      Evaluation of non-instrumented nucleic acid amplification by loop-mediated isothermal amplification (NINA-LAMP) for the diagnosis of malaria in Northwest Ethiopia

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Malaria is a major public health problem in sub-Saharan African countries including Ethiopia. Early and accurate diagnosis followed by prompt and effective treatment is among the various tools available for prevention, control and elimination of malaria. This study aimed to evaluate the performance of non-instrumented nucleic acid amplification loop-mediated isothermal amplification (NINA-LAMP) compared to standard thick and thin film microscopy and nested PCR as gold standard for the sensitive diagnosis of malaria in Northwest Ethiopia.

          Methods

          A cross-sectional study was conducted in North Gondar, Ethiopia from March to July 2014. Eighty-two blood samples were collected from malaria suspected patients visiting Kola Diba Health Centre and analysed for Plasmodium parasites by microscopy, NINA-LAMP and nested PCR. The NINA-LAMP method was performed using the Loopamp™ Malaria Pan/Pf detection kits for detecting DNA of the genus Plasmodium and more specifically Plasmodium falciparum using an electricity-free heater. Diagnostic accuracy outcome measures (analytical sensitivity, specificity, predictive values, and Kappa scores) of NINA-LAMP and microscopy were compared to nested PCR.

          Results

          A total of 82 samples were tested in the primary analysis. Using nested PCR as reference, the sensitivity and specificity of the primary NINA-LAMP assay were 96.8% (95% confidence interval (CI), 83.2% - 99.5%) and 84.3% (95% CI, 71.4% - 92.9%), respectively for detection of Plasmodium genus, and 100% (95% CI, 75.1% - 100%) and 81.2% (95% CI, 69.9% - 89.6%), respectively for detection of P. falciparum parasite. Microscopy demonstrated sensitivity and specificity of 93.6% (95% CI, 78.5% - 99.0%) and 98.0% (95% CI, 89.5% - 99.7%), respectively for the detection of Plasmodium parasites. Post-hoc repeat NINA-LAMP analysis showed improvement in diagnostic accuracy, which was comparable to nested PCR performance and superior to microscopy for detection at both the Plasmodium genus level and P. falciparum parasites.

          Conclusion

          NINA-LAMP is highly sensitive for the diagnosis of malaria and detection of Plasmodium parasite infection at both the genus and species level when compared to nested PCR. NINA-LAMP is more sensitive than microscopy for the detection of P. falciparum and differentiation from non-falciparum species and may be a critical diagnostic modality in efforts to eradicate malaria from areas of low endemicity.

          Related collections

          Most cited references36

          • Record: found
          • Abstract: found
          • Article: not found

          Microfluidic diagnostic technologies for global public health.

          The developing world does not have access to many of the best medical diagnostic technologies; they were designed for air-conditioned laboratories, refrigerated storage of chemicals, a constant supply of calibrators and reagents, stable electrical power, highly trained personnel and rapid transportation of samples. Microfluidic systems allow miniaturization and integration of complex functions, which could move sophisticated diagnostic tools out of the developed-world laboratory. These systems must be inexpensive, but also accurate, reliable, rugged and well suited to the medical and social contexts of the developing world.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            A review of malaria diagnostic tools: microscopy and rapid diagnostic test (RDT).

            The absolute necessity for rational therapy in the face of rampant drug resistance places increasing importance on the accuracy of malaria diagnosis. Giemsa microscopy and rapid diagnostic tests (RDTs) represent the two diagnostics most likely to have the largest impact on malaria control today. These two methods, each with characteristic strengths and limitations, together represent the best hope for accurate diagnosis as a key component of successful malaria control. This review addresses the quality issues with current malaria diagnostics and presents data from recent rapid diagnostic test trials. Reduction of malaria morbidity and drug resistance intensity plus the associated economic loss of these two factors require urgent scaling up of the quality of parasite-based diagnostic methods. An investment in anti-malarial drug development or malaria vaccine development should be accompanied by a parallel commitment to improve diagnostic tools and their availability to people living in malarious areas.
              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              World Malaria Report

                Bookmark

                Author and article information

                Contributors
                mesem2000@gmail.com
                yanbule@gmail.com
                agbayih@ucalgary.ca
                sisaygetie@yahoo.com
                ggebeyaw@yahoo.com
                dguelig@path.org
                rburton@path.org
                plabarre@path.org
                drpillai@ucalgary.ca
                Journal
                Malar J
                Malar. J
                Malaria Journal
                BioMed Central (London )
                1475-2875
                28 January 2015
                28 January 2015
                2015
                : 14
                : 44
                Affiliations
                [ ]Department of Medical Laboratory Sciences, College of Medicine and Health Sciences, Wollo University, Dessie, Ethiopia
                [ ]School of Medicine, College of Health Sciences and Medicine, Wolaita Sodo University, Wolaita, Ethiopia
                [ ]Department of Pathology and Laboratory Medicine, University of Calgary, Calgary, Canada
                [ ]Department of Medical Parasitology, School of Biomedical and Laboratory Sciences, College of Medicine and Health Sciences, University of Gondar, Gondar, Ethiopia
                [ ]PATH, Seattle, USA
                Article
                559
                10.1186/s12936-015-0559-9
                4323137
                25626339
                0caa85a8-b52d-4d64-b265-33c139f054be
                © Sema et al.; licensee BioMed Central. 2015

                This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

                History
                : 21 October 2014
                : 13 January 2015
                Categories
                Research
                Custom metadata
                © The Author(s) 2015

                Infectious disease & Microbiology
                Infectious disease & Microbiology

                Comments

                Comment on this article