      Low radiation dose 256-MDCT angiography of the carotid arteries: effect of hybrid iterative reconstruction technique on noise, artifacts, and image quality.

      European Journal of Radiology
      Angiography, methods, Artifacts, Carotid Artery Diseases, radiography, Female, Humans, Male, Middle Aged, Multidetector Computed Tomography, Radiation Dosage, Radiation Protection, Radiographic Image Interpretation, Computer-Assisted, Reproducibility of Results, Sensitivity and Specificity, Signal-To-Noise Ratio

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          To evaluate the effect of hybrid iterative reconstruction on qualitative and quantitative parameters at low dose carotid CTA. 44 consecutive patients were enrolled in the study. First group (n=22) was examined under 120 kV 250 mAs, second group (n = 22) - 100 kV 250 mAs. CT images in first group were reconstructed only with the filtered back projection (FBP). CT data in second group were reconstructed both with FBP and three levels of hybrid iterative reconstruction algorithm (iDose). We compared quantitative and qualitative parameters among the two groups and among four different reconstructions in second group. Effective dose in 120 kV and 100 kV group was 7.18 ± 1.19 mSv and 4.14 ± 1.03 mSv, respectively (p<0.0001). Mean arterial attenuation was about 25% higher in second group (236.5 ± 46 HU vs. 302.6 ± 32.7 HU; p<0.0001). Image noise at the level of humeral belt was 32.5 ± 12.5 in 100 kV group and 26.3 ± 13.3 in 120 kV (p = 0.115). Average noise decreased when using 3 levels of iDose up to 23.6 ± 6.4, 17.7 ± 5.6 and 13.7 ± 5.1, respectively (p = 0.00001). Mean CNR increased to 10.38 ± 3.87, 14.5 ± 5.21 and 18.32 ± 8.61, respectively (p<0.05). The presence of artifacts on the level of humeral belt in 120 kV group was 14%, in 100 kV - 41% (p = 0.002). The difference in visual scores between standard and low-dose protocol was significant (p = 0.008). When applying iterative reconstruction the frequency of streak artifacts decreased dramatically (p<0.0001). Most studies had excellent quality with no artifacts while using highest level of iDose. According to the results of our study low dose CT angiography using hybrid iterative reconstruction may provide sufficient image quality and allows for significant reduction of patient dose. Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.

          Related collections

          Author and article information

          Comments

          Comment on this article