      Deterioration of the fixation segment’s stress distribution and the strength reduction of screw holding position together cause screw loosening in ALSR fixed OLIF patients with poor BMD

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The vertebral body’s Hounsfield unit (HU) value can credibly reflect patients’ bone mineral density (BMD). Given that poor bone-screw integration initially triggers screw loosening and regional differences in BMD and strength in the vertebral body exist, HU in screw holding planes should better predict screw loosening. According to the stress shielding effect, the stress distribution changes in the fixation segment with BMD reduction should be related to screw loosening, but this has not been identified. We retrospectively collected the radiographic and demographic data of 56 patients treated by single-level oblique lumbar interbody fusion (OLIF) with anterior lateral single rod (ALSR) screw fixation. BMD was identified by measuring HU values in vertebral bodies and screw holding planes. Regression analyses identified independent risk factors for cranial and caudal screw loosening separately. Meanwhile, OLIF with ALSR fixation was numerically simulated; the elastic modulus of bony structures was adjusted to simulate different grades of BMD reduction. Stress distribution changes were judged by computing stress distribution in screws, bone-screw interfaces, and cancellous bones in the fixation segment. The results showed that HU reduction in vertebral bodies and screw holding planes were independent risk factors for screw loosening. The predictive performance of screw holding plane HU is better than the mean HU of vertebral bodies. Cranial screws suffer a higher risk of screw loosening, but HU was not significantly different between cranial and caudal sides. The poor BMD led to stress concentrations on both the screw and bone-screw interfaces. Biomechanical deterioration was more severe in the cranial screws than in the caudal screws. Additionally, lower stress can also be observed in fixation segments’ cancellous bone. Therefore, a higher proportion of ALSR load transmission triggers stress concentration on the screw and bone-screw interfaces in patients with poor BMD. This, together with decreased bony strength in the screw holding position, contributes to screw loosening in osteoporotic patients biomechanically. The trajectory optimization of ALSR screws based on preoperative HU measurement and regular anti-osteoporosis therapy may effectively reduce the risk of screw loosening.

          Related collections

          Most cited references74

          • Record: found
          • Abstract: found
          • Article: not found

          Opportunistic screening for osteoporosis using abdominal computed tomography scans obtained for other indications.

          Osteoporosis is a prevalent but underdiagnosed condition.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Trabecular bone modulus-density relationships depend on anatomic site.

            One outstanding issue regarding the relationship between elastic modulus and density for trabecular bone is whether the relationship depends on anatomic site. To address this, on-axis elastic moduli and apparent densities were measured for 142 specimens of human trabecular bone from the vertebra (n=61), proximal tibia (n=31), femoral greater trochanter (n=23), and femoral neck (n=27). Specimens were obtained from 61 cadavers (mean+/-SD age=67+/-15 years). Experimental protocols were used that minimized end-artifact errors and controlled for specimen orientation. Tissue moduli were computed for a subset of 18 specimens using high-resolution linear finite element analyses and also using two previously developed theoretical relationships (Bone 25 (1999) 481; J. Elasticity 53 (1999) 125). Resultant power law regressions between modulus and density did depend on anatomic site, as determined via an analysis of covariance. The inter-site differences were among the leading coefficients (p 0.08), which ranged 1.49-2.18. At a given density, specimens from the tibia had higher moduli than those from the vertebra (p=0.01) and femoral neck (p=0.002); those from the trochanter had higher moduli than the vertebra (p=0.02). These differences could be as large as almost 50%, and errors in predicted values of modulus increased by up to 65% when site-dependence was ignored. These results indicate that there is no universal modulus-density relationship for on-axis loading. Tissue moduli computed using methods that account for inter-site architectural variations did not differ across site (p>0.15), suggesting that the site-specificity in apparent modulus-density relationships may be attributed to differences in architecture.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Comparison of eight published static finite element models of the intact lumbar spine: predictive power of models improves when combined together.

              Finite element (FE) model studies have made important contributions to our understanding of functional biomechanics of the lumbar spine. However, if a model is used to answer clinical and biomechanical questions over a certain population, their inherently large inter-subject variability has to be considered. Current FE model studies, however, generally account only for a single distinct spinal geometry with one set of material properties. This raises questions concerning their predictive power, their range of results and on their agreement with in vitro and in vivo values. Eight well-established FE models of the lumbar spine (L1-5) of different research centers around the globe were subjected to pure and combined loading modes and compared to in vitro and in vivo measurements for intervertebral rotations, disc pressures and facet joint forces. Under pure moment loading, the predicted L1-5 rotations of almost all models fell within the reported in vitro ranges, and their median values differed on average by only 2° for flexion-extension, 1° for lateral bending and 5° for axial rotation. Predicted median facet joint forces and disc pressures were also in good agreement with published median in vitro values. However, the ranges of predictions were larger and exceeded those reported in vitro, especially for the facet joint forces. For all combined loading modes, except for flexion, predicted median segmental intervertebral rotations and disc pressures were in good agreement with measured in vivo values. In light of high inter-subject variability, the generalization of results of a single model to a population remains a concern. This study demonstrated that the pooled median of individual model results, similar to a probabilistic approach, can be used as an improved predictive tool in order to estimate the response of the lumbar spine. Copyright © 2014 Elsevier Ltd. All rights reserved.
                Bookmark

                Author and article information

                Contributors
                Journal
                Front Bioeng Biotechnol
                Front Bioeng Biotechnol
                Front. Bioeng. Biotechnol.
                Frontiers in Bioengineering and Biotechnology
                Frontiers Media S.A.
                2296-4185
                30 August 2022
                2022
                : 10
                : 922848
                Affiliations
                [1] 1 Department of Orthopedic Surgery and Orthopedic Research Institute , West China Hospital/West China School of Medicine for Sichuan University , Chengdu, China
                [2] 2 Department of Imaging , West China Hospital , Chengdu, China
                Author notes

                Edited by: Keitaro Matsukawa, Murayama Medical Center (NHO), Japan

                Reviewed by: Xin Zhao, Shanghai Jiao Tong University, China

                Panagiotis Chatzistergos, Staffordshire University, United Kingdom

                Wen Yuan, Shanghai Changzheng Hospital, China

                Yun Peng, NuVasive, United States

                *Correspondence: Yue-Ming Song, sym_cd@ 123456163.com ; Jian-Cheng Zeng, tomzeng5@ 123456126.com
                [ † ]

                These authors have contributed equally to this work and share first authorship

                This article was submitted to Biomechanics, a section of the journal Frontiers in Bioengineering and Biotechnology

                Article
                922848
                10.3389/fbioe.2022.922848
                9468878
                36110315
                75543dce-0872-4782-b3fb-a35fccf9156b
                Copyright © 2022 Li, Yang, Xie, Song, Song and Zeng.

                This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

                History
                : 18 April 2022
                : 29 July 2022
                Categories
                Bioengineering and Biotechnology
                Original Research

                oblique lumbar interbody fusion,screw loosening,biomechanical deterioration,anterior lateral single rod fixation,screw holding plane,stress distribution

                Comments

                Comment on this article