      Viscoelastic Hemostatic Assays: Moving from the Laboratory to the Site of Care—A Review of Established and Emerging Technologies

      review-article

      Read this article at

          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Viscoelastic-based techniques to evaluate whole blood hemostasis have advanced substantially since they were first developed over 70 years ago but are still based upon the techniques first described by Dr. Hellmut Hartert in 1948. Today, the use of thromboelastography, the method of testing viscoelastic properties of blood coagulation, has moved out of the research laboratory and is now more widespread, used commonly during surgery, in emergency departments, intensive care units, and in labor wards. Thromboelastography is currently a rapidly growing field of technological advancement and is attracting significant investment. This review will first describe the history of the viscoelastic testing and the established first-generation devices, which were developed for use within the laboratory. This review will then describe the next-generation hemostasis monitoring devices, which were developed for use at the site of care for an expanding range of clinical applications. This review will then move on to experimental technologies, which promise to make viscoelastic testing more readily available in a wider range of clinical environments in the endeavor to improve patient care.

          Related collections

          Most cited references43

          • Record: found
          • Abstract: found
          • Article: not found

          Coagulation monitoring: current techniques and clinical use of viscoelastic point-of-care coagulation devices.

          Perioperative monitoring of blood coagulation is critical to better understand causes of hemorrhage, to guide hemostatic therapies, and to predict the risk of bleeding during the consecutive anesthetic or surgical procedures. Point-of-care (POC) coagulation monitoring devices assessing the viscoelastic properties of whole blood, i.e., thrombelastography, rotation thrombelastometry, and Sonoclot analysis, may overcome several limitations of routine coagulation tests in the perioperative setting. The advantage of these techniques is that they have the potential to measure the clotting process, starting with fibrin formation and continue through to clot retraction and fibrinolysis at the bedside, with minimal delays. Furthermore, the coagulation status of patients is assessed in whole blood, allowing the plasmatic coagulation system to interact with platelets and red cells, and thereby providing useful additional information on platelet function. Viscoelastic POC coagulation devices are increasingly being used in clinical practice, especially in the management of patients undergoing cardiac and liver surgery. Furthermore, they provide useful information in a large variety of clinical scenarios, e.g., massive hemorrhage, assessment of hypo- and hypercoagulable states, guiding pro- and anticoagulant therapies, and in diagnosing of a surgical bleeding. A surgical etiology of bleeding has to be considered when viscoelastic test results are normal. In summary, viscoelastic POC coagulation devices may help identify the cause of bleeding and guide pro- and anticoagulant therapies. To ensure optimal accuracy and performance, standardized procedures for blood sampling and handling, strict quality controls and trained personnel are required.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Intraoperative changes in blood coagulation and thrombelastographic monitoring in liver transplantation.

            The blood coagulation system of 66 consecutive patients undergoing consecutive liver transplantations was monitored by thrombelastograph and analytic coagulation profile. A poor preoperative coagulation state, decrease in levels of coagulation factors, progressive fibrinolysis, and whole blood clot lysis were observed during the preanhepatic and anhepatic stages of surgery. A further general decrease in coagulation factors and platelets, activation of fibrinolysis, and abrupt decrease in levels of factors V and VIII occurred before and with reperfusion of the homograft. Recovery of blood coagulability began 30-60 min after reperfusion of the graft liver, and coagulability had returned toward baseline values 2 hr after reperfusion. A positive correlation was shown between the variables of thrombelastography and those of the coagulation profile. Thrombelastography was shown to be a reliable and rapid monitoring system. Its use was associated with a 33% reduction of blood and fluid infusion volume, whereas blood coagulability was maintained without an increase in the number of blood product donors.
              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              The use of viscoelastic haemostatic assays in the management of major bleeding

                Bookmark

                Author and article information

                Journal
                Diagnostics (Basel)
                Diagnostics (Basel)
                diagnostics
                Diagnostics
                MDPI
                2075-4418
                21 February 2020
                February 2020
                : 10
                : 2
                : 118
                Affiliations
                [1 ]Haemonetics Corporation, Boston, MA 02110, USA; MMurphy@ 123456Haemonetics.com
                [2 ]Haemonetics SA, Signy CH, 1274 Signy-Centre, Switzerland; joao.dias@ 123456haemonetics.com
                Author notes
                [* ]Correspondence: jan.hartmann@ 123456haemonetics.com ; Tel.: +1-781-348-7396
                Author information
                https://orcid.org/0000-0003-4978-8014
                https://orcid.org/0000-0002-1150-4357
                Article
                diagnostics-10-00118
                10.3390/diagnostics10020118
                7167835
                32098161
                650948e4-735a-405c-8654-1916db81f97b
                © 2020 by the authors.

                Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license ( http://creativecommons.org/licenses/by/4.0/).

                History
                : 28 January 2020
                : 19 February 2020
                Categories
                Review

                blood,coagulation,hemostasis,point of care,rotem,teg,thromboelastography,vha,viscoelastic testing