      Effect of electronic apex locators on cardiac pacemaker function.

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The purpose of this study was to assess the effects of five electronic apex locators on pacemaker function in vitro. A Biotronik Actros DR+ pacemaker was evaluated at maximum sensitivity on a flat bench top. The pacemaker lead, electronic apex locator, and oscilloscope were connected across a 150-ohm resistor. Pace monitoring was carried out with a Biotronik EPR 1000 programmer and a Tektronix TDS 220 2-channel digital real-time oscilloscope. Four of five electronic apex locators tested did not cause inhibition or interfere with normal pacemaker function. It seems that electronic apex locators can be used safely in patients with pacemakers.