced cardiomyopathy’ however
are incompletely understood. 4D flow cardiac MRI affords the evaluation of a series
of advanced cardiac haemodynamic parameters such as kinetic energy (KE), turbulent
KE, particle tracing and vortex visualisation [22]. These parameters are predominantly
research tools but have been suggested as subclinical markers of LV dysfunction with
reductions in average LV KE and end diastolic KE observed in patients with ischaemic
heart disease and little or no LV dysfunction [28,29]. More recently it has been shown
in heart failure patients with dyssynchrony from left bundle branch block (LBBB) that
LV filling forces are more orthogonal to main LV flow direction during early diastole
and the direct flow entering the LV has lower KE when compared to those without LBBB
[30,31]. Suwa et al. have also demonstrated changes in vortex size and core locations
during diastole in patients with heart failure suggesting vortex formation plays a
role in LV ejection and filling [32]. Therefore these metrics may allow us to evaluate
how flow haemodynamics change in pacing induced dyssynchrony and may contribute to
the pathophysiology of pacing induced left ventricular dysfunction and development
of heart failure. Indeed recent work using echocardiographic particle image velocimetry
has demonstrated that blood flow momentum and KE dissipation are altered with RV apical
pacing and associated with deterioration in global longitudinal strain, highlighting
the potential role that altered flow dynamics may play in adverse cardiac remodelling
over the longer term in these patients [33]. Although full evaluation of intra-cardiac
flows in patients with pacemakers would require manufacturers to allow greater flexibility
in device programming within the MRI environment.
4.6
Limitations
There were several limitations to our study. The number of patients recruited to this
study remains small and the implanted pacemakers were from a limited number of manufacturers
with MRI conditional models. This study did not evaluate pulmonary valvular flow as
the relevant right ventricular outflow tract cines for retrospective valve tracking
planning were not acquired. The artefact created by the RV pacing lead meant tricuspid
stroke volume was overestimated. Although excluding these miscalculated pixels meant
that stroke volumes were consistent with aortic and mitral valves this could have
important implications for calculating regurgitant volumes across the tricuspid valve,
particularly if this occurs in close proximity to the pacing lead. The 4D flow sequence
used in this study was not respiratory navigated. However respiratory navigated sequences
have a longer acquisition time and this may preclude their application in clinical
workflows. Furthermore in healthy volunteers the use of respiratory motion compensation
has been shown to have no significant effect on intra-cardiac flow quantification
[34]. This study did not recruit patients with significant valvular heart disease,
especially patients with tricuspid regurgitation. Future studies will need to establish
the reliability of 4D flow in quantifying pulmonary and tricuspid flow in pacemaker
patients particularly as inclusion of miscalculated pixels due to the RV pacing lead
may augment derived stroke volumes. This is not as relevant for the left heart as
the artefacts are minimal. Larger studies are required to fully evaluate safety of
4D flow cardiac MRI across a wider range of devices including cardiac resynchronisation
pacemakers and implanted cardioverter-defibrillators.
5
Conclusion
Whole-heart, 4D flow cardiac MRI in patients with MRI conditional pacemakers is feasible.
Retrospective valve tracking techniques allow assessment of stroke volumes, particularly
across left sided heart valves, irrespective of pacing mode and are comparable to
stroke volumes obtained using cine imaging. Further research is needed in patients
with defibrillators and cardiac resynchronisation devices to evaluate whether better
device optimisation is possible by 4D flow guided cardiac haemodynamics.
Grant support
AD is funded by
10.13039/501100000327
Heart Research UK, United Kingdom
(RG2668/18/20). ED is supported by a grant from the
10.13039/501100000274
British Heart Foundation, United Kingdom
(FS/13/71/30378). SP is supported by a grant from the
10.13039/501100000274
British Heart Foundation, United Kingdom
(CH/16/2/32089).
Disclosures
The authors have reported that they have no relationships relevant to the contents
of this paper to disclose.