      Acuros CTS: A fast, linear Boltzmann transport equation solver for computed tomography scatter – Part I: Core algorithms and validation

      research-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose

          To describe Acuros ® CTS, a new software tool for rapidly and accurately estimating scatter in x-ray projection images by deterministically solving the linear Boltzmann transport equation (LBTE).

          Methods

          The LBTE describes the behavior of particles as they interact with an object across spatial, energy, and directional (propagation) domains. Acuros CTS deterministically solves the LBTE by modeling photon transport associated with an x-ray projection in three main steps: 1) Ray tracing photons from the x-ray source into the object where they experience their first scattering event and form scattering sources. 2) Propagating photons from their first scattering sources across the object in all directions to form second scattering sources, then repeating this process until all high order scattering sources are computed by using the Source Iteration method. 3) Ray tracing photons from scattering sources within the object to the detector, accounting for the detector’s energy and anti-scatter grid responses.

          To make this process computationally tractable, a combination of analytical and discrete methods is applied. The three domains are discretized using the Linear Discontinuous Finite Elements, Multigroup, and Discrete Ordinates methods, respectively, which confer the ability to maintain the accuracy of a continuous solution. Furthermore, through implementation in CUDA, we sought to exploit the parallel computing capabilities of graphics processing units (GPUs) to achieve the speeds required for clinical utilization.

          Acuros CTS was validated against Geant4 Monte Carlo simulations using two digital phantoms: 1) a water phantom containing lung, air, and bone inserts (WLAB phantom) and 2) a pelvis phantom derived from a clinical CT data set. For these studies, we modeled the TrueBeam ® (Varian Medical Systems, Palo Alto, CA) kV imaging system with a source energy of 125kVp. The imager comprised a 600 μm-thick Cesium Iodide (CsI) scintillator and a 10:1 one-dimensional anti-scatter grid. For the WLAB studies, the full-fan geometry without a bowtie filter was used (with and without the anti-scatter grid). For the pelvis phantom studies, a half-fan geometry with bowtie was used (with the anti-scatter grid). Scattered and primary photon fluences and energies deposited in the detector were recorded.

          Results

          The Acuros CTS and Monte Carlo results demonstrated excellent agreement. For the WLAB studies, the average percent difference between the Monte Carlo- and Acuros-generated scattered photon fluences at the face of the detector was −0.7%. After including the detector response, the average percent differences between the Monte Carlo- and Acuros- generated scatter fractions (SF) were −0.1% without the grid and 0.6% with the grid. For the digital pelvis simulation, the Monte Carlo- and Acuros-generated SFs agreed to within 0.1% on average, despite the scatter-to-primary ratios (SPRs) being as high as 5.5. The Acuros CTS computation time for each scatter image was ~1 second using a single GPU.

          Conclusions

          Acuros CTS enables a fast and accurate calculation of scatter images by deterministically solving the LBTE thus offering a computationally attractive alternative to Monte Carlo methods. Part II describes the application of Acuros CTS to scatter correction of CBCT scans on the TrueBeam system.

          Related collections

          Author and article information

          Journal
          0425746
          5648
          Med Phys
          Med Phys
          Medical physics
          0094-2405
          2473-4209
          15 March 2018
          06 April 2018
          May 2018
          01 May 2019
          : 45
          : 5
          : 1899-1913
          Affiliations
          [1 ]Varian Medical Systems, Palo Alto, CA 94304 USA
          Author notes
          [* ]Corresponding author: Adam Wang, PhD, Varian Medical Systems, 3120 Hansen Way, G205, Palo Alto, CA 94034, adam.wang@ 123456varian.com
          Article
          PMC5948176 PMC5948176 5948176 nihpa948726
          10.1002/mp.12850
          5948176
          29509970
          c3fdcc50-8ac5-4f38-ae15-6e70940b560f
          History
          Categories
          Article

          Comments

          Comment on this article