      Relevant Aspects of the Dental Implant Design on the Insertion Torque, Resonance Frequency Analysis (RFA) and Micromobility: An In Vitro Study

      , , , , ,
      Journal of Clinical Medicine
      MDPI AG

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The major problems for the osseointegration of dental implants are the loosening of the screw that fixes the dental implant to the abutment and the micromovements that are generated when mechanical loads are applied. In this work, torque differences in the tightening and loosening of the connection screws after 1 cycle, 10 cycles and 1000 cycles for 4 dental implants with 2 external and 2 internal connections were analyzed. The loosening of 240 implants (60 for each system) was determined using high-precision torsimeters and an electromechanical testing machine. A total of 60 dental implants for each of the 4 systems were inserted into fresh bovine bone to determine the micromovements. The implant stability values (ISQ) were determined by RFA. The mechanical loads were performed at 30° from 20 N to 200 N. By means of the Q-star technique, the micromovements were determined. It was observed that, for a few cycles, the loosening of the screw did not exceed a loss of tightening of 10% for both connections. However, for 1000 cycles, the loss for the external connection was around 20% and for the internal connection it was 13%. The micromovements showed a lineal increase with the applied load for the implant systems studied. An external connection presented greater micromotions for each level of applied load and lower ISQ values than internal ones. An excellent lineal correlation between the ISQ and micromobility was observed. These results may be very useful for clinicians in the selection of the type of dental implant, depending on the masticatory load of the patient as well as the consequences of the insertion torque of the dental implant and its revisions.

          Related collections

          Most cited references46

          • Record: found
          • Abstract: found
          • Article: not found

          Role of primary stability for successful osseointegration of dental implants: Factors of influence and evaluation.

          A secure implant primary (mechanical) stability is positively associated with a successful implant integration and long-term successful clinical outcome. Therefore, it is essential to assess the initial stability at different time-points to ensure a successful osseointegration. The present study critically reviews the factors that may play a role in achieving a successful initial stability in dental implants. Databases were searched from 1983 up to and including October 2013 using different combinations of various keywords. Bone quality and quantity, implant geometry, and surgical technique adopted may significantly influence primary stability and overall success rate of dental implants.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Resonance frequency analysis of implants subjected to immediate or early functional occlusal loading. Successful vs. failing implants.

            The objective of this study was to analyze the development of implant stability by repeated resonance frequency analysis (RFA) measurements during 1 year in 23 patients treated according to an immediate/early-loading protocol. The objective was also to evaluate the possible differences between failing and successful implants. Eighty-one Brånemark System implants were placed in 23 patients for immediate/early-occlusal loading in all jaw regions. Thirty of the implants were placed in extraction sockets and 62 were subjected to GBR procedures. Apart from clinical and radiographic examinations, the patients were followed with RFA at placement, prosthesis connection and after 1-3, 6 and 12 months. Statistical analyses were carried out to study the possible differences between implants that failed during the study period and implants that remained successful. Nine implants failed (11.2%) during the 1 year of loading. RFA showed a distinct different pattern between the implants that remained stable and the implants that were lost. The implants that failed during the course of the study showed a significantly lower stability already after 1 month. Within the limitations of this study, it is concluded that failing implants show a continuous decrease of stability until failure. Low RFA levels after 1 and 2 months seem to indicate an increased risk for future failure. This information may be used to avoid implant failure in the future by unloading implants with decreasing degree of stability with time as diagnosed with the RFA technique.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Relationship between implant preload and screw loosening on implant-supported prostheses.

              The purpose of this study was to determine whether varying the preload on the implant-abutment complex would affect screw loosening under simulated loading conditions. Abutment screws in sample models were tightened to 25, 30, 35, and 40 N-cm. One group of samples was allowed to stand for 3 hours after being torqued and then loosened. Another group of samples was retorqued after 10 minutes with the same initial torque value and then allowed to stand for 3 hours before loosening. For the load group of samples, the abutments were torqued into place, retorqued after 10 minutes, and a load applied for 3 hours before loosening. Cyclic loading was carried out using a servo-hydraulic testing machine with the values cycled between 1 and 26 pounds and the load applied directly to the abutments. Analysis of variance, analysis of covariance, and linear regression analysis was performed. Within the parameters of this in vitro investigation, the following recommendations can be made: (1) retightening abutment screws 10 minutes after the initial torque applications should be routinely performed and (2) increasing the torque value for abutment screws above 30 N-cm can be beneficial for abutment-implant stability and to decrease screw loosening.
                Bookmark

                Author and article information

                Contributors
                (View ORCID Profile)
                (View ORCID Profile)
                (View ORCID Profile)
                (View ORCID Profile)
                Journal
                JCMOHK
                Journal of Clinical Medicine
                JCM
                MDPI AG
                2077-0383
                February 2023
                January 20 2023
                : 12
                : 3
                : 855
                Article
                10.3390/jcm12030855
                4c6de0f0-3435-4de7-827d-ff74376860d4