      Slice-thickness artifacts in gray-scale ultrasound.

      Journal of clinical ultrasound : JCU
      Diagnostic Errors, Humans, Ultrasonography

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          We have become increasingly aware of the presence of a type of image artifact normally appearing in anechoic areas (eg, cyst, bladder, gallbladder) and giving the appearance of "sludge" or "debris." These artifactual echoes may be caused by the fact that the finite width of the transducer beam pattern produces a finite thickness of the patient scan plane. All echoes produced in this "thick" scan plane are misinterpreted as being due to structures in the normally assumed "thin" scan plane. We have tested and verified this hypothesis by simulating soft tissue interfaces with 400 grit silicon carbide sandpaper in a water tank. A set of clues are proposed to enable the rapid identification of these artifacts.