      Using gamma index to flag changes in anatomy during image‐guided radiation therapy of head and neck cancer

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          During radiation therapy of head and neck cancer, the decision to consider replanning a treatment because of anatomical changes has significant resource implications. We developed an algorithm that compares cone‐beam computed tomography ( CBCT) image pairs and provides an automatic alert as to when remedial action may be required. Retrospective CBCT data from ten head and neck cancer patients that were replanned during their treatment was used to train the algorithm on when to recommend a repeat CT simulation (re‐ CT). An additional 20 patients (replanned and not replanned) were used to validate the predictive power of the algorithm. CBCT images were compared in 3D using the gamma index, combining Hounsfield Unit ( HU) difference with distance‐to‐agreement ( DTA), where the CBCT study acquired on the first fraction is used as the reference. We defined the match quality parameter ( MQP x) as a difference between the x th percentiles of the failed‐pixel histograms calculated from the reference gamma comparison and subsequent comparisons, where the reference gamma comparison is taken from the first two CBCT images acquired during treatment. The decision to consider re‐ CT was based on three consecutive MQP values being less than or equal to a threshold value, such that re‐ CT recommendations were within ±3 fractions of the actual re‐ CT order date for the training cases. Receiver‐operator characteristic analysis showed that the best trade‐off in sensitivity and specificity was achieved using gamma criteria of 3 mm DTA and 30 HU difference, and the 80 th percentile of the failed‐pixel histogram. A sensitivity of 82% and 100% was achieved in the training and validation cases, respectively, with a false positive rate of ~30%. We have demonstrated that gamma analysis of CBCT‐acquired anatomy can be used to flag patients for possible replanning in a manner consistent with local clinical practice guidelines.

          Related collections

          Most cited references16

          • Record: found
          • Abstract: found
          • Article: not found

          A technique for the quantitative evaluation of dose distributions.

          The commissioning of a three-dimensional treatment planning system requires comparisons of measured and calculated dose distributions. Techniques have been developed to facilitate quantitative comparisons, including superimposed isodoses, dose-difference, and distance-to-agreement (DTA) distributions. The criterion for acceptable calculation performance is generally defined as a tolerance of the dose and DTA in regions of low and high dose gradients, respectively. The dose difference and DTA distributions complement each other in their useful regions. A composite distribution has recently been developed that presents the dose difference in regions that fail both dose-difference and DTA comparison criteria. Although the composite distribution identifies locations where the calculation fails the preselected criteria, no numerical quality measure is provided for display or analysis. A technique is developed to unify dose distribution comparisons using the acceptance criteria. The measure of acceptability is the multidimensional distance between the measurement and calculation points in both the dose and the physical distance, scaled as a fraction of the acceptance criteria. In a space composed of dose and spatial coordinates, the acceptance criteria form an ellipsoid surface, the major axis scales of which are determined by individual acceptance criteria and the center of which is located at the measurement point in question. When the calculated dose distribution surface passes through the ellipsoid, the calculation passes the acceptance test for the measurement point. The minimum radial distance between the measurement point and the calculation points (expressed as a surface in the dose-distance space) is termed the gamma index. Regions where gamma > 1 correspond to locations where the calculation does not meet the acceptance criteria. The determination of gamma throughout the measured dose distribution provides a presentation that quantitatively indicates the calculation accuracy. Examples of a 6 MV beam penumbra are used to illustrate the gamma index.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119.

            AAPM Task Group 119 has produced quantitative confidence limits as baseline expectation values for IMRT commissioning. A set of test cases was developed to assess the overall accuracy of planning and delivery of IMRT treatments. Each test uses contours of targets and avoidance structures drawn within rectangular phantoms. These tests were planned, delivered, measured, and analyzed by nine facilities using a variety of IMRT planning and delivery systems. Each facility had passed the Radiological Physics Center credentialing tests for IMRT. The agreement between the planned and measured doses was determined using ion chamber dosimetry in high and low dose regions, film dosimetry on coronal planes in the phantom with all fields delivered, and planar dosimetry for each field measured perpendicular to the central axis. The planar dose distributions were assessed using gamma criteria of 3%/3 mm. The mean values and standard deviations were used to develop confidence limits for the test results using the concept confidence limit = /mean/ + 1.96sigma. Other facilities can use the test protocol and results as a basis for comparison to this group. Locally derived confidence limits that substantially exceed these baseline values may indicate the need for improved IMRT commissioning.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Quantification of volumetric and geometric changes occurring during fractionated radiotherapy for head-and-neck cancer using an integrated CT/linear accelerator system.

              Many patients receiving fractionated radiotherapy (RT) for head-and-neck cancer have marked anatomic changes during their course of treatment, including shrinking of the primary tumor or nodal masses, resolving postoperative changes/edema, and changes in overall body habitus/weight loss. We conducted a pilot study to quantify the magnitude of these anatomic changes with systematic CT imaging. Fourteen assessable patients were enrolled in this pilot study. Eligible patients had to have a pathologic diagnosis of head-and-neck cancer, be treated with definitive external beam RT, and had have gross primary and/or cervical nodal disease measuring at least 4 cm in maximal diameter. All patients were treated using a new commercial integrated CT-linear accelerator system (EXaCT) that allows CT imaging at the daily RT sessions while the patient remains immobilized in the treatment position. CT scans were acquired three times weekly during the entire course of RT, and both gross tumor volumes (GTVs: primary tumor and involved lymph nodes) and normal tissues (parotid glands, spinal canal, mandible, and external contour) were manually contoured on every axial slice. Volumetric and positional changes relative to a central bony reference (the center of mass of the C2 vertebral body) were determined for each structure. Gross tumor volumes decreased throughout the course of fractionated RT, at a median rate of 0.2 cm(3) per treatment day (range, 0.01-1.95 cm(3)/d). In terms of the percentage of the initial volume, the GTVs decreased at a median rate of 1.8%/treatment day (range, 0.2-3.1%/d). On the last day of treatment, this corresponded to a median total relative loss of 69.5% of the initial GTV (range, 9.9-91.9%). In addition, the center of the mass of shrinking tumors changed position with time, indicating that GTV loss was frequently asymmetric. At treatment completion, the median center of the mass displacement (after corrections for daily setup variation) was 3.3 mm (range, 0-17.3 mm). Parotid glands also decreased in volume (median, 0.19 cm(3)/d range, 0.04-0.84 cm(3)/d), and generally shifted medially (median, 3.1 mm; range, 0-9.9 mm) with time. This medial displacement of the parotid glands correlated highly with the weight loss that occurred during treatment. Measurable anatomic changes occurred throughout fractionated external beam RT for head-and-neck cancers. These changes in the external contour, shape, and location of the target and critical structures appeared to be significant during the second half of treatment (after 3-4 weeks of treatment) and could have potential dosimetric impact when highly conformal treatment techniques are used. These data may, therefore, be useful in the development of an adaptive RT scheme (periodic adjustment of the conformal treatment plan) that takes into account such treatment-related anatomic changes. In theory, such a strategy would maximize the therapeutic ratio of RT.
                Bookmark

                Author and article information

                Contributors
                bryan.schaly@lhsc.on.ca
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                13 September 2017
                November 2017
                : 18
                : 6 ( doiID: 10.1002/acm2.2017.18.issue-6 )
                : 79-87
                Affiliations
                [ 1 ] Departments of Oncology and Medical Biophysics Western University London ON Canada
                [ 2 ] Physics and Engineering London Regional Cancer Program London ON Canada
                [ 3 ] Schulich School of Medicine and Dentistry Western University London ON Canada
                [ 4 ] Department of Radiation Oncology London Regional Cancer Program London ON Canada
                [ 5 ] Department of Radiation Therapy London Regional Cancer Program London ON Canada
                Author notes
                [*] [* ] Author to whom correspondence should be addressed. Bryan Schaly

                E‐mail: bryan.schaly@ 123456lhsc.on.ca

                Article
                ACM212180
                10.1002/acm2.12180
                5689936
                28901659
                a831a736-3688-4484-a9ed-f966029b02d9
                © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

                This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 25 May 2017
                : 04 August 2017
                : 09 August 2017
                Page count
                Figures: 8, Tables: 0, Pages: 9, Words: 6822
                Categories
                87.53.Jw
                Radiation Oncology Physics
                Radiation Oncology Physics
                Custom metadata
                2.0
                acm212180
                November 2017
                Converter:WILEY_ML3GV2_TO_NLMPMC version:5.2.5 mode:remove_FC converted:16.11.2017

                adaptive radiation therapy,anatomical variations,gamma index,head and neck cancer,image‐guided radiation therapy

                Comments

                Comment on this article