      Subsurface ultrasound microscopic imaging of the intact eye.

      Ophthalmology
      Anterior Chamber, anatomy & histology, Cornea, Eye, Humans, Image Processing, Computer-Assisted, Microscopy, methods, Retina, Sclera, Ultrasonics

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The authors have developed a method of obtaining images of cross sections of the intact eye at microscopic resolution. High-frequency ultrasound transducers (100 MHz) have been developed and incorporated into imaging devices. These devices are capable of producing images to a depth of 4 mm at an axial and lateral resolution approaching 20 microns. Resolution exceeds that of current combined A- and B-scan imaging devices by a factor of approximately 10. Microscopic images of ocular structures including Schlemm's canal, cornea, iris, ciliary muscles, and retina have been produced in eye bank eyes. These studies show the feasibility of developing an apparatus to be used in the clinical setting for examining anterior structures of the eye not visible by current techniques.