The aim of this study was to validate a method for the determination of acethaldehyde, methanol, ethanol, acetone and isopropanol employing solid-phase microextraction associated to gas chromatography with flame ionization detection. The operational conditions of SPME were optimized by response surface analysis. The calibration curves for all compounds were linear with r² > 0.9973. Accuracy (89.1-109.0%), intra-assay precision (1.8-8.5%) and inter-assay precision (2.2-8.2%) were acceptable. The quantification limit was 50 µg/mL. The method was applied to the meaurement of ethanol in blood and oral fluid of a group of volunteers. Oral fluid ethanol concentrations were not directly correlated with blood concentrations.