      Fusion of real-time 3D transesophageal echocardiography and cardiac fluoroscopy imaging in transapical catheter-based mitral paravalvular leak closure

      brief-report

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Introduction Real-time 3D transesophageal echocardiography (TEE) with cardiac fluoroscopy X-ray imaging is well introduced into practice and is used to monitor and assist catheter-based cardiac operations [1–6]. Cardiac fluoroscopy imaging is useful to monitor and control delivery tools into the desired site of the heart during transcatheter closure of paravalvular defects [3, 7, 8]. Fluoroscopy is superior to 3D TEE in visualization of the catheters, wires, devices, and calcified and metallic structures. The main disadvantages of fluoroscopy are the inability to determine the three-dimensional anatomy of intracardiac tissues, exposure to radiation and the need for contrast media. 3D TEE is a standard method to visualize intracardiac structural pathology and measure its hemodynamic significance. It is also used to control intracardiac manipulation of various instruments as well as paravalvular leak closure devices [8]. 3D TEE provides spatial intracardiac anatomy imaging in multiple planes, but its advantages are limited by echo shadowing, which reduces visualization of the catheters and metallic structures. Also experienced personnel are needed to achieve high quality 3D imaging and guide the procedures [7]. Merging these two imaging techniques combines the advantages of fluoroscopy and 3D TEE during paravalvular leak closure. As to date there are no studies comparing mitral paravalvular leak closure with or without fusion technology, we hypothesize that periprocedural fusion of cardiac fluoroscopy and real-time 3D TEE may make intracardiac procedures more controlled, precise and easier. Also it may reduce patients’ and professionals’ exposure to radiation [7]. We present three cases of transapical catheter-based mitral paravalvular leak closure. All patients were treated with devices specifically designed and manufactured by Occlutech Holding, Switzerland for paravalvlar leak closure. The Occlutech® Paravalvular Leak Device represents a novel device with unique rectangular- and square-shaped designs. Occlutech® Paravalvular Leak Device was CE marked in 2014 and is the first transcatheter device indicated and approved for aortic and mitral PVL closure. Procedures were performed with the assistance of fusion (EchoNavigator system, Phillips Medical Systems, Netherlands) of cardiac fluoroscopy imaging (Phillips AlluraClarity, Phillips Medical Systems, Netherlands) and real-time 3D TEE (Phillips Epiq7, Phillips Medical Systems, Netherlands). Procedure description The procedure was carried out in a hybrid operating room. Standard cardiac surgery monitoring was used. Prior to surgical draping, two self-adhesive external defibrillator pads were applied to the patient’s back. With the patient in a supine position under general anesthesia and single lumen endotracheal intubation, left anterolateral thoracotomy (5–6 cm) was performed at the level of the left ventricle apex (preoperative TEE guidance). The pericardium was identified and opened. Blunt dissection of the adhesions was performed. Stay sutures were placed at both edges of the pericardium and hinged to the edges of the skin under the retractor. Two “U” shape sutures reinforced with Teflon pledgets (5 × 15 mm) were placed and secured with the tourniquets perpendicular to each other at the apex of the heart. Then heparin was intravenously administered (150 U/kg), aiming for an activated clotting time of 250 to 300 s. A needle puncture between four pledgets was performed and the guidewire was introduced into the left ventricle with subsequent short 6 Fr catheter sheath insertion. The table and C-arm are oriented for the TEE probe to be visible in the fluoroscopy field; this maneuver allows the fusion software EchoNavigator system to “recognize” the TEE probe and co-register its position with the table and angulation of the C-arm. When co-registration is “successful” the C-arm and TEE probe can be manipulated with the fused images maintained [9]. Real-time 3D TEE is performed to identify and confirm the exact location, size, depth and shape of the paravalvular leak. The location of the paravalvular leak is marked in the EchoNavigator system to aid the pathway of the guidewire crossing the paravalvular leak [10]. A hydrophilic guidewire was used to pass through the defect with the help of a guiding catheter. The guiding catheter was advanced through the leak and replaced with a stiff wire. The delivery sheath was chosen according to the size of the occluder. The guiding catheter is removed and the delivery sheath is advanced through the defect. The paravalvular leak closure device was deployed under control of real-time 3D TEE and fluoroscopy. Position, orientation and hemodynamic effect of the closure device were checked. If the function of the prosthetic valve was not compromised, occluder discs were fully expanded and the paravalvular regurgitation closure device detached from the delivery system. Following a successful paravalvular leak closure, catheters and sheaths were removed from the left ventricle and protamine sulfate administered. “U” shape sutures were securely tightened and the pericardium was closed with three or two interrupted sutures, leaving “communication” with the pleural space to allow blood to drain from the pericardial cavity. The pleural cavity was drained with one drain; usually it was inserted into one or two intercostal spaces below the incision. Patients were transferred to the cardiac intensive therapy unit and monitored for bleeding and hemodynamic instability. If no complication occurred patients were weaned from the ventilator, extubated and transferred to the ward on the next postoperative day. Patients treated Detailed descriptions of the patients’ demographic, clinical and procedural data are presented in Table I. Table I Patients’ demographic, clinical and procedural data Patients’ data Patient 1 Patient 2 Patient 3 Age [years] 61 57 64 Gender Female Male Female Mitral prosthesis description 27 mmBileaflet mechanical 31 mmBileaflet mechanical 29 mmBileaflet mechanical 6 min walking distance [m] 165 357 480 GFR [ml/min/1.73 m2] 46 79 78 Hemoglobin [g/l] 89 81 85 LDH [U/l] 331 430 818 Haptoglobin [g/l] 0.68 < 0.08 < 0.08 LVEF [%] 40 45 45 Systolic pulmonary artery pressure [mm Hg] 65 50 42 EuroSCORE [%] 10.56 3.7 7.08 PVL dimensions [mm] 9 × 4 8 × 5 2 × 2 4 × 3 2 × 4 3 × 2 PVL location according to the clock and segments of the mitral valve leaflets 3 o’clock(A3) 2 o’clock(A3) 6 o’clock(P2) 11 o’clock(A1) 2 o’clock(A3) 4 o’clock(P3) Severity of PVL regurgitation Severe Severe Insignificant Moderate Mild Mild Indications for PVL closure Heart failure and anemia Anemia Anemia Procedural data:  Occlutech PLD occluder article no. 61PLD12W 61PLD12W Not closed 60PLD5W 60PLD4W 62PLD3T  Occluder waist size [mm] 12 × 5 12 × 5 5 × 5 5 × 5 3 × 3  Size of the atrial [distal] disc of the occluder [mm] 21 21 14 × 14 13 × 13 11.5 × 11.5  Size of the ventricular (proximal) disc of the occlude [mm] 19 19 12.5 × 12 11.5 × 11.5 10 × 10  Occluder type Waist Waist Waist Waist Twist  Occluder shape Rectangular Rectangular Square Square Square  Sheath size [F] 9 9 6 6 6  Occluder delivery system Flex Pusher a item no. 50FP120 & 50FP120L 50FP120 & 50FP120L 50FP100 & 50FP100L 50FP100 & 50FP100L 50FP100 & 50FP100L  Occluder type Waist Waist  Cumulative fluoroscopy time 8 min 27 s 36 min 3 s 29 min 46 s  Total cumulative air kerma [mGy] 70.67 169.69 157.7 GFR – glomerular filtration rate, LDH – lactate dehydrogenase, LVEF – left ventricular ejection fraction, PVL – paravalvular leak, PLD – a trade name for percutaneous trans-catheter device for paravalvular leak closure, a trade name for delivery system for paravalvular leak closure device. 1st patient Thirteen years ago the 61-year-old woman underwent mitral valve replacement with a 27 mm bileaflet mechanical valve prosthesis and radiofrequency maze procedure for severe rheumatic mitral stenosis and long standing persistent atrial fibrillation. Aspiration of the pleural effusion was complicated by severe bleeding from the lung and was treated by revision through the right thoracotomy. The patient developed sepsis (Stenotrophomonas maltophilia) with subsequent severe mitral paravalvular leak. Four years later the patient underwent redo surgery and mitral valve paravalvular leak suture repair through a right thoracotomy. Intraoperative TEE did not show any residual paravalvular leak. Unfortunately at discharge following that surgery on transthoracic echocardiography a paravalvular leak at the same location with severe mitral regurgitation was noted. On the basis of the patient’s refusal, redo surgery was declined. For the following 9 years the patient was managed medically. With the patient’s functional capacity gradually declining, she agreed to an elective transapical catheter-based procedure for mitral paravalvular leak. The patient underwent an uneventful transapical catheter-based mitral paravalvular leak closure procedure (Figure 1). Immediate intraoperative hemodynamic improvement was noted; after closure of the paravalvular leak the central venous pressure decreased from 20 mm Hg to 9 mm Hg, and systolic pulmonary artery (PA) pressure from 65 mm Hg to 40 mm Hg. Figure 1 Patient 1 periprocedural fusion of fluoroscopy and real-time 3D TEE. A – Preprocedural (top) and postprocedural (bottom) 3D TEE images of the mitral prosthesis. The arrows point to the orifice of the paravalvular leak (PVL) located at 2 o’clock (surgical view) and the occluder position after the deployment. B – Four fused fluoroscopy and real-time TEE images. The X-plain color Doppler images of PVL regurgitant flow (top) are oriented according to C-arm position (left bottom) and are fused with the fluoroscopy image (bottom). The location of PVL is marked by a red circle. C – Guidewire passage through the paravalvular defect. The fluoroscopy image (on the bottom) demonstrates wire crossing the red circle; 3D TEE image (top) confirms the guidewire passage through the defect. The arrow in the 3D TEE image shows the guidewire crossing the defect Ao – position of the aorta. The postoperative course was uneventful and the patient was discharged from hospital on the 11th postoperative day. Discharge echo showed no paravalvular regurgitation and good left ventricle systolic function (LVEF > 55%). 2nd patient A 57-year-old man with previously known moderate aortic stenosis developed infective endocarditis of the aortic and mitral valves. He underwent aortic and mitral valve replacement with 21 mm and 31 mm bileaflet mechanical prostheses respectively. Three weeks after the operation the patient was diagnosed with severe anemia. The cause of anemia was established following transesophageal echocardiography which revealed severe mitral regurgitation due to paravalvular leak. The patient was scheduled for elective transapical catheter-based paravalvular defect closure. He underwent an uneventful transapical catheter-based mitral paravalvular leak closure procedure (Figure 2) with complete closure of the defect. On intraoperative TEE no regurgitant flow across the occluded defect was noted. His postoperative course was uneventful. The patient was discharged on the 11th postoperative day. Discharge echocardiography showed good left ventricle systolic function (LVEF > 50%), systolic PA pressure of 55 mm Hg, trivial regurgitation through the occluded defect and trivial regurgitation through the untreated defect at 6 o’clock. Figure 2 Patient 2 periprocedural fusion of fluoroscopy and real-time 3D TEE. A – The preprocedural color flow Doppler (top) and 3D images (bottom) demonstrates bileaflet mechanical mitral valve prostheses (P) and the paravalvular leak located at 2 o’clock (arrow). B – Two dimensional X-plain images (top), X-plain echo images are oriented according to C-arm position (left bottom) and are fused with the fluoroscopy image (right bottom). Position of the prostheses is marked by a green circle in all images. The red circle identifies the location of the defect. C – Guidewire passage through the paravalvular defect. The fluoroscopy image (on the bottom) demonstrates wire crossing the red circle; 3D TEE image (top) confirms the guidewire passage through the defect. The arrow in the 3D TEE image shows the guidewire crossing the defect Ao – position of the aorta. 3rd patient Two years ago a 64-year-old woman underwent aortic and mitral valve replacement with 25 mm and 29 mm mechanical prostheses respectively for rheumatic stenosis. Fourteen months after surgery she developed hemolytic anemia and presented with symptoms of general weakness. Transesophageal echocardiography revealed severe mitral regurgitation due to three paravalvular defects. For the 4 following months she was managed medically and 18 months after initial surgery the patient underwent a redo conventional operation and suture repair for mitral paravalvular leak. Unfortunately, discharge transthoracic echocardiography revealed reoccurrence of severe mitral paravalvular regurgitation. She was scheduled for elective transapical catheter-based mitral paravalvular leak closure a month later. The patient underwent an uneventful transapical catheter-based mitral paravalvular leak closure procedure (Figure 3) with complete occlusion of all three defects. At discharge on transthoracic echocardiography no regurgitant flow across the occluded defects and good left ventricle systolic function were noted. The patient was discharged on the 6th postoperative day with no postoperative complications. Figure 3 Patient 3 periprocedural fusion of fluoroscopy and real-time 3D TEE. A – The preprocedural 3D TEE color flow Doppler image (top) and 3D TEE image (bottom) with paravalvular defects closed with occluders. Arrows point to the defects. B – Real-time 3D TEE image (top left and bottom left) demonstrating ventricular perspective of the mitral prosthesis where colored (red, white, green) circles indicate paravalvular defects. Realtime 3D TEE image (top right) demonstrates atrial perspective of the mitral prosthesis where colored (red, white, green) circles indicate paravalvular defects. Fused fluoroscopy and real-time TEE image (bottom right) demonstrates wire crossing the green circle; 3D TEE image (top right) confirms the guidewire passage through the defect (arrow pointing to the guidewire) Ao – position of the aorta. Comment In recent years, catheter-based closure of the mitral paravalvular leak has advanced globally. Nevertheless, due to the absence of standardized methods to quantify the regurgitation of the paravalvular leak, disagreement on the definition of success, tortuous anatomy of the defects and recent absence of dedicated devices for the closure of paravalvular leaks, the results of this method of treatment were debatable [8, 11–13]. The success rate of catheter-based paravalvular leak closure via the transapical approach reaches 100% compared to 66–78% in the trans-septal approach [11]. The closure of a mitral paravalvular leak using the transapical approach and real-time fluoroscopy and 3D TEE is efficient and safe [14]. Recent technological advances such as the ability to fuse fluoroscopy and real-time 3D TEE, as well as the development of devices dedicated to paravalvular leak closure, allow closure of a mitral paravalvular leak to be performed more conveniently [8, 15]. Fusion is needed to guide the pathway of the pass through the paravalvular lesion. We have presented three cases of mitral paravalvular leak closure where this complex approach was applied. All three procedures showed technical and procedural success. Conclusions Fusion of fluoroscopy and real-time 3D TEE is a useful method in catheter-based paravalvular leak closure via the transapical approach.

          Related collections

          Most cited references15

          • Record: found
          • Abstract: found
          • Article: not found

          Transcatheter reduction of paravalvular leaks: a systematic review and meta-analysis.

          Significant paravalvular leak (PVL) after surgical valve replacement can result in intractable congestive heart failure and hemolytic anemia. Because repeat surgery is performed in only few patients, transcatheter reduction of PVL is emerging as an alternative option, but its safety and efficacy remain uncertain. In this study we sought to assess whether a successful transcatheter PVL reduction is associated with an improvement in clinical outcomes.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Registration of 3D trans-esophageal echocardiography to X-ray fluoroscopy using image-based probe tracking.

            Two-dimensional (2D) X-ray imaging is the dominant imaging modality for cardiac interventions. However, the use of X-ray fluoroscopy alone is inadequate for the guidance of procedures that require soft-tissue information, for example, the treatment of structural heart disease. The recent availability of three-dimensional (3D) trans-esophageal echocardiography (TEE) provides cardiologists with real-time 3D imaging of cardiac anatomy. Increasingly X-ray imaging is now supported by using intra-procedure 3D TEE imaging. We hypothesize that the real-time co-registration and visualization of 3D TEE and X-ray fluoroscopy data will provide a powerful guidance tool for cardiologists. In this paper, we propose a novel, robust and efficient method for performing this registration. The major advantage of our method is that it does not rely on any additional tracking hardware and therefore can be deployed straightforwardly into any interventional laboratory. Our method consists of an image-based TEE probe localization algorithm and a calibration procedure. While the calibration needs to be done only once, the GPU-accelerated registration takes approximately from 2 to 15s to complete depending on the number of X-ray images used in the registration and the image resolution. The accuracy of our method was assessed using a realistic heart phantom. The target registration error (TRE) for the heart phantom was less than 2mm. In addition, we assess the accuracy and the clinical feasibility of our method using five patient datasets, two of which were acquired from cardiac electrophysiology procedures and three from trans-catheter aortic valve implantation procedures. The registration results showed our technique had mean registration errors of 1.5-4.2mm and 95% capture range of 8.7-11.4mm in terms of TRE.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Hybrid Imaging During Transcatheter Structural Heart Interventions

              Fusion of different imaging modalities has gained increasing popularity over the last decade. However, most fusions are done between static rather than dynamic images. In order to adequately visualize the complex three-dimensional structures of the beating heart, high-temporal and spatial image resolutions are mandatory. Currently, only the combination of transesophageal echocardiography with fluoroscopy allows real-time image fusion of high quality during structural heart disease (SHD) interventions. The use of markers as well as real-time image overlay greatly facilitates communication between SHD team members and potentially increases procedural success while reducing radiation dose and use of contrast. However, to date there is only limited evidence that fusion imaging improves safety and outcomes of SHD interventions. This review highlights the benefits of fusion imaging during SHD interventions such as transseptal puncture and closure of atrial septal defects and left atrial appendage as well as interventions on the mitral and aortic valve.
                Bookmark

                Author and article information

                Journal
                Postepy Kardiol Interwencyjnej
                Postepy Kardiol Interwencyjnej
                PWKI
                Postępy w Kardiologii Interwencyjnej = Advances in Interventional Cardiology
                Termedia Publishing House
                1734-9338
                1897-4295
                25 September 2017
                2017
                : 13
                : 3
                : 263-268
                Affiliations
                [1 ]Department of Cardiovascular Medicine, Vilnius University, Vilnius, Lithuania
                [2 ]Cardiovascular Department, Humanitas Gavazzeni, Bergamo, Italy
                Author notes
                Corresponding author: Aleksejus Zorinas, Heart Surgery Centre, Vilnius University Hospital Santariskiu Klinikos, Santariskiu 2, 08406 Vilnius, Lithuania, phone: +370 52365000, fax: +370 52365111, mobile: +370 698 29079. e-mail: aleksejus.zorinas@ 123456santa.lt
                Article
                30625
                10.5114/aic.2017.70200
                5644046
                95ad518b-4f25-48c1-917b-30af18710ad5
                Copyright: © 2017 Termedia Sp. z o. o.

                This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License, allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.

                History
                : 11 April 2017
                : 17 June 2017
                Categories
                Short Communication

                Comments

                Comment on this article