      A highly sensitive prostate-specific antigen immunosensor based on gold nanoparticles/PAMAM dendrimer loaded on MWCNTS/chitosan/ionic liquid nanocomposite.

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          We have developed a sensitive electrochemical immunosensor for the detection of prostate-specific antigen (PSA), based on covalently immobilizing of anti-PSA and redox mediator (thionine) onto gold nanoparticles-incorporated polyamidoamine dendrimer (AuNPs-PAMAM) and multiwalled carbon nanotubes/ionic liquid/chitosan nanocomposite (MWCNTs/IL/Chit) as the support platform. The MWCNTs/IL/Chit nanocomposite and synthesized AuNPs were characterized using SEM and TEM microscopy techniques. Greatly amplified immunoassay was established by sandwiching the antigen between anti-PSA immobilized on the MWCNTs/IL/Chit/AuNPs-PAMAM interface and anti-PSA labeled with horseradish peroxidase (HRP-labeled anti-PSA) as secondary antibody. Phtaloyl chloride (Ph) was used as linking agent for the subsequent immobilization of AuNPs-PAMAM onto platform and anti-PSA antibody and thionine onto AuNPs-PAMAM dendrimer. The increased electrocatalytic reduction of H2O2 by HRP was monitored by differential pulse voltammetry technique. Under optimized condition the calibration curve for PSA concentration was linear up to 80 ng ml(-1) with detection limit (signal-to-noise ratio of 3) of 1 pg ml(-1). AuNPs-PAMAM dendrimer as platform not only increased the amount of thionine and PSA antibody but also the electron transfer process accelerated by encapsulated AuNPs. Moreover, the proposed PSA immunosensor exhibited excellent stability and reproducibility. Accurate detection of PSA in human serum samples was demonstrated by comparison to standard ELISA assays. In addition, impedance technique was used as simple, rapid, low cost label free analytical method for PSA measurement with detection limit of 0.5 ng ml(-1) at concentration range up to 25 ng ml(-1). The results indicate that the present protocol is quite promising in developing other electrochemical immunosensors.

          Related collections

          Author and article information

          Journal
          Biosens Bioelectron
          Biosensors & bioelectronics
          Elsevier BV
          1873-4235
          0956-5663
          Feb 15 2014
          : 52
          Affiliations
          [1 ] Department of Chemistry, University of Kurdistan, 66177-15175, Sanandaj, Iran.
          Article
          S0956-5663(13)00553-8
          10.1016/j.bios.2013.08.012
          24016535
          0726d759-a17e-4ff1-9e21-57946ed48c0b
          History

          Carbon nanotube,Chitosan,Dendrimer,Electrochemical immunosensor,Gold nanoparticles,Ionic liquid,Prostate specific antigen

          Comments

          Comment on this article