      Nintendo Wii remote controllers for head posture measurement: accuracy, validity, and reliability of the infrared optical head tracker.

      Investigative ophthalmology & visual science
      Adult, Diagnosis, Computer-Assisted, instrumentation, Female, Head, physiology, Humans, Infrared Rays, diagnostic use, Male, Movement, Play and Playthings, Posture, Reproducibility of Results, Video Games, Young Adult

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          To evaluate the accuracy, validity, and reliability of a newly developed infrared optical head tracker (IOHT) using Nintendo Wii remote controllers (WiiMote; Nintendo Co. Ltd., Kyoto, Japan) for measurement of the angle of head posture. The IOHT consists of two infrared (IR) receivers (WiiMote) that are fixed to a mechanical frame and connected to a monitoring computer via a Bluetooth communication channel and an IR beacon that consists of four IR light-emitting diodes (LEDs). With the use of the Cervical Range of Motion (CROM; Performance Attainment Associates, St. Paul, MN) as a reference, one- and three-dimensional (1- and 3-D) head postures of 20 normal adult subjects (20-37 years of age; 9 women and 11 men) were recorded with the IOHT. In comparison with the data from the CROM, the IOHT-derived results showed high consistency. The measurements of 1- and 3-D positions of the human head with the IOHT were very close to those of the CROM. The correlation coefficients of 1- and 3-D positions between the IOHT and the CROM were more than 0.99 and 0.96 (P < 0.05, Pearson's correlation test), respectively. Reliability tests of the IOHT for the normal adult subjects for 1- and 3-D positions of the human head had 95% limits of agreement angles of approximately ±4.5° and ±8.0°, respectively. The IOHT showed strong concordance with the CROM and relatively good test-retest reliability, thus proving its validity and reliability as a head-posture-measuring device. Considering its high performance, ease of use, and low cost, the IOHT has the potential to be widely used as a head-posture-measuring device in clinical practice.

          Related collections

          Author and article information

          Comments

          Comment on this article