      Effects of implant thread design on primary stability—a comparison between single- and double-threaded implants in an artificial bone model

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Primary implant stability is essential for osseointegration. To increase stability without changing the implant size, the thread length must be extended by reducing pitch, using a double-threaded implant, or reducing pitch/lead and lead angle to half that of a single-threaded implant.

          Materials and methods

          We tested the stabilities of these configurations using artificial bone. A 1.2-mm pitch, single-threaded implant (12S) was the control. We tested a 0.6-mm pitch/1.2-mm-lead double-threaded implant (06D) and a 0.6-mm pitch/lead single-threaded implant (06S). We compared stabilities by measuring insertion torque, removal torque, and the implant stability quotient (ISQ). Damage to bone tissue caused by the implants was evaluated using microscopy and morphometric analysis.

          Results

          We show that 06D and 06S significantly improved stability compared with the 12S reference. The stability of 06S was significantly greater compared with that of 06D, except for ISQ. The three implants were associated with bone tissue damage characterized by debris and voids surrounding the implant/bone interface. The 06D caused the most tissue damage, followed by 06S and then 12S.

          Conclusion

          These findings indicate that primary stability was significantly improved by changing the implant size, extending the thread length with reduced pitch/lead, and reducing the lead angle to half that of a single-threaded implant compared with a double-threaded implant.

          Related collections

          Most cited references28

          • Record: found
          • Abstract: found
          • Article: not found

          Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man.

          A total of 2895 threaded, cylindrical titanium implants have been inserted into the mandible or the maxilla and 124 similar implants have been installed in the tibial, temporal or iliac bones in man for various bone restorative procedures. The titanium screws were implanted without the use of cement, using a meticulous technique aiming at osseointegration--a direct contact between living bone and implant. Thirty-eight stable and integrated screws were removed for various reasons from 18 patients. The interface zone between bone and implant was investigated using X-rays, SEM, TEM and histology. The SEM study showed a very close spatial relationship between titanium and bone. The pattern of the anchorage of collagen filaments to titanium appeared to be similar to that of Sharpey's fibres to bone. No wear products were seen in the bone or soft tissues in spite of implant loading times up to 90 months. The soft tissues were also closely adhered to the titanium implant, thereby forming a biological seal, preventing microorganism infiltration along the implant. The implants in many cases had been allowed to permanently penetrate the gingiva and skin. This caused no adverse tissue effects. An intact bone-implant interface was analyzed by TEM, revealing a direct bone-to-implant interface contact also at the electron microscopic level, thereby suggesting the possibility of a direct chemical bonding between bone and titanium. It is concluded that the technique of osseointegration is a reliable type of cement-free bone anchorage for permanent prosthetic tissue substitutes. At present, this technique is being tried in clinical joint reconstruction. In order to achieve and to maintain such a direct contact between living bone and implant, threaded, unalloyed titanium screws of defined finish and geometry were inserted using a delicate surgical technique and were allowed to heal in situ, without loading, for a period of at least 3--4 months.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Assessment of implant stability as a prognostic determinant.

            N Meredith (2025)
            This paper aims to establish the parameters necessary to monitor successful implant placement and osseointegration.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Role of primary stability for successful osseointegration of dental implants: Factors of influence and evaluation.

              A secure implant primary (mechanical) stability is positively associated with a successful implant integration and long-term successful clinical outcome. Therefore, it is essential to assess the initial stability at different time-points to ensure a successful osseointegration. The present study critically reviews the factors that may play a role in achieving a successful initial stability in dental implants. Databases were searched from 1983 up to and including October 2013 using different combinations of various keywords. Bone quality and quantity, implant geometry, and surgical technique adopted may significantly influence primary stability and overall success rate of dental implants.
                Bookmark

                Author and article information

                Contributors
                blacklab87@gmail.com
                Journal
                Int J Implant Dent
                Int J Implant Dent
                International Journal of Implant Dentistry
                Springer Berlin Heidelberg (Berlin/Heidelberg )
                2198-4034
                20 August 2020
                20 August 2020
                December 2020
                : 6
                : 42
                Affiliations
                [1 ]GRID grid.410714.7, ISNI 0000 0000 8864 3422, Department of Implant Dentistry, School of Dentistry, , Showa University, ; 2-1-1 Kitasenzoku Ota-ku, Tokyo, 145-8515 Japan
                [2 ]GRID grid.265073.5, ISNI 0000 0001 1014 9130, Oral Implantology and Regenerative Dental Medicine, Department of Masticatory Function Rehabilitation, Division of Oral Health Sciences, Graduate School, , Tokyo Medical and Dental University, ; 1-5-45 Yushima, Bunkyo-ku, Tokyo, 113-8510 Japan
                Article
                239
                10.1186/s40729-020-00239-1
                7438427
                32815046
                f7d4bb9a-e081-4543-89b5-3a4f8f45c30f
                © The Author(s) 2020

                Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

                History
                : 13 February 2020
                : 7 June 2020
                Categories
                Research
                Custom metadata
                © The Author(s) 2020

                primary stability,single-threaded implant,double-threaded implant,artificial bone,insertion torque,removal torque,low-density bone