      Avaliação das condições de funcionamento dos equipamentos de raios X médico em serviços de radiologia no Estado da Paraíba, durante os anos de 2008 e 2009 Translated title: Evaluation of medical X-ray machines in Paraíba state radiology centers between 2008 and 2009

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          OBJETIVO: Avaliar a coincidência entre o campo luminoso e o campo de radiação, e o alinhamento do feixe de radiação dos equipamentos de raios X médico no Estado da Paraíba, por meio de dois testes de controle de qualidade. MATERIAIS E MÉTODOS: Foram utilizados os dispositivos para os testes de tamanho de campo e de alinhamento, um chassi carregado, trena e nível de bolha. Os testes de exatidão do sistema de colimação e de alinhamento do raio central do feixe de raios X foram realizados durante as inspeções sanitárias em serviços de radiologia nos anos de 2008 e 2009. RESULTADOS: No ano de 2008, 121 equipamentos de raios X foram testados no Estado da Paraíba. No ano de 2009, passaram pelos testes 117 equipamentos. Deste universo, 86 foram selecionados para a comparação por terem sido avaliados tanto no ano de 2008 como em 2009, sendo observada uma melhoria de 18,60% (n = 16) nos resultados dos testes realizados de um ano para outro. CONCLUSÃO Conclusão: Pode-se concluir que o percentual de equipamentos apresentando problemas no seu desempenho sofreu uma diminuição entre 2008 e 2009, não sendo observado programa de garantia de qualidade em nenhum dos serviços de radiologia do Estado da Paraíba.

          Translated abstract

          OBJECTIVE: To evaluate the light and radiation fields congruence and radiation beam alignment in medical X-ray equipment in Paraíba state radiology centers by means of two quality control tests. MATERIALS AND METHODS: A loaded cassette, a measuring tape and a bubble level were utilized in the field size and alignment testing. The evaluation of collimation systems accuracy and X-ray beam alignment was undertaken during health inspections performed in radiology centers between 2008 and 2009. RESULTS: In 2008, 121 X-ray machines were evaluated in the Paraíba state. In 2009, 117 machines were tested. From this universe, 86 machines were selected for comparison, since they were evaluated both in 2008 and 2009, with 18.60% (n = 16) showing test results improvement from one year to another. CONCLUSION: The percentage of problematic X-ray machines decreased between 2008 and 2009, notwithstanding no quality assurance program has been observed in Paraíba state radiology centers.

          Related collections

          Most cited references16

          • Record: found
          • Abstract: not found
          • Book: not found

          Epidemiologia: teoria e prática

            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Análise dos resultados do programa de gerenciamento da qualidade dos equipamentos radiológicos

              Bookmark
              • Record: found
              • Abstract: not found
              • Book: not found

              Radiodiagnóstico médico: desempenho de equipamentos e segurança

              (2005)
                Bookmark

                Author and article information

                Journal
                rb
                Radiologia Brasileira
                Radiol Bras
                Publicação do Colégio Brasileiro de Radiologia e Diagnóstico por Imagem (São Paulo, SP, Brazil )
                1678-7099
                August 2011
                : 44
                : 4
                : 244-248
                Affiliations
                [01] João Pessoa PB orgnameAgência Estadual de Vigilância Sanitária da Paraíba Brasil
                Article
                S0100-39842011000400010 S0100-3984(11)04400410
                bc49ce9e-0156-43e8-b6fb-2f59901cb2aa

                This work is licensed under a Creative Commons Attribution 4.0 International License.

                History
                : 20 October 2010
                : 29 June 2011
                Page count
                Figures: 0, Tables: 0, Equations: 0, References: 16, Pages: 5
                Product

                SciELO Brazil

                Categories
                Artigos Originais

                Quality control,Equipamento,Raios X,Controle de qualidade,Equipment,X-rays

                Comments

                Comment on this article