      Diagnostic Value of Contrast-Enhanced Spectral Mammography in Comparison to Magnetic Resonance Imaging in Breast Lesions

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Objective

          The aim of this study was to evaluate the diagnostic value between contrast-enhanced spectral mammography (CESM) and breast magnetic resonance imaging (MRI) in breast disease.

          Methods

          Two hundred thirty-five patients who were suspected of having breast abnormalities by clinical examination or mammography underwent CESM and MRI examination. Using histopathologic results as the criterion standard, the diagnostic performance of CESM and MRI was investigated. The areas under receiver operating characteristic curves were applied to analyze diagnostic efficiency. The Pearson correlation coefficients between CESM versus pathology and MRI versus pathology were calculated.

          Results

          Two hundred sixty-three breast lesions were found in 235 patients, in which 177 were malignant and 86 were benign. By evaluating the diagnostic value, sensitivity, positive predictive value, negative predictive value, and false-negative rate from CESM examination were comparable to those from MRI (91.5%, 94.7%, 83.7%, and 8.5% vs 91.5%, 90.5%, 82.1%, and 8.5%). Importantly, the accuracy and the specificity were higher for CESM than those for MRI (81% and 89.5% vs 80.2% and 71.7%), whereas the false-positive rate was lower (10.5% vs 19.8%). The areas under receiver operating characteristic curves of CESM and MRI were 0.950 and 0.939, displaying the equivalent diagnostic efficiency ( P = 0.48).

          For the agreement between measurements, mean tumor sizes were 3.1 cm for CESM and 3.4 cm for MRI compared with 3.2 cm on histopathologic results. The Pearson correlation coefficient of CESM versus histopathology ( r = 0.774, P = 0.000) was consistent with MRI versus histopathology ( r = 0.771, P = 0.000).

          Conclusions

          Our results show better accuracy, specificity, and false-positive rate of CESM in breast cancer detection than MRI. Contrast-enhanced spectral mammography displayed a good correlation with histopathology in assessing the lesion size of breast cancer, which is consistent with MRI.

          Related collections

          Most cited references12

          • Record: found
          • Abstract: found
          • Article: found
          Is Open Access

          Early Prediction and Evaluation of Breast Cancer Response to Neoadjuvant Chemotherapy Using Quantitative DCE-MRI1

          The purpose is to compare quantitative dynamic contrast-enhanced (DCE) magnetic resonance imaging (MRI) metrics with imaging tumor size for early prediction of breast cancer response to neoadjuvant chemotherapy (NACT) and evaluation of residual cancer burden (RCB). Twenty-eight patients with 29 primary breast tumors underwent DCE-MRI exams before, after one cycle of, at midpoint of, and after NACT. MRI tumor size in the longest diameter (LD) was measured according to the RECIST (Response Evaluation Criteria In Solid Tumors) guidelines. Pharmacokinetic analyses of DCE-MRI data were performed with the standard Tofts and Shutter-Speed models (TM and SSM). After one NACT cycle the percent changes of DCE-MRI parameters K trans (contrast agent plasma/interstitium transfer rate constant), v e (extravascular and extracellular volume fraction), k ep (intravasation rate constant), and SSM-unique τ i (mean intracellular water lifetime) are good to excellent early predictors of pathologic complete response (pCR) vs. non-pCR, with univariate logistic regression C statistics value in the range of 0.804 to 0.967. v e values after one cycle and at NACT midpoint are also good predictors of response, with C ranging 0.845 to 0.897. However, RECIST LD changes are poor predictors with C = 0.609 and 0.673, respectively. Post-NACT K trans, τ i, and RECIST LD show statistically significant (P < .05) correlations with RCB. The performances of TM and SSM analyses for early prediction of response and RCB evaluation are comparable. In conclusion, quantitative DCE-MRI parameters are superior to imaging tumor size for early prediction of therapy response. Both TM and SSM analyses are effective for therapy response evaluation. However, the τ i parameter derived only with SSM analysis allows the unique opportunity to potentially quantify therapy-induced changes in tumor energetic metabolism.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Contrast-enhanced spectral mammography in patients referred from the breast cancer screening programme.

            Feasibility studies have shown that contrast-enhanced spectral mammography (CESM) increases diagnostic accuracy of mammography. We studied diagnostic accuracy of CESM in patients referred from the breast cancer screening programme, who have a lower disease prevalence than previously published papers on CESM.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Comparison between Breast MRI and Contrast-Enhanced Spectral Mammography

              Background The main goal of this study was to compare contrast-enhanced spectral mammography (CESM) and breast magnetic resonance imaging (MRI) with histopathological results and to compare the sensitivity, accuracy, and positive and negative predictive values for both imaging modalities. Material/Methods After ethics approval, CESM and MRI examinations were performed in 102 patients who had suspicious lesions described in conventional mammography. All visible lesions were evaluated independently by 2 experienced radiologists using BI-RADS classifications (scale 1–5). Dimensions of lesions measured with each modality were compared to postoperative histopathology results. Results There were 102 patients entered into CESM/MRI studies and 118 lesions were identified by the combination of CESM and breast MRI. Histopathology confirmed that 81 of 118 lesions were malignant and 37 were benign. Of the 81 malignant lesions, 72 were invasive cancers and 9 were in situ cancers. Sensitivity was 100% with CESM and 93% with breast MRI. Accuracy was 79% with CESM and 73% with breast MRI. ROC curve areas based on BI-RADS were 0.83 for CESM and 0.84 for breast MRI. Lesion size estimates on CESM and breast MRI were similar, both slightly larger than those from histopathology. Conclusions Our results indicate that CESM has the potential to be a valuable diagnostic method that enables accurate detection of malignant breast lesions, has high negative predictive value, and a false-positive rate similar to that of breast MRI.
                Bookmark

                Author and article information

                Journal
                J Comput Assist Tomogr
                J Comput Assist Tomogr
                RCT
                Journal of Computer Assisted Tomography
                Lippincott Williams & Wilkins
                0363-8715
                1532-3145
                Mar-Apr 2019
                07 December 2018
                : 43
                : 2
                : 245-251
                Affiliations
                From the [* ]Department of Radiology and
                []Interventional Therapy Ward, Yantai Yuhuangding Hospital; and
                []The Affiliated Yantai Yuhuangding Hospital of Qingdao University, Yantai, China.
                Author notes
                [*]Correspondence to: Xiaoxiao Chi, Department of Radiology, Yantai Yuhuangding Hospital, No. 20, Yuhuangding East Road, Yantai, Shandong Province 264000 China(e-mail: chixiaoxiao87@ 123456126.com ).
                Article
                RCT50594 00013
                10.1097/RCT.0000000000000832
                6426358
                30531546
                c678785e-4f0c-46ca-9100-8352f0e5575a
                Copyright © 2018 The Author(s). Published by Wolters Kluwer Health, Inc.

                This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND), where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal.

                History
                : 30 May 2018
                : 9 October 2018
                Page count
                Pages: 0
                Categories
                Breast Imaging
                Custom metadata
                TRUE

                breast cancer,contrast agent,contrast-enhanced spectral mammography,magnetic resonance imaging