      Metal artifact reduction in CT, a phantom study: subjective and objective evaluation of four commercial metal artifact reduction algorithms when used on three different orthopedic metal implants

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Metal implants may introduce severe artifacts in computed tomography (CT) images. Over the last few years dedicated algorithms have been developed in order to reduce metal artifacts in CT images.

          Purpose

          To investigate and compare metal artifact reduction algorithms (MARs) from four different CT vendors when imaging three different orthopedic metal implants.

          Material and Methods

          Three clinical metal implants were attached to the leg of an anthropomorphic phantom: cobalt-chrome; stainless steel; and titanium. Four commercial MARs were investigated: SmartMAR (GE); O-MAR (Philips); iMAR (Siemens); and SEMAR (Toshiba). The images were evaluated subjectively by three observers and analyzed objectively by calculating the fraction of pixels with CT number above 500 HU in a region of interest around the metal. The average CT number and image noise were also measured.

          Results

          Both subjective evaluation and objective analysis showed that MARs reduced metal artifacts and improved the image quality for CT images containing metal implants of steel and cobalt-chrome. When using MARs on titanium, all MARs introduced new visible artifacts.

          Conclusion

          The effect of MARs varied between CT vendors and different metal implants used in orthopedic surgery. Both in subjective evaluation and objective analysis the effect of applying MARs was most obvious on steel and cobalt-chrome implants when using SEMAR from Toshiba followed by SmartMAR from GE. However, MARs may also introduce new image artifacts especially when used on titanium implants. Therefore, it is important to reconstruct all CT images containing metal with and without MARs.

          Related collections

          Author and article information

          Journal
          Acta Radiologica
          Acta Radiol
          SAGE Publications
          0284-1851
          1600-0455
          September 2018
          January 08 2018
          September 2018
          : 59
          : 9
          : 1110-1118
          Affiliations
          [1 ]Department of Oncology and Medical Physics, Haukeland University Hospital, Bergen, Norway
          [2 ]Department of Radiology, Haukeland University Hospital, Bergen, Norway
          Article
          10.1177/0284185117751278
          29310445
          aa00d388-5b35-42c9-af80-2421de66ac36
          © 2018

          http://journals.sagepub.com/page/policies/text-and-data-mining-license

          History

          Comments

          Comment on this article