      Permutation entropy in intraoperative ECoG of brain tumour patients in awake tumour surgery– a robust parameter to separate consciousness from unconsciousness

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Awake craniotomies represent an essential opportunity in the case of lesions in eloquent areas. Thus, optimal surveillance of the patient during different stages of sedation, as well as the detection of seizure activity during brain surgery, remains difficult, as skin electrodes for electroencephalographic (EEG) analysis are not applicable in most cases. We assessed the applicability of ECoG to monitor different stages of sedation, as well as the influence of different patient characteristics, such as tumour volume, size, entity, and age or gender on permutation entropy (PeEn). We conducted retrospective analysis of the ECoG data of 16 patients, who underwent awake craniotomies because of left-sided brain tumours at our centre between 2014 and 2016. PeEn could be easily calculated and compared using frontal and parietal cortical electrodes. A comparison of PeEn scores showed significantly higher values in awake patients than in patients under anaesthesia (p ≤ 0.004) and significantly higher ones in the state of transition than under general anaesthesia (p = 0.023). PeEn scores in frontal and parietal leads did not differ significantly, making them both applicable for continuous surveillance during brain surgery. None of the following clinical characteristics showed significant correlation with PeEn scores: tumour volume, WHO grade, first or recurrent tumour, gender, and sex. Being 50 years or older led to significantly lower values in parietal leads but not in frontal leads. ECoG and a consecutive analysis of PeEn are feasible and suitable for the continuous surveillance of patients during awake craniotomies. Hence, the analysis is not influenced by patients’ clinical characteristics.

          Related collections

          Most cited references32

          • Record: found
          • Abstract: found
          • Article: not found

          Permutation entropy: a natural complexity measure for time series.

          We introduce complexity parameters for time series based on comparison of neighboring values. The definition directly applies to arbitrary real-world data. For some well-known chaotic dynamical systems it is shown that our complexity behaves similar to Lyapunov exponents, and is particularly useful in the presence of dynamical or observational noise. The advantages of our method are its simplicity, extremely fast calculation, robustness, and invariance with respect to nonlinear monotonous transformations.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Bispectral index monitoring to prevent awareness during anaesthesia: the B-Aware randomised controlled trial.

            Awareness is an uncommon complication of anaesthesia, affecting 0.1-0.2% of all surgical patients. Bispectral index (BIS) monitoring measures the depth of anaesthesia and facilitates anaesthetic titration. In this trial we determined whether BIS-guided anaesthesia reduced the incidence of awareness during surgery in adults. We did a prospective, randomised, double-blind, multicentre trial. Adult patients at high risk of awareness were randomly allocated to BIS-guided anaesthesia or routine care. Patients were assessed by a blinded observer for awareness at 2-6 h, 24-36 h, and 30 days after surgery. An independent committee, blinded to group identity, assessed every report of awareness. The primary outcome measure was confirmed awareness under anaesthesia at any time. Of 2463 eligible and consenting patients, 1225 were assigned to the BIS group and 1238 to the routine care group. There were two reports of awareness in the BIS-guided group and 11 reports in the routine care group (p=0.022). BIS-guided anaesthesia reduced the risk of awareness by 82% (95% CI 17-98%). BIS-guided anaesthesia reduces the risk of awareness in at-risk adult surgical patients undergoing relaxant general anaesthesia. With a cost of routine BIS monitoring at US16 dollars per use in Australia and a number needed to treat of 138, the cost of preventing one case of awareness in high-risk patients is about 2200 dollars.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Permutation entropy of the electroencephalogram: a measure of anaesthetic drug effect.

              It would be useful to have an open-source electroencephalographic (EEG) index of gamma-amino-butyric acid (GABA)-ergic anaesthetic drug effect that is resistant to eye-blink artifact, responds rapidly to changes in EEG pattern, and can be linked to underlying neurophysiological and neuropharmacological mechanisms that control the conscious state. The EEG waveform can be described as a sequence of ordinal patterns. The permutation entropy (PE) describes the relative occurrence of each of these patterns. It is high ( approximately 1.0) when the signal has predominantly high frequencies and low ( approximately 0.4) when the signal consists of only low frequencies. The response of the PE to various computer-generated EEG-like waveforms was assessed. A composite PE index (CPEI) was developed, which was the sum of two simple PEs and included a small measurement-noise threshold (0.5 microV). We also applied the CPEI to two small pilot EEG data sets from patients receiving sevoflurane (n=21) or propofol (n=9) anaesthesia. With minimal pre-processing or artifact rejection, the CPEI reliably tracked the anaesthetic-related EEG changes, namely loss of high frequencies, spindle-like waves, and delta waves. Using NONMEM, it was possible to construct adequate pharmacokinetic-pharmacodynamic models from the data. The CPEI was comparable with models derived using the bispectral index [BIS R(2)=0.88 (0.08) vs CPEI R(2)=0.91 (0.06) for the propofol data] and M-entropy indices [M-entropy R(2)=0.91 (0.06) vs CPEI R(2)=0.87 (0.09) for the sevoflurane data]. PE of the EEG shows promise as a simple measure of GABAergic anaesthetic drug effect.
                Bookmark

                Author and article information

                Contributors
                Nicole.Lange@tum.de
                Journal
                Sci Rep
                Sci Rep
                Scientific Reports
                Nature Publishing Group UK (London )
                2045-2322
                11 November 2019
                11 November 2019
                2019
                : 9
                : 16482
                Affiliations
                [1 ]ISNI 0000 0004 0477 2438, GRID grid.15474.33, Department of Neurosurgery, , Klinikum rechts der Isar, Technical University Munich, ; Munich, Germany
                [2 ]ISNI 0000 0004 0477 2438, GRID grid.15474.33, Department of Neuroradiology, , Klinikum rechts der Isar, Technical University Munich, ; Munich, Germany
                [3 ]ISNI 0000 0004 0477 2438, GRID grid.15474.33, Department of Anesthesiology, , Klinikum rechts der Isar, Technical University Munich, ; Munich, Germany
                Article
                52949
                10.1038/s41598-019-52949-1
                6848144
                31712616
                e3c002ac-9fee-4e2b-a9be-6f5cf2025f5d
                © The Author(s) 2019

                Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.

                History
                : 12 March 2019
                : 25 October 2019
                Categories
                Article
                Custom metadata
                © The Author(s) 2019

                Uncategorized
                adaptive clinical trial,neurology
                Uncategorized
                adaptive clinical trial, neurology

                Comments

                Comment on this article