      Prosthesis-patient Mismatch - What Cardiac Anesthesiologists Need to Know?

      review-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Prosthesis-patient Mismatch (PPM) is not uncommon with an incidence reported up to 70% after aortic valve (AV) replacement. Severe forms of PPM are less common (up to 20%); PPM can lead to increased short- and long-term morbidity and mortality. It is important to discriminate PPM from other forms of prosthetic valve dysfunction. Sometimes, prosthetic valve degenerative disease may coexist with PPM. Echocardiography plays an important role in the prevention and diagnosis of PPM. Preemptive strategies to prevent PPM include insertion of newer generation prosthetic valves with better hemodynamic characteristics, stentless prosthesis, aortic root enlargement to insert a larger prosthesis, aortic homograft, and transcutaneous AV implantation. We present an illustrative case and review the literature on PPM pertinent to anesthesiologists.

          Related collections

          Most cited references57

          • Record: found
          • Abstract: found
          • Article: not found

          The problem of valve prosthesis-patient mismatch.

          Valve prostheses have played an important part in the past two decades in the management of patients with valvular heart disease. However, many of the devices used in valve replacement have introduced new clinical problems. This paper deals with some of the problems associated with valve replacement, including one not previously emphasized--valve prosthesis-patient mismatch, which may cause obstruction to ventricular outflow and/or inflow.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            The impact of prosthesis-patient mismatch on long-term survival after aortic valve replacement: a systematic review and meta-analysis of 34 observational studies comprising 27 186 patients with 133 141 patient-years.

            Numerous studies have linked prosthesis-patient mismatch (PPM) after aortic valve replacement (AVR) to adverse outcomes. Its correlation with long-term survival has been described but with contradicting results. This systematic review and meta-analysis of observational studies aims to determine the hazard of PPM after AVR. The Medline and EMBase databases were searched for English-language original publications. Two researchers independently screened studies and extracted data. Pooled estimates were obtained by random effects model. Subgroup analyses were performed to detect sources of heterogeneity. The search yielded 348 potentially relevant studies; 34 were included comprising 27 186 patients and 133 141 patient-years. Defined by the universally accredited indexed effective orifice area <0.85 cm(2)/m(2), 44.2% of patients were categorized as having PPM. In 34.2 and 9.8% of patients moderate (0.65-0.85 cm(2)/m(2)) and severe (<0.65 cm(2)/m(2)) PPM was present, respectively. Prosthesis-patient mismatch was associated with a statistically significant increase in all-cause mortality (HR = 1.34, 95% CI: 1.18-1.51), but only a trend to an increase in cardiac-related mortality (HR = 1.51, 95% CI: 0.88-2.60) was recognized. Analysis by severity of PPM demonstrated that both moderate and severe PPM increased all-cause mortality (HR = 1.19, 95% CI: 1.07-1.33 and HR = 1.84, 95% CI: 1.38-2.45) and cardiac-related mortality (HR = 1.32, 95% CI: 1.02-1.71 and HR = 6.46, 95% CI: 2.79-14.97). Further analyses showed a consistent effect over separate time intervals during follow-up. Prosthesis-patient mismatch is associated with an increase in all-cause and cardiac-related mortality over long-term follow-up. We recommend that current efforts to prevent PPM should receive more emphasis and a widespread acceptance to improve long-term survival after AVR.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Prosthesis-patient mismatch: definition, clinical impact, and prevention.

              Prosthesis-patient mismatch (PPM) is present when the effective orifice area of the inserted prosthetic valve is too small in relation to body size. Its main haemodynamic consequence is to generate higher than expected gradients through normally functioning prosthetic valves. This review updates the present knowledge about the impact of PPM on clinical outcomes. PPM is common (20-70% of aortic valve replacements) and has been shown to be associated with worse haemodynamic function, less regression of left ventricular hypertrophy, more cardiac events, and lower survival. Moreover, as opposed to most other risk factors, PPM can largely be prevented by using a prospective strategy at the time of operation.
                Bookmark

                Author and article information

                Journal
                Ann Card Anaesth
                Ann Card Anaesth
                ACA
                Annals of Cardiac Anaesthesia
                Medknow Publications & Media Pvt Ltd (India )
                0971-9784
                0974-5181
                Apr-Jun 2017
                : 20
                : 2
                : 234-242
                Affiliations
                [1] Department of Anesthesiology, University of Pittsburgh, Pittsburgh, PA, USA
                [1 ] Board Certified Cardiac Anesthesiologist, Tabriz, Iran
                Author notes
                Address for correspondence: Dr. Kathirvel Subramaniam, Associate Professor, Department of Anesthesiology, University of Pittsburgh, Pittsburgh, PA, USA. E-mail: subramaniamk@ 123456upmc.edu
                Article
                ACA-20-234
                10.4103/aca.ACA_9_17
                5408531
                28393786
                9a863e0b-2330-4be2-9e12-faa6b5443563
                Copyright: © 2017 Annals of Cardiac Anaesthesia

                This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.

                History
                Categories
                Review Article

                aortic root enlargement,aortic valve,indexed effective orifice area,patient-prosthesis mismatch

                Comments

                Comment on this article