      Use of the Impella microaxial blood pump for ablation of hemodynamically unstable ventricular tachycardia.

      Journal of Cardiovascular Electrophysiology
      Adult, Aged, Catheter Ablation, instrumentation, Equipment Design, Equipment Failure Analysis, Female, Heart-Assist Devices, Humans, Hypotension, etiology, prevention & control, Male, Middle Aged, Miniaturization, Tachycardia, Ventricular, complications, surgery, Treatment Outcome

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Ablation for ventricular tachycardia remains a challenge with suboptimal procedural success rates. One of the major causes of difficulty is precipitous hypotension when ventricular tachycardia is induced precluding even rapid mapping of the arrhythmia. We report the successful use of the Impella microcirculatory axial blood flow pump in 3 patients with hemodynamically unstable ventricular tachycardia that allowed successful completion of the procedure. In these 3 patients, there was no evidence of Impella-related valvular disturbance, iatrogenic ventricular arrhythmias, or interference with mapping and ablation catheter movement.