      Noninvasive measurement of cardiac stroke volume using pulse wave velocity and aortic dimensions: a simulation study

      research-article
      BioMedical Engineering OnLine
      BioMed Central

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Concerns about the cost-effectiveness of invasive hemodynamic monitoring in critically ill patients using pulmonary artery catheters motivate a renewed search for effective noninvasive methods to measure stroke volume. This paper explores a new approach based on noninvasively measured pulse wave velocity, pulse contour, and ultrasonically determined aortic cross sectional area.

          Methods

          The Bramwell-Hill equation relating pulse wave velocity to aortic compliance is applied. At the time point on the noninvasively measured pulse contour, denoted t h, when pulse amplitude has fallen midway between systolic and diastolic values, the portion of stroke volume remaining in the aorta, and in turn the entire stroke volume, can be estimated from the compliance and the pulse waveform. This approach is tested and refined using a numerical model of the systemic circulation including the effects of blood inertia, nonlinear compliance, aortic tapering, varying heart rate, and varying myocardial contractility, in which noninvasively estimated stroke volumes were compared with known stroke volumes in the model.

          Results

          The Bramwell-Hill approach correctly allows accurate calculation of known, constant aortic compliance in the numerical model. When nonlinear compliance is present the proposed noninvasive technique overestimates true aortic compliance when pulse pressure is large. However, a reasonable correction for nonlinearity can be derived and applied to restore accuracy for normal and for fast heart rates (correlation coefficient > 0.98).

          Conclusions

          Accurate estimates of cardiac stroke volume based on pulse wave velocity are theoretically possible and feasible. The precision of the method may be less than desired, owing to the dependence of the final result on the square of measured pulse wave velocity and the first power of ultrasonically measured aortic cross sectional area. However, classical formulas for propagation of random errors suggest that the method may still have sufficient precision for clinical applications. It remains as a challenge for experimentalists to explore further the potential of noninvasive measurement of stroke volume using pulse wave velocity. The technique is non-proprietary and open access in full detail, allowing future users to modify and refine the method as guided by practical experience.

          Related collections

          Most cited references31

          • Record: found
          • Abstract: found
          • Article: not found

          The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators.

          To examine the association between the use of right heart catheterization (RHC) during the first 24 hours of care in the intensive care unit (ICU) and subsequent survival, length of stay, intensity of care, and cost of care. Prospective cohort study. Five US teaching hospitals between 1989 and 1994. A total of 5735 critically ill adult patients receiving care in an ICU for 1 of 9 prespecified disease categories. Survival time, cost of care, intensity of care, and length of stay in the ICU and hospital, determined from the clinical record and from the National Death Index. A propensity score for RHC was constructed using multivariable logistic regression. Case-matching and multivariable regression modeling techniques were used to estimate the association of RHC with specific outcomes after adjusting for treatment selection using the propensity score. Sensitivity analysis was used to estimate the potential effect of an unidentified or missing covariate on the results. By case-matching analysis, patients with RHC had an increased 30-day mortality (odds ratio, 1.24; 95% confidence interval, 1.03-1.49). The mean cost (25th, 50th, 75th percentiles) per hospital stay was $49 300 ($17 000, $30 500, $56 600) with RHC and $35 700 ($11 300, $20 600, $39 200) without RHC. Mean length of stay in the ICU was 14.8 (5, 9, 17) days with RHC and 13.0 (4, 7, 14) days without RHC. These findings were all confirmed by multivariable modeling techniques. Subgroup analysis did not reveal any patient group or site for which RHC was associated with improved outcomes. Patients with higher baseline probability of surviving 2 months had the highest relative risk of death following RHC. Sensitivity analysis suggested that a missing covariate would have to increase the risk of death 6-fold and the risk of RHC 6-fold for a true beneficial effect of RHC to be misrepresented as harmful. In this observational study of critically ill patients, after adjustment for treatment selection bias, RHC was associated with increased mortality and increased utilization of resources. The cause of this apparent lack of benefit is unclear. The results of this analysis should be confirmed in other observational studies. These findings justify reconsideration of a randomized controlled trial of RHC and may guide patient selection for such a study.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            A Note on the Delta Method

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Comparison of aortic pulse wave velocity measured by three techniques: Complior, SphygmoCor and Arteriograph.

              New 2007 European Society of Hypertension guidelines recommend measuring arterial stiffness in patients with arterial hypertension, suggesting a carotid-femoral pulse wave velocity over 12 m/s as an estimate of subclinical organ damage. Considering this cutoff point, it is worth exploring whether or not there are significant differences in results obtained using various techniques for measuring aortic pulse wave velocity. The aim of the study was to compare aortic pulse wave velocity measurements using Complior, SphygmoCor, and Arteriograph devices, and to assess the effect of pulse wave transit time and traveled distance on pulse wave velocity values. Aortic pulse wave velocity was measured on a single visit, using these devices, in randomized order, in a group of 64 patients with grade 1 or 2 arterial hypertension. Aortic pulse wave velocity measured using Complior (10.1 +/- 1.7 m/s) was significantly higher than that obtained using SphygmoCor (8.1 +/- 1.1 m/s) or Arteriograph (8.6 +/- 1.3 m/s). No differences were noted between pulse wave velocity measurements using SphygmoCor and Arteriograph. Between-method comparison revealed that differences in traveled distance were significant: Complior versus Arteriograph [0.09 m, Confidence interval (CI): 0.08-0.12 m, P < 0.05], Complior versus SphygmoCor (0.15 m, CI: 0.13-0.16 m, P < 0.05), Arteriograph versus SphygmoCor (0.05 m, CI: 0.03-0.07 m, P < 0.05). No between-method differences were found for transit times. Differences in pulse wave velocity obtained by compared devices resulted primarily from using various methods for measuring traveled distance. It appears reasonable to establish uniform principles for the measurement of traveled distance. Because a large number of prognosis/survival studies used direct distance between carotid and femoral sites of pulse wave recording, this distance should be mostly recommended.
                Bookmark

                Author and article information

                Contributors
                babbs@purdue.edu
                Journal
                Biomed Eng Online
                Biomed Eng Online
                BioMedical Engineering OnLine
                BioMed Central (London )
                1475-925X
                19 September 2014
                19 September 2014
                2014
                : 13
                : 1
                : 137
                Affiliations
                Weldon School of Biomedical Engineering, Purdue University, 206 South Martin Jische Drive, West Lafayette, Indiana 47907-2032 USA
                Article
                895
                10.1186/1475-925X-13-137
                4271357
                25238910
                e38f6d61-8dae-42d1-9ff5-9e116c0257db
                © Babbs; licensee BioMed Central Ltd. 2014

                This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

                History
                : 9 June 2014
                : 16 September 2014
                Categories
                Research
                Custom metadata
                © The Author(s) 2014

                Biomedical engineering
                Biomedical engineering

                Comments

                Comment on this article