      A comparison of trans-tibial amputee suction and vacuum socket conditions.

      Prosthetics and Orthotics International
      Adult, Amputation Stumps, Amputees, rehabilitation, Artificial Limbs, Biomechanical Phenomena, Body Mass Index, Female, Gait, physiology, Humans, Male, Middle Aged, Prosthesis Design, Prosthesis Fitting, Sampling Studies, Sensitivity and Specificity, Stress, Mechanical, Tibia, surgery, Vacuum

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Daily volume loss of the stump leads to a poor fit of the prosthetic socket. A method of preventing this volume loss and maintaining a good fit was developed. A vacuum (-78 kPa) was drawn on the expulsion port of a total surface-bearing suction socket to hold the liner tightly against the socket. Stump volume of 10 trans-tibial amputees was measured prior to and immediately after a 30 minute walk with normal and vacuum socket conditions. Under the normal condition, the limb lost an average of 6.5% of its volume during the walk. In contrast, with the liner held tightly by vacuum, the limb gained an average of 3.7% in volume. It is believed that the difference observed between conditions resulted from a greater negative pressure developed during the swing phase of gait with the vacuum condition. X-rays revealed that the limb and tibia pistoned 4 mm and 7 mm less, respectively, under the vacuum condition. The combination of reduced pistoning and maintenance of volume is thought to account for the more symmetrical gait observed with the vacuum.

          Related collections

          Author and article information

          Comments

          Comment on this article