      Method for Accurate Quantitation of Volatile Organic Compounds in Urine Using Point of Collection Internal Standard Addition

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          A method to achieve accurate measurement of unmetabolized volatile organic compounds (VOCs) in urine was developed and characterized. The method incorporates a novel preanalytical approach of adding isotopically labeled internal standard (ISTD) analogues directly to the collection container at the point of collection to compensate for analyte loss to the headspace and the collection container surfaces. Using this approach, 45 toxic VOCs ranging in water solubility and boiling point were evaluated and analyzed by headspace solid-phase microextraction/gas chromatography–mass spectrometry. Results show that urine VOCs could be equally lost to the container headspace as to the container surface suggesting similarity of these two regions as partition phases. Surface adsorption loss was found to trend with compound water solubility. In particular, with no headspace, more nonpolar VOCs experienced substantial losses (e.g., 48% for hexane) in a standard 120 mL urine cup at concentrations in the low- and sub-ppb range. The most polar VOCs evaluated (e.g., tetrahydrofuran) showed no significant loss. Other commonly practiced methods for urine sample collection and analysis such as aliquoting, specimen freezing, and use of surrogate ISTD were found to significantly bias results. With this method, we achieved errors ranging from −8.0 to 4.8% of spiked urine specimens. Paired urine and blood specimens from cigarette smokers were compared to assess this method.

          Related collections

          Most cited references30

          • Record: found
          • Abstract: found
          • Article: not found

          Limit of blank, limit of detection and limit of quantitation.

          * Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) are terms used to describe the smallest concentration of a measurand that can be reliably measured by an analytical procedure. * LoB is the highest apparent analyte concentration expected to be found when replicates of a blank sample containing no analyte are tested. LoB = mean(blank) + 1.645(SD(blank)). * LoD is the lowest analyte concentration likely to be reliably distinguished from the LoB and at which detection is feasible. LoD is determined by utilising both the measured LoB and test replicates of a sample known to contain a low concentration of analyte. * LoD = LoB + 1.645(SD (low concentration sample)). * LoQ is the lowest concentration at which the analyte can not only be reliably detected but at which some predefined goals for bias and imprecision are met. The LoQ may be equivalent to the LoD or it could be at a much higher concentration.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Interpretation of urine results used to assess chemical exposure with emphasis on creatinine adjustments: a review.

            This paper reviews the process of elimination of creatinine (CRE), and the limitations presented when using it to express urine concentrations. This literature review leads to three conclusions: (1) CRE excretion is subject to wide fluctuations due to specific internal and external factors; (2) the use of CRE to correct chemical concentrations in urine will not necessarily improve the correlation to the exposure dose for all chemicals (it may, in fact, worsen the result); and (3) other means of expressing urine concentration may offer greater accuracy towards estimating individually absorbed dose.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Human hydration indices: acute and longitudinal reference values.

              It is difficult to describe hydration status and hydration extremes because fluid intakes and excretion patterns of free-living individuals are poorly documented and regulation of human water balance is complex and dynamic. This investigation provided reference values for euhydration (i.e., body mass, daily fluid intake, serum osmolality; M +/- SD); it also compared urinary indices in initial morning samples and 24-hr collections. Five observations of 59 healthy, active men (age 22 +/- 3 yr, body mass 75.1 +/- 7.9 kg) occurred during a 12-d period. Participants maintained detailed records of daily food and fluid intake and exercise. Results indicated that the mean total fluid intake in beverages, pure water, and solid foods was >2.1 L/24 hr (range 1.382-3.261, 95% confidence interval 0.970-3.778 L/24 hr); mean urine volume was >1.3 L/24 hr (0.875-2.250 and 0.675-3.000 L/24 hr); mean urine specific gravity was >1.018 (1.011-1.027 and 1.009-1.030); and mean urine color was > or = 4 (4-6 and 2-7). However, these men rarely (0-2% of measurements) achieved a urine specific gravity below 1.010 or color of 1. The first morning urine sample was more concentrated than the 24-h urine collection, likely because fluids were not consumed overnight. Furthermore, urine specific gravity and osmolality were strongly correlated (r2 = .81-.91, p < .001) in both morning and 24-hr collections. These findings provide euhydration reference values and hydration extremes for 7 commonly used indices in free-living, healthy, active men who were not exercising in a hot environment or training strenuously.
                Bookmark

                Author and article information

                Journal
                ACS Omega
                ACS Omega
                ao
                acsodf
                ACS Omega
                American Chemical Society
                2470-1343
                04 May 2021
                18 May 2021
                : 6
                : 19
                : 12684-12690
                Affiliations
                [1]Tobacco and Volatiles Branch, Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention , 4770 Buford Highway MS S103−3, Atlanta, Georgia 30341, United States
                Author notes
                [* ]Email address: mzz7@ 123456cdc.gov . Tel.: +17704880185. Fax: +17704880181.
                Article
                10.1021/acsomega.1c00854
                8154218
                34056420
                f57cb614-b85e-4e53-bef2-5f589ea973c6
                Not subject to U.S. Copyright. Published 2021 by American Chemical Society

                Permits non-commercial access and re-use, provided that author attribution and integrity are maintained; but does not permit creation of adaptations or other derivative works ( https://creativecommons.org/licenses/by-nc-nd/4.0/).

                History
                : 19 February 2021
                : 22 April 2021
                Categories
                Article
                Custom metadata
                ao1c00854
                ao1c00854

                Comments

                Comment on this article