8
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: not found

      Initial Experience and Feasibility of the New Low-Profile Stingray Catheter as Part of the Antegrade Dissection and Re-Entry Revascularization Strategy for Coronary Chronic Total Occlusions

      research-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Case series

          Patient: Male, 77 • Male, 57 • Male, 73

          Final Diagnosis: Coronary chronic total occlusion

          Symptoms: Angina pectoris

          Medication: —

          Clinical Procedure: Percutaneous coronary intervention of coronary chronic total occlusions

          Specialty: Cardiology

          Objective:

          Unusual setting of medical care

          Background:

          During antegrade dissection re-entry (ADR) of chronic total occlusions (CTO), the first-generation Stingray catheter requires the use of large-bore guides (sheathless 7.5 Fr or 8 Fr), which increases the risk for access site-related complications and compromises radial approaches. Smaller guiding sizes necessitate long guidewires (e.g., 300 cm) or guidewire extensions for catheter advancement or removal. However, friction between guides and the Stingray catheter can result in unstable guidewire position or unintentional removal. Furthermore, failure to deliver the catheter at the distal re-entry zone is a common problem. To overcome issues of deliverability and reduce the need for pre-dilatations, with its inherent risk of creating subintimal hematomas, the Stingray low-profile (LP) balloon catheter was developed.

          Case Report:

          We describe 3 cases of successful application of the novel Stingray LP catheter during ADR. In all cases, 7 Fr guiding catheters were successfully used in combination with the device. The lower profile facilitated a good exchange and delivery of the device, without the need for balloon pre-dilatations in 2 cases. This resulted in a limited subintimal plane, enabling a smooth puncture into the true lumen. One case presented with extreme levels of calcification and tortuosity, resulting in a high degree of friction, despite the lower catheter profile. No in-hospital coronary or access site-related complications occurred.

          Conclusions:

          This case report illustrates the feasibility of the Stingray LP catheter for the treatment of CTOs via the ADR technique. The lower profile of the catheter potentially increases the deliverability, safety, and exchangeability of the device.

          Related collections

          Most cited references6

          • Record: found
          • Abstract: found
          • Article: not found

          A percutaneous treatment algorithm for crossing coronary chronic total occlusions.

          Coronary chronic total occlusions (CTOs) are frequently identified during coronary angiography and remain the most challenging lesion group to treat. Patients with CTOs are frequently left unrevascularized due to perceptions of high failure rates and technical complexity even if they have symptoms of coronary disease or ischemia. In this review, the authors describe a North American contemporary approach for percutaneous coronary interventions for CTO. Two guide catheters are placed to facilitate seamless transition between antegrade wire-based, antegrade dissection re-entry-based, and retrograde (wire or dissection re-entry) techniques, the "hybrid" interventional strategy. After dual coronary injection is performed, 4 angiographic parameters are assessed: 1) clear understanding of location of the proximal cap using angiography or intravascular ultrasonography; 2) lesion length; 3) presence of branches, as well as size and quality of the target vessel at the distal cap; and 4) suitability of collaterals for retrograde techniques. On the basis of these 4 characteristics, an initial strategy and rank order hierarchy for technical approaches is established. Radiation exposure, contrast utilization, and procedure time are monitored throughout the procedure, and thresholds are established for intraprocedural strategy conversion to maximize safety, efficiency, and effectiveness.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Use of a novel crossing and re-entry system in coronary chronic total occlusions that have failed standard crossing techniques: results of the FAST-CTOs (Facilitated Antegrade Steering Technique in Chronic Total Occlusions) trial.

            This study sought to examine the efficacy and safety of 3 novel devices to recanalize coronary chronic total occlusions (CTOs).
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Subintimal dissection/reentry strategies in coronary chronic total occlusion interventions.

              Subintimal dissection/reentry techniques are increasingly being used for crossing coronary chronic total occlusions both antegradely (using a knucle wire or the Bridgepoint system) or retrogradely (using the controlled antegrade and retrograde tracking and dissection, and reverse controlled antegrade and retrograde tracking and dissection technique). Subintimal dissection/reentry techniques can increase procedural success rates, but their subsequent clinical outcomes are poorly studied, and they appear to be associated with high rates of in-stent restenosis and repeat target lesion revascularization. In the present review, we describe in detail the chronic total occlusions subintimal dissection/reentry techniques, clarify the related terminology and summarize the published studies in this area and the current gaps of knowledge.
                Bookmark

                Author and article information

                Journal
                Am J Case Rep
                Am J Case Rep
                amjcaserep
                The American Journal of Case Reports
                International Scientific Literature, Inc.
                1941-5923
                2017
                31 January 2017
                : 18
                : 104-109
                Affiliations
                [1 ]Faculty of Medicine and Life Sciences, Universiteit Hasselt, Hasselt, Belgium
                [2 ]Department of Cardiology, Ziekenhuis Oost-Limburg, Genk, Belgium
                Author notes

                Authors’ Contribution:

                [A]

                Study Design

                [B]

                Data Collection

                [C]

                Statistical Analysis

                [D]

                Data Interpretation

                [E]

                Manuscript Preparation

                [F]

                Literature Search

                [G]

                Funds Collection

                Conflict of interest: Jo Dens receives grants from TopMedical (Distributor of Asahi Intecc Co. materials), Boston Scientific, Vascular Solutions, and Orbus Neich for teaching courses and proctoring. Joren Maeremans is researcher for the Limburg Clinical Research Program UHasselt-ZOL-Jessa, supported by the foundation Limburg Sterk Merk, Hasselt University, Ziekenhuis Oost-Limburg and Jessa Hospital. For the remaining author, none were declared.

                Corresponding Author: Jo Dens, e-mail: jo.dens@ 123456zol.be
                Article
                902178
                10.12659/AJCR.902178
                5297402
                28138123
                aba7ebe3-94dc-4af0-852b-6d14477e1c77
                © Am J Case Rep, 2017

                This work is licensed under Creative Common Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0)

                History
                : 31 October 2016
                : 16 November 2016
                Categories
                Articles

                atherosclerosis,coronary artery disease,coronary occlusion,percutaneous coronary intervention

                Comments