      Dosimetry Evaluation of Treatment Planning Systems in Patient-Specific 3D Printed Anthropomorphic Phantom for Breast Cancer after Mastectomy using a Single-Beam 3D-CRT Technique for Megavoltage Electron Radiation Therapy

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background:

          The patient-specific 3D printed anthropomorphic phantom is used for breast cancer after mastectomy developed by the laboratory of medical physics and biophysics, Department of Physics, Institut Teknologi Sepuluh Nopember, Indonesia. This phantom is applied to simulate and measure the radiation interactions occurring in the human body either using the treatment planning system (TPS) or direct measurement with external beam therapy (EBT) 3 film.

          Objective:

          This study aimed to provide dose measurements in the patient-specific 3D printed anthropomorphic phantom using a TPS and direct measurements using single-beam three-dimensional conformal radiation therapy (3DCRT) technique with electron energy of 6 MeV.

          Material and Methods:

          In this experimental study, the patient-specific 3D printed anthropomorphic phantom was used for post-mastectomy radiation therapy. TPS on the phantom was conducted using a 3D-CRT technique with RayPlan 9A software. The single-beam radiation was delivered to the phantom with an angle perpendicular to the breast plane at 337.3° at 6 MeV with a total prescribed dose of 5000 cGy/25 fractions with 200 cGy per fraction.

          Results:

          The doses at planning target volume (PTV) and right lung confirmed a non-significant difference both for TPS and direct measurement with P-values of 0.074 and 0.143, respectively. The dose at the spinal cord showed statistically significant differences with a P-value of 0.002. The result presented a similar skin dose value using either TPS or direct measurement.

          Conclusion:

          The patient-specific 3D printed anthropomorphic phantom for breast cancer after mastectomy on the right side has good potential as an alternative to the evaluation of dosimetry for radiation therapy.

          Related collections

          Most cited references24

          • Record: found
          • Abstract: found
          • Article: not found

          Use of normal tissue complication probability models in the clinic.

          The Quantitative Analysis of Normal Tissue Effects in the Clinic (QUANTEC) review summarizes the currently available three-dimensional dose/volume/outcome data to update and refine the normal tissue dose/volume tolerance guidelines provided by the classic Emami et al. paper published in 1991. A "clinician's view" on using the QUANTEC information in a responsible manner is presented along with a description of the most commonly used normal tissue complication probability (NTCP) models. A summary of organ-specific dose/volume/outcome data, based on the QUANTEC reviews, is included. Copyright 2010 Elsevier Inc. All rights reserved.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Report of AAPM Task Group 235 Radiochromic Film Dosimetry: An Update to TG-55.

            The use of radiochromic film (RCF) dosimetry in radiation therapy is extensive due to its high level of achievable accuracy for a wide range of dose values and its suitability under a variety of measurement conditions. However, since the publication of the 1998 AAPM Task Group 55, Report No. 63 on RCF dosimetry, the chemistry, composition, and readout systems for RCFs have evolved steadily. There are several challenges in using the new RCFs, readout systems and validation of the results depending on their applications. Accurate RCF dosimetry requires understanding of RCF selection, handling and calibration methods, calibration curves, dose conversion methods, correction methodologies as well as selection, operation and quality assurance (QA) programs of the readout systems. Acquiring this level of knowledge is not straight forward, even for some experienced users. This Task Group report addresses these issues and provides a basic understanding of available RCF models, dosimetric characteristics and properties, advantages and limitations, configurations, and overall elemental compositions of the RCFs that have changed over the past 20 yr. In addition, this report provides specific guidelines for data processing and analysis schemes and correction methodologies for clinical applications in radiation therapy.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Evaluation of Gafchromic® EBT3 films characteristics in therapy photon, electron and proton beams.

              To evaluate the uncertainties and characteristics of radiochromic film-based dosimetry system using the EBT3 model Gafchromic(®) film in therapy photon, electron and proton beams. EBT3 films were read using an EPSON Expression 10000XL/PRO scanner. They were irradiated in five beams, an Elekta SL25 6 MV and 18 MV photon beam, an IBA 100 MeV 5 × 5 cm(2) proton beam delivered by pencil-beam scanning, a 60 MeV fixed proton beam and an Elekta SL25 6 MeV electron beam. Reference dosimetry was performed using a FC65-G chamber (Elekta beam), a PPC05 (IBA beam) and both Markus 1916 and PPC40 Roos ion-chambers (60 MeV proton beam). Calibration curves of the radiochromic film dosimetry system were acquired and compared within a dose range of 0.4-10 Gy. An uncertainty budget was estimated on films irradiated by Elekta SL25 by measuring intra-film and inter-film reproducibility and uniformity; scanner uniformity and reproducibility; room light and film reading delay influences. The global uncertainty on acquired optical densities was within 0.55% and could be reduced to 0.1% by placing films consistently at the center of the scanner. For all beam types, the calibration curves are within uncertainties of measured dose and optical densities. The total uncertainties on calibration curve due to film reading and fitting were within 1.5% for photon and proton beams. For electrons, the uncertainty was within 2% for dose superior to 0.8 Gy. The low combined uncertainty observed and low beam and energy-dependence make EBT3 suitable for dosimetry in various applications. Copyright © 2012 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
                Bookmark

                Author and article information

                Journal
                J Biomed Phys Eng
                J Biomed Phys Eng
                Journal of Biomedical Physics & Engineering
                Shiraz University of Medical Sciences (Iran )
                2251-7200
                June 2023
                01 June 2023
                : 13
                : 3
                : 217-226
                Affiliations
                [1 ] Department of Physics, Institut Teknologi Sepuluh Nopember, Kampus ITS - Sukolilo Surabaya 60111, East Java, Indonesia
                [2 ] Medical Physicist of Radiotherapy Installation, Naval Hospital Dr. Ramelan, Surabaya 60244, East Java, Indonesia
                Author notes
                *Corresponding author: Endarko Endarko, Department of Physics, Institut Teknologi Sepuluh Nopember, Kampus ITS - Sukolilo Surabaya 60111, East Java, Indonesia
                Article
                JBPE-13-3
                10.31661/jbpe.v0i0.2111-1428
                10258204
                c54232ef-8f1b-41e2-bd14-7f2802db3d96
                Copyright: © Journal of Biomedical Physics and Engineering

                This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 Unported License, ( http://creativecommons.org/licenses/by-nc/4.0/ ) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 20 March 2022
                : 8 November 2021
                Categories
                Original Article

                electrons, energy, dosimetry, ebt3, phantom, radiation, mastectomy, three-dimensional therapy