      Pulsatility Imaging of Saccular Aneurysm Model by 64-Slice CT with Dynamic Multiscan Technique

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The feasibility of imaging pulsatility in an aneurysm model with the high-resolution dynamic multiscan technique of 64-slice computed tomography (CT) was studied. A pulsatile aneurysm phantom was constructed and imaged with dynamic multiscan technique. The aneurysm model was filled with iodinated contrast material (250 Hounsfield Units) and was scanned with use of a gantry rotation time of 0.33 seconds, slice thickness of 1.2 mm, effective coverage of 24 mm, and total imaging time of 4 seconds. Images were reconstructed at 50-msec intervals. The visualization of wall motion was qualitatively evaluated by direct comparison of four-dimensional images versus phantom motion. Pulsatility imaging without perceptible artifact or need for cardiac gating was achieved with the use of this technique.