      Computerized Casts for Orthodontic Purpose Using Powder-Free Intraoral Scanners: Accuracy, Execution Time, and Patient Feedback

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Introduction

          Intraoral scanners allow direct images of oral situation, with fewer steps than conventional impressions. The purpose of this study was to compare the accuracy of digital impressions, traditional impressions, and digitalization of full-arch gypsum models, to evaluate timing of different methods and finally to study perception of patients about conventional and digital impression techniques.

          Methods

          Dental arches of fourteen patients were evaluated by alginate impression, titanium dioxide powder-free intraoral scanning (Trios, 3Shape), and digitalization obtained from gypsum models using the same scanner. Conventional and digital techniques were evaluated through measurements (lower and upper arch anteroposterior length, lower and upper intercanine distance, and lower and upper intermolar distance) with a caliber for analogic models and using a computer software for digital models (Ortho Analyzer, Great Lakes Orthodontics). In addition, chairside and processing times were recorded. Finally, each patient completed a VAS questionnaire to evaluate comfort. Statistical analyses were performed with ANOVA and Tukey tests for accuracy measurements and paired t-test for times and VAS scores. Significance was predetermined at P < 0.05.

          Results

          The measurements obtained with intraoral scanning, gypsum models after conventional impression, and digitalized gypsum models were not significantly different. Both chairside and processing times of digital scanning were shorter than the traditional method. VAS reporting patients comfort were significantly higher when evaluating digital impression.

          Conclusions

          Intraoral scanners used for orthodontic applications provide useful data in clinical practice, comparable to conventional impression. This technology is more time efficient than traditional impression and comfortable for patients. Further evolution with more accurate and faster scanners could in future replace traditional impression methods.

          Related collections

          Most cited references31

          • Record: found
          • Abstract: found
          • Article: found
          Is Open Access

          Editorial

            Bookmark
            • Record: found
            • Abstract: found
            • Article: found
            Is Open Access

            The complete digital workflow in fixed prosthodontics: a systematic review

            Background The continuous development in dental processing ensures new opportunities in the field of fixed prosthodontics in a complete virtual environment without any physical model situations. The aim was to compare fully digitalized workflows to conventional and/or mixed analog-digital workflows for the treatment with tooth-borne or implant-supported fixed reconstructions. Methods A PICO strategy was executed using an electronic (MEDLINE, EMBASE, Google Scholar) plus manual search up to 2016–09-16 focusing on RCTs investigating complete digital workflows in fixed prosthodontics with regard to economics or esthetics or patient-centered outcomes with or without follow-up or survival/success rate analysis as well as complication assessment of at least 1 year under function. The search strategy was assembled from MeSH-Terms and unspecific free-text words: {((“Dental Prosthesis” [MeSH]) OR (“Crowns” [MeSH]) OR (“Dental Prosthesis, Implant-Supported” [MeSH])) OR ((crown) OR (fixed dental prosthesis) OR (fixed reconstruction) OR (dental bridge) OR (implant crown) OR (implant prosthesis) OR (implant restoration) OR (implant reconstruction))} AND {(“Computer-Aided Design” [MeSH]) OR ((digital workflow) OR (digital technology) OR (computerized dentistry) OR (intraoral scan) OR (digital impression) OR (scanbody) OR (virtual design) OR (digital design) OR (cad/cam) OR (rapid prototyping) OR (monolithic) OR (full-contour))} AND {(“Dental Technology” [MeSH) OR ((conventional workflow) OR (lost-wax-technique) OR (porcelain-fused-to-metal) OR (PFM) OR (implant impression) OR (hand-layering) OR (veneering) OR (framework))} AND {((“Study, Feasibility” [MeSH]) OR (“Survival” [MeSH]) OR (“Success” [MeSH]) OR (“Economics” [MeSH]) OR (“Costs, Cost Analysis” [MeSH]) OR (“Esthetics, Dental” [MeSH]) OR (“Patient Satisfaction” [MeSH])) OR ((feasibility) OR (efficiency) OR (patient-centered outcome))}. Assessment of risk of bias in selected studies was done at a ‘trial level’ including random sequence generation, allocation concealment, blinding, completeness of outcome data, selective reporting, and other bias using the Cochrane Collaboration tool. A judgment of risk of bias was assigned if one or more key domains had a high or unclear risk of bias. An official registration of the systematic review was not performed. Results The systematic search identified 67 titles, 32 abstracts thereof were screened, and subsequently, three full-texts included for data extraction. Analysed RCTs were heterogeneous without follow-up. One study demonstrated that fully digitally produced dental crowns revealed the feasibility of the process itself; however, the marginal precision was lower for lithium disilicate (LS2) restorations (113.8 μm) compared to conventional metal-ceramic (92.4 μm) and zirconium dioxide (ZrO2) crowns (68.5 μm) (p < 0.05). Another study showed that leucite-reinforced glass ceramic crowns were esthetically favoured by the patients (8/2 crowns) and clinicians (7/3 crowns) (p < 0.05). The third study investigated implant crowns. The complete digital workflow was more than twofold faster (75.3 min) in comparison to the mixed analog-digital workflow (156.6 min) (p < 0.05). No RCTs could be found investigating multi-unit fixed dental prostheses (FDP). Conclusions The number of RCTs testing complete digital workflows in fixed prosthodontics is low. Scientifically proven recommendations for clinical routine cannot be given at this time. Research with high-quality trials seems to be slower than the industrial progress of available digital applications. Future research with well-designed RCTs including follow-up observation is compellingly necessary in the field of complete digital processing. Electronic supplementary material The online version of this article (10.1186/s12903-017-0415-0) contains supplementary material, which is available to authorized users.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Orthodontic measurements on digital study models compared with plaster models: a systematic review.

              The aim of this study is to evaluate the validity of the use of digital models to assess tooth size, arch length, irregularity index, arch width and crowding versus measurements generated on hand-held plaster models with digital callipers in patients with and without malocclusion. Studies comparing linear and angular measurements obtained on digital and standard plaster models were identified by searching multiple databases including MEDLINE, LILACS, BBO, ClinicalTrials.gov, the National Research Register and Pro-Quest Dissertation Abstracts and Thesis database, without restrictions relating to publication status or language of publication. Two authors were involved in study selection, quality assessment and the extraction of data. Items from the Quality Assessment of Studies of Diagnostic Accuracy included in Systematic Reviews checklist were used to assess the methodological quality of included studies. No meta-analysis was conducted. Comparisons between measurements of digital and plaster models made directly within studies were reported, and the difference between the (repeated) measurement means for digital and plaster models were considered as estimates. Seventeen relevant studies were included. Where reported, overall, the absolute mean differences between direct and indirect measurements on plaster and digital models were minor and clinically insignificant. Orthodontic measurements with digital models were comparable to those derived from plaster models. The use of digital models as an alternative to conventional measurement on plaster models may be recommended, although the evidence identified in this review is of variable quality. © 2010 John Wiley & Sons A/S.
                Bookmark

                Author and article information

                Contributors
                Journal
                Biomed Res Int
                Biomed Res Int
                BMRI
                BioMed Research International
                Hindawi
                2314-6133
                2314-6141
                2018
                23 April 2018
                : 2018
                : 4103232
                Affiliations
                Unit of Orthodontics and Paediatric Dentistry, Section of Dentistry, Department of Clinical, Surgical, Diagnostic and Paediatric Sciences, University of Pavia, Pavia, Italy
                Author notes

                Academic Editor: Cristiana Corsi

                Author information
                http://orcid.org/0000-0003-3445-3289
                http://orcid.org/0000-0002-2760-0124
                Article
                10.1155/2018/4103232
                5937598
                29850512
                f9c3cc59-98e2-4c75-8509-58ebd082e01e
                Copyright © 2018 Maria Francesca Sfondrini et al.

                This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 10 November 2017
                : 2 March 2018
                : 20 March 2018
                Categories
                Research Article

                Comments

                Comment on this article