Vitek 2 (bioMérieux Inc., Durham, NC) is a widely used commercial antimicrobial susceptibility test system. We compared the MIC results obtained using the Vitek 2 AST-GN69 and AST-XN06 cards to those obtained by CLSI broth microdilution (BMD) for 255 isolates of Enterobacteriaceae, including 25 isolates of carbapenem-resistant Enterobacteriaceae. In total, 25 antimicrobial agents were examined. For 10 agents, the MIC data were evaluated using two sets of breakpoints: (i) the Vitek 2 breakpoints, which utilized the 2009 FDA breakpoints at the time of the study and are equivalent to the 2009 CLSI M100-S19 breakpoints, and (ii) the 2014 CLSI M100-S24 breakpoints. There was an overall 98.7% essential agreement (EA). The categorical agreement was 95.5% (CA) using the Vitek 2 breakpoints and 95.7% using the CLSI breakpoints. There was 1 very major error (VME) (0.05%) observed using the Vitek 2 breakpoints (cefazolin) and 8 VMEs (0.5%) using the CLSI breakpoints (2 each for aztreonam, cefepime, and ceftriaxone, and 1 for cefazolin and ceftazidime). Fifteen major errors (MEs) (0.4%) were noted using the Vitek 2 breakpoints and 8 (0.5%) using the CLSI breakpoints. Overall, the Vitek 2 performance was comparable to that of BMD for testing a limited number of Enterobacteriaceaecommonly isolated by clinical laboratories. Ongoing studies are warranted to assess performance in isolates with emerging resistance.
See how this article has been cited at scite.ai
scite shows how a scientific paper has been cited by providing the context of the citation, a classification describing whether it supports, mentions, or contrasts the cited claim, and a label indicating in which section the citation was made.