          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Accurate and consistent viral load (VL) quantitation of HIV type 1 (HIV-1), hepatitis B virus (HBV), and hepatitis C virus (HCV) is important for diagnosis and clinical monitoring. Assay results have to be concordant and compatible across laboratories. We evaluated the performance of three Aptima assays (Hologic, San Diego, CA, USA) and compared their VL values with corresponding cobas 6800 assay (Roche Diagnostics, Mannheim, Germany) results, using 840 clinical samples.

          Methods

          The correlation between VL results obtained using the two assays was evaluated in terms of analytical sensitivity, precision/reproducibility, linearity, and cross-reactivity. Agreement rates were determined using kappa statistics. The overall agreement of VL values was examined using Passing–Bablok regression analysis.

          Results

          All CVs were within 5%; the assays had good precision for detecting all three viruses. The linearity of quantitation assessed using three AccuSpan linearity panels (Seracare, Milford, MA, USA), was excellent for the Aptima assays. For HIV-1 and HCV, the results of both assays showed excellent agreement (κ=0.89 and 0.90, respectively) while for HBV, the results showed good agreement (κ=0.69). For analytical sensitivity, the VLs required for a 100% detection rate of HIV-1, HBV, and HCV were 20 copies/mL, 7.5 IU/mL, and 5.0 IU/mL, respectively. The results for HIV-1, HBV, and HCV obtained using both assays correlated strongly (R 2=0.97, 0.93, and 0.95, respectively).

          Conclusions

          The cobas 6800 and Aptima assays, with fully automated and high-throughput molecular platforms for HIV-1, HBV, and HCV VL measurements, show good analytical performance and a strong correlation between results. The study results suggest that the assays can be used interchangeably for long-term monitoring of chronic infections.

          Related collections

          Most cited references34

          • Record: found
          • Abstract: found
          • Article: not found

          EASL 2017 Clinical Practice Guidelines on the management of hepatitis B virus infection.

          Hepatitis B virus (HBV) infection remains a global public health problem with changing epidemiology due to several factors including vaccination policies and migration. This Clinical Practice Guideline presents updated recommendations for the optimal management of HBV infection. Chronic HBV infection can be classified into five phases: (I) HBeAg-positive chronic infection, (II) HBeAg-positive chronic hepatitis, (III) HBeAg-negative chronic infection, (IV) HBeAg-negative chronic hepatitis and (V) HBsAg-negative phase. All patients with chronic HBV infection are at increased risk of progression to cirrhosis and hepatocellular carcinoma (HCC), depending on host and viral factors. The main goal of therapy is to improve survival and quality of life by preventing disease progression, and consequently HCC development. The induction of long-term suppression of HBV replication represents the main endpoint of current treatment strategies, while HBsAg loss is an optimal endpoint. The typical indication for treatment requires HBV DNA >2,000IU/ml, elevated ALT and/or at least moderate histological lesions, while all cirrhotic patients with detectable HBV DNA should be treated. Additional indications include the prevention of mother to child transmission in pregnant women with high viremia and prevention of HBV reactivation in patients requiring immunosuppression or chemotherapy. The long-term administration of a potent nucleos(t)ide analogue with high barrier to resistance, i.e., entecavir, tenofovir disoproxil or tenofovir alafenamide, represents the treatment of choice. Pegylated interferon-alfa treatment can also be considered in mild to moderate chronic hepatitis B patients. Combination therapies are not generally recommended. All treated and untreated patients should be monitored for treatment response and adherence, and the risk of progression and development of complications. HCC remains the major concern for treated chronic hepatitis B patients. Several subgroups of patients with HBV infection require specific focus. Future treatment strategies to achieve 'cure' of disease and new biomarkers are discussed.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Update on prevention, diagnosis, and treatment of chronic hepatitis B: AASLD 2018 hepatitis B guidance.

              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              EASL Recommendations on Treatment of Hepatitis C 2018

                Bookmark

                Author and article information

                Journal
                Ann Lab Med
                Ann Lab Med
                Annals of Laboratory Medicine
                Korean Society for Laboratory Medicine
                2234-3806
                2234-3814
                1 July 2022
                1 July 2022
                1 July 2022
                : 42
                : 4
                : 447-456
                Affiliations
                [1 ]Department of Laboratory Medicine, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea
                [2 ]Department of Laboratory Medicine, Hallym University Sacred Heart Hospital, Anyang, Korea
                Author notes
                Corresponding author: Sinyoung Kim, M.D., Ph.D. Department of Laboratory Medicine, Severance Hospital, Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemun-gu, Seoul 03722, Korea Tel: +82-2-2228-2452 Fax: +82-2-2227-8353 E-mail: sykim@ 123456yuhs.ac
                Author information
                https://orcid.org/0000-0001-8458-1495
                https://orcid.org/0000-0003-0078-1145
                https://orcid.org/0000-0002-2609-8945
                Article
                alm-42-4-447
                10.3343/alm.2022.42.4.447
                8859551
                35177565
                90d5d83f-6775-47d4-b4a9-36e89cd3d2bb
                © Korean Society for Laboratory Medicine

                This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License ( http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 4 March 2021
                : 6 June 2021
                : 28 January 2022
                Categories
                Original Article
                Diagnostic Immunology

                Clinical chemistry
                hiv,hepatitis b virus,hepatitis c virus,viral load,aptima assay,cobas 6800 assay,performance,agreement,analytical sensitivity