388
views
0
recommends
+1 Recommend
1 collections
    1
    shares

      2023 Journal Citation Reports Journal Impact Factor is 0.9. Scopus Citescore 0.8. 

      Interested in becoming a CVIA published author?

      • Platinum Open Access with no APCs. 
      • Fast peer review/Fast publication online after article acceptance.

      Submissions should be made electronically at: https://mc04.manuscriptcentral.com/cvia-journal.

      Please refer to the Author Guidelines at https://cvia-journal.org/instructions-to-authors/ before submission.

       

      scite_
       
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Valve-in-Valve Transcatheter Aortic Valve Replacement in a High-Risk Patient with a Biocor Bioprosthesis and a Flail Prosthetic Valve Leaflet

      Published
      case-report
      Bookmark

            Abstract

            An 80-year-old woman with a history of surgical aortic valve replacement with a 21 mm St. Jude Medical Biocor porcine aortic valve 14 years prior presented with New York Heart Association (NYHA) class III symptoms, severe aortic insufficiency from a degenerated prosthesis, and a large echocardiographic mobile mass representing a highly mobile prosthetic leaflet. The patient worsened to NYHA class IV symptoms despite medical management. The Society of Thoracic Surgery mortality risk score was extremely high. However, a valve-in-valve transcatheter aortic valve replacement (TAVR) was found to be a reasonable option. We used a 20 mm SAPIEN 3 Ultra valve (Edwards Lifesciences Inc., Irvine, CA, USA) with a SENTINEL embolic protection device (Boston Scientific, Marlborough, MA, USA). During valve deployment, the echocardiographic mobile mass was visually pinned between the new TAVR valve and the surgical bioprosthetic valve. No large embolic debris was noted within the embolic protection device, and the patient remained without any new focal neurologic deficits in the perioperative period and at the 30-day follow-up. The severe aortic insufficiency resolved, and the patient clinically improved to NYHA class II symptoms.

            Main article text

            Introduction

            Prosthetic aortic valves have differing durability and longevity. Bioprosthetic valve (BPV) life is finite, averaging 15 years in the elderly population [1]. Several factors of valve degeneration are not well understood; however, calcification appears to contribute to degeneration and leaflet tears [24]. Within the last 10 years, the number of aortic valve replacement (AVR) procedures has doubled [5], and as the quantity of AVR procedures increases, clinicians will inevitably face increased prosthetic valve failures. Herein, we present a successful case of relatively urgent valve-in-valve (ViV) transcatheter AVR (TAVR) with embolic protection in a high-risk patient with BPV failure and a large echocardiographic mobile mass.

            Case Report

            An 80-year-old woman with a 21 mm St. Jude Medical Biocor porcine aortic valve replacement in 2006 presented to her primary cardiologist with new worsening dyspnea on mild exertion. The patient’s medical history was significant for type 2 diabetes mellitus, hypertension, hyperlipidemia, remote ischemic stroke, chronic kidney disease, anemia, chronic obstructive pulmonary disease, and atrial fibrillation. Thoracic echocardiography at that time revealed a large (2 cm×0.75 cm) extremely mobile mass within the BPV associated with severe aortic insufficiency (AI), moderate mitral insufficiency, and a preserved left ventricular ejection fraction. Coronary angiography demonstrated mild nonobstructive coronary artery disease. The patient was treated empirically with broad-spectrum antimicrobial therapy for possible infective endocarditis, despite a lack of major Duke criteria or any stigmata of infective endocarditis.

            The patient was referred to the structural heart team at our institution, and transesophageal echocardiography was performed. Transesophageal echocardiography revealed that the mobile mass most likely represented a prosthetic flail leaflet without vegetation (see supplementary video files 1 and 2). The patient was evaluated by the palliative care team for hospice evaluation given her other comorbidities. However, she insisted on an aggressive plan to treat her worsening New York Heart Association (NYHA) class IV symptoms and associated paroxysmal nocturnal dyspnea despite increasing doses of diuretics.

            With the patient having an estimated Society of Thoracic Surgery (STS) mortality risk score greater than 30%, a ViV TAVR was thought to be the best option. We suspected that a ViV TAVR would result in patient prosthesis mismatch (PPM) given the patient’s significant risk factors: ViV TAVR, female sex, and prosthesis smaller than 23 mm [6]. However, the estimated surgical mortality appeared greater than the risk of PPM with ViV TAVR. The patient and her family elected to proceed with TAVR as her comorbidities did not make her risk of death within 1 year prohibitive for the procedure.

            The patient underwent ViV TAVR with a 20 mm SAPIEN 3 Ultra valve (Edwards Lifesciences Inc., Irvine, CA, USA). To reduce stroke risk, we used a SENTINEL cerebral embolic protection (Boston Scientific, Marlborough, MA, USA), a filterlike device. During valve deployment, the echocardiographic mobile mass was visually pinned between the new TAVR valve and the surgical BPV (see supplementary video files 3–5). No large embolism was filtered from the SENTINEL device (Figure 1). BPV fracture before TAVR valve placement has been shown to improve the valve’s hemodynamics and decrease PPM in patients, although risks include coronary obstruction and ischemic stroke [7, 8]. We suspected that BPV fracture could result in increased embolic risk of the flail leaflet and therefore it was not performed.

            Figure 1

            Embolic Protection Device (Boston Scientific).

            Small amount of debris filtered out with the embolic protection device in place during the TAVR.

            The patient remained without new focal neurologic deficits in the perioperative period and at the 30-day follow-up. Postoperative echocardiography demonstrated resolved AI with stable mitral insufficiency. The aortic valve mean gradient was elevated as expected at 19 mmHg, indicative of at least moderate PPM. The patient’s symptoms changed from NYHA class IV to NYHA class II, and the patient was discharged home.

            Discussion

            ViV TAVR for degenerative prosthetic valves is a reasonable option for many patients, especially when there is a high risk of repeated surgical AVR (SAVR). ViV TAVR has been performed with success in other cases with mobile masses or flail leaflets, although in one case the mass did embolize and was retrieved [9, 10]. In our patient’s case, ViV TAVR was a life-saving measure that improved her quality of life. Early detection of valve degeneration is preferred for optimal patient outcomes, as emergency repeated AVR is associated with mortality of 22.6% compared with 1.4% for elective repeated AVR [11]. Unfortunately, degenerative prosthetic valves are often discovered late, when patients have high operative risk.

            The goal of prosthetic valve surveillance should be to detect significant valvular degeneration early so that the patient can undergo lower-risk procedures. Prosthetic valve surveillance and evaluation is performed with an echocardiogram, which is indicated for any change in symptoms in an AVR patient. A routine annual echocardiogram is recommended at 5 and 10 years after the surgical implantation, even if the patient is asymptomatic [12]. With an inconclusive echocardiogram in a symptomatic patient, further workup is indicated with other imaging modalities, such as cinefluoroscopy, cardiac computed tomography, or cardiac magnetic resonance imaging [13].

            When a BPV fails, AVR options are a redo SAVR procedure or a ViV TAVR [14]. Individualized patient care is necessary when one is weighing the risks and benefits of redo SAVR versus ViV TAVR. The higher operative morbidity and mortality associated with redo SAVR must be weighed against the risk of significant PPM, permanent pacemaker placement, and coronary obstruction with ViV TAVR [6, 15].

            Newer studies suggest that ViV TAVR is associated with decreased risk of stroke, myocardial infarction, and major bleeding, as well as reduction in all-cause mortality during the perioperative period and no significant difference in all-cause mortality at 2 years [16]. However, ViV TAVR is more likely to cause PPM because the new valve is deployed within the ring of the existing BPV, which leaves less space for the new TAVR valve orifice area. Ultimately both perioperative mortality and overall mortality increase proportionally to PPM severity [17]. Therefore, it is important to understand the predictors of severe PPM when one is considering ViV TAVR versus redo SAVR. Overall predictors of PPM are a valve prosthesis diameter of 23 mm or less, female sex, a ViV procedure, young age, severe mitral or tricuspid insufficiency, larger body surface area, lower ejection fraction, and atrial fibrillation [6].

            In addition to PPM, each patient’s stroke risk should be individually assessed. ViV TAVR stroke risk has not been fully established, but recent studies do not show any increased stroke risk in comparison with native valve TAVR [18]. Use of the SENTINEL embolic protection device in TAVR procedures has been reported to significantly lower the risk of stroke, death, and major bleeding at 30 days [19]. In this case of a clearly elevated risk of stroke, a ViV TAVR with embolic protection was successfully performed without new focal neurologic deficits in the perioperative period or at the 30-day follow-up.

            Ethics Statement

            The study was approved by the Ethics Committee of AdventHealth Tampa Pepin Heart Institute.

            Conflict of Interest

            The authors have no related conflicts of interest.

            Citation Information

            References

            1. , , , , , . Carpentier–Edwards supraannular porcine bioprosthesis: second-generation prosthesis in aortic valve replacement. Ann Thorac Surg 2001;71(5):S224–7.

            2. , , , , . Aortic bioprosthetic valve durability: incidence, mechanisms, predictors, and management of surgical and transcatheter valve degeneration. J Am Coll Cardiol 2017;70(8):1013–28.

            3. , , . Bioprosthetic heart valves: impact of implantation on biomaterials. Int Sch Res Notices 2013;2013:728791.

            4. , . Prosthetic heart valves: selection of the optimal prosthesis and long-term management. Circulation 2009;119(7):1034–48.

            5. , , , , , , et al. Changing trends in aortic valve procedures over the past ten years – from mechanical prosthesis via stented bioprosthesis to TAVI procedures – analysis of 50,846 aortic valve cases based on a Polish national cardiac surgery database. J Thorac Dis 2019; 11(6):2340.

            6. , , , , , , et al. Prosthesis–patient mismatch in patients undergoing transcatheter aortic valve replacement: from the STS/ACC TVT registry. J Am Coll Cardiol 2018;72(22):2701–11.

            7. , , , , , , et al. Bioprosthetic valve fracture improves the hemodynamic results of valve-in-valve transcatheter aortic valve replacement. Circ Cardiovasc Interv 2017;10(7):e005216.

            8. , , , , , , et al. Complications of bioprosthetic valve fracture as an adjunct to valve-in-valve TAVR. Struct Heart 2019;3(2):92–9.

            9. , , , , . High-risk trans-catheter aortic valve replacement in a failed freestyle valve with low coronary height: a case report. Cardiol Ther 2017;6(1):145–50.

            10. , , , . Embolization of aortic valve leaflet during valve-in-valve transcatheter aortic valve implantation: a case report. Eur Heart J Case Rep 2020;4(1):1–5.

            11. , , , , , , et al. Reoperative surgery for degenerated aortic bioprostheses: predictors for emergency surgery and reoperative mortality. Eur J Cardiothorac Surg 2000;17(2):134–9.

            12. Writing Committee Members, , , , , , et al. 2020 ACC/AHA guideline for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. J Am Coll Cardiol 2021;77(4):e25–197. doi:[Cross Ref]. Erratum in: J Am Coll Cardiol 2021;77(4):509. Erratum in: J Am Coll Cardiol 2021;77(9):1275.

            13. , , , , , , et al. Recommendations for the imaging assessment of prosthetic heart valves: a report from the European Association of Cardiovascular Imaging endorsed by the Chinese Society of Echocardiography, the Inter-American Society of Echocardiography, and the Brazilian Department of Cardiovascular Imaging. Eur Heart J Cardiovasc Imaging 2016;17(6):589–90.

            14. , , , , . Modes of bioprosthetic valve failure: a narrative review. Curr Opin Cardiol 2020;35(2):123–32.

            15. , , , , , , et al. The Society of Thoracic Surgeons 2008 cardiac surgery risk models: part 2 – isolated valve surgery. Ann Thorac Surg 2009;88(1):S23–42.

            16. , , , , , , et al. Transcatheter valve-in-valve aortic valve replacement as an alternative to surgical re-replacement. J Am Coll Cardiol 2020;76(5):489–99.

            17. , , , , , . Predictors and outcomes of prosthesis-patient mismatch after aortic valve replacement. JACC Cardiovasc Imaging 2016;9(8):924–33.

            18. , , , , . Does valve in valve TAVR carry a higher risk for thromboembolic events compared to native valve TAVR?. Catheter Cardiovasc Interv 2020;95(5):1017–21.

            19. , , , . Clinical outcomes of sentinel cerebral protection system use during transcatheter aortic valve replacement: a systematic review and meta-analysis. Cardiovasc Revasc Med 2020;21(6):717–22.

            Supplementary Material:

            This paper offers supplementary material which can be found at the following link: https://cvia-journal.org/supplementary-figures-2/

            Author and article information

            Journal
            CVIA
            Cardiovascular Innovations and Applications
            CVIA
            Compuscript (Ireland )
            2009-8782
            2009-8618
            May 2022
            May 2022
            : 6
            : 3
            : 191-194
            Affiliations
            [1] 1AdventHealth Tampa Pepin Heart Institute, Cardiology, 3100 East Fletcher Avenue, Tampa, FL 33613, USA
            [2] 2DeBusk College of Osteopathic Medicine, Lincoln Memorial University, Cardiology, 6965 Cumberland Gap Parkway, Harrogate, TN 37752, USA
            [3] 3AdventHealth Tampa Pepin Heart Institute, Cardiothoracic Surgery, 3100 East Fletcher Avenue, Tampa, FL 33613, USA
            Author notes
            Correspondence: Dr. Oliver Abela, AdventHealth Tampa Pepin Heart Institute, Cardiology, 3100 East Fletcher Avenue, Tampa, FL 33613, USA, E-mail: abelaoli@ 123456gmail.com
            Article
            cvia.2021.0031
            10.15212/CVIA.2021.0031
            e297d599-f26d-4e6f-b366-da7bb63e1ebc
            Copyright © 2022 Cardiovascular Innovations and Applications

            This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 Unported License (CC BY-NC 4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. See https://creativecommons.org/licenses/by-nc/4.0/.

            History
            : 22 July 2021
            : 10 October 2021
            : 21 October 2021
            Categories
            Case Report

            General medicine,Medicine,Geriatric medicine,Transplantation,Cardiovascular Medicine,Anesthesiology & Pain management
            TAVR,Bioprosthetic,Regurgitation

            Comments

            Comment on this article