      Short-term safety and feasibility of a practical approach to combined atrial and ventricular physiological pacing: An initial single-center experience

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Traditional right atrial appendage (RAA) pacing accentuates conduction disturbances as opposed to Bachmann bundle pacing (BBP).

          Objective

          The purpose of this study was to evaluate the feasibility, efficacy, and safety of routine anatomically guided high right atrial septal (HRAS) pacing with activation of Bachmann bundle combined with routine left bundle branch area pacing (LBBAP).

          Methods

          This retrospective single-center study included 96 consecutive patients who underwent 1 of 2 strategies: physiological pacing (PP) (n = 32) with HRAS and LBBAP leads and conventional pacing (CP) (n = 64) with traditional RAA and right ventricular apical leads. Baseline characteristics, sensing, pacing thresholds, and impedances were recorded at implantation and follow-up.

          Results

          The PP and CP cohorts were of similar age (74.2 ± 13.8 years vs 73.9 ± 9.9 years) and sex (28.1% vs 40.6% female). There were no differences in procedural time (95.0 ± 31.4 minutes vs 86.5 ± 33.3 minutes; P = .19) or fluoroscopy time (12.1 ± 4.5 minutes vs 12.3 ± 13.5 minutes; P = .89) between cohorts. After excluding patients who received >2 leads, these parameters became significantly shorter in the CP cohort. The PP cohort exhibited higher atrial pacing thresholds (1.5 ± 1.1 mV vs 0.8 ± 0.3 mV; P <.001) and lower p waves (1.8 ± 0.8 mV vs 3.8 ± 2.3 mV; P <.001) at implantation and at follow-up. In the PP cohort, 72% of implants met criteria for BBP; of the ventricular leads, 94% demonstrated evidence of LBBAP. One lead-related complication occurred in each cohort.

          Conclusion

          Routine placement of leads in the HRAS is a feasible and safe alternative to standard RAA pacing, allowing for BBP in 72% of patients. HRAS pacing can be combined with LBBAP as a routine strategy.

          Related collections

          Most cited references10

          • Record: found
          • Abstract: found
          • Article: not found

          Permanent His-bundle pacing is feasible, safe, and superior to right ventricular pacing in routine clinical practice.

          Right ventricular pacing (RVP) has been associated with heart failure and increased mortality. His-bundle pacing (HBP) is more physiological but requires a mapping catheter or a backup right ventricular lead and is technically challenging.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Left Bundle Branch Area Pacing for Cardiac Resynchronization Therapy

            The aim of this study was to assess the feasibility and outcomes of left bundle branch area pacing (LBBAP) in patients eligible for cardiac resynchronization therapy (CRT) in an international, multicenter, collaborative study.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Trends in permanent pacemaker implantation in the United States from 1993 to 2009: increasing complexity of patients and procedures.

              This study sought to define contemporary trends in permanent pacemaker use by analyzing a large national database. The Medicare National Coverage Determination for permanent pacemaker, which emphasized single-chamber pacing, has not changed significantly since 1985. We sought to define contemporary trends in permanent pacemaker use by analyzing a large national database. We queried the Nationwide Inpatient Sample to identify permanent pacemaker implants between 1993 and 2009 using the International Classification of Diseases-Ninth Revision-Clinical Modification procedure codes for dual-chamber (DDD), single-ventricular (VVI), single-atrial (AAI), or biventricular (BiV) devices. Annual permanent pacemaker implantation rates and patient demographics were analyzed. Between 1993 and 2009, 2.9 million patients received permanent pacemakers in the United States. Overall use increased by 55.6%. By 2009, DDD use increased from 62% to 82% (p < 0.001), whereas single-chamber ventricular pacemaker use fell from 36% to 14% (p = 0.01). Use of DDD devices was higher in urban, nonteaching hospitals (79%) compared with urban teaching hospitals (76%) and rural hospitals (72%). Patients with private insurance (83%) more commonly received DDD devices than Medicaid (79%) or Medicare (75%) recipients (p < 0.001). Patient age and Charlson comorbidity index increased over time. Hospital charges ($2011) increased 45.3%, driven by the increased cost of DDD devices. There is a steady growth in the use of permanent pacemakers in the United States. Although DDD device use is increasing, whereas single-chamber ventricular pacemaker use is decreasing. Patients are becoming older and have more medical comorbidities. These trends have important health care policy implications. Copyright © 2012 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
                Bookmark

                Author and article information

                Contributors
                Journal
                Heart Rhythm O2
                Heart Rhythm O2
                Heart Rhythm O2
                Elsevier
                2666-5018
                16 May 2024
                June 2024
                16 May 2024
                : 5
                : 6
                : 357-364
                Affiliations
                [1]Helmsley Electrophysiology Center, Department of Cardiology, Icahn School of Medicine at Mount Sinai, New York, New York
                Author notes
                [] Address reprint requests and correspondence: Dr Jacob Koruth, Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, One Gustave L. Levy Place, Box 1030, New York, NY 10029. jacob.koruth@ 123456mountsinai.org
                Article
                S2666-5018(24)00097-7
                10.1016/j.hroo.2024.04.002
                11228114
                e09f4ba7-6679-4de8-891d-ce2b03e5eb5c
                © 2024 Heart Rhythm Society. Published by Elsevier Inc.

                This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).

                History
                Categories
                Clinical
                Devices

                atrialmyocardium,bachmann bundle,interatrial septum,interventricular septum,left bundle,pacing,right atrial appendage

                Comments

                Comment on this article