      Performance evaluation and clinical application of the UA-5600 automatic dry chemistry urine analyzer in renal diseases

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Urine testing as a routine screening programme, abnormal test results can be suggestive to clinicians but can sometimes be overlooked, and the establishment of a diagnostic model can better assist clinicians in identifying potential problems. BLD (blood), LEU (leukocyte), PRO (protein) and GLU (glucose) are the four most important parameters in urine testing, and the accuracy of their results is a key concern for clinicians, so it is essential to verify the accuracy of their results. In this study, we evaluated the analytical and clinical performance of Mindray’s automatic urine dry chemistry analyzer, the UA-5600 (Hereinafter referred to as the (UA-5600), and the test strips configured with the instrument, and developed a machine-learning (ML) model for kidney disease screening from the results of 11 parameters output from the UA-5600 with the aim of detecting abnormal urine test results.

          Methods

          Urine samples from outpatients and inpatients at The First Affiliated Hospital of Sun Yat-sen University were collected from August to September 2022 to evaluate the performance of the Mindray UA-5600 dry chemistry analyzer and test strips. The evaluation of the UA-5600 and its test strips focused on the agreement of the urine BLD and LEU readings with the RBC (red blood cell) and WBC (white blood cell) counts obtained by the Mindray EH-2090 urine formed element analyzer. We also compared the PRO and GLU readings with the results of the Mindray BS-2800M biochemistry analyzer. Urine samples from outpatients and inpatients were retrospectively analysed and grouped according to LIS diagnosis. Additionally, eight ML models for kidney disease screening were developed using 11 parameters measured by the UA-5600. And the model was validated by the validation set.

          Results

          The UA-5600 had an 89.55% concordance rate for BLD and a 91.04% concordance rate for LEU compared to the EH-2090 analyzer. When benchmarked against the BS-2800M, the concordance rates for PRO and GLU were 94.14% and 95.20%, respectively. A total of 1,691 samples were used for the construction of the ML models, of which 346 patients (135 males and 211 females, age range: 18 to 98 years) diagnosed with renal disease, and 1,345 patients (397 males and 948 females, age range: 18 to 92 years) with non-renal disease diagnosed with other conditions. Notably, the Naïve Bayes (NB) model, which was built from the UA-5600 parameters, demonstrated superior predictive capabilities for renal disease, with an area under the receiver operating characteristic curve of 0.9470, a sensitivity of 0.7767, and a specificity of 0.9457.

          Conclusions

          The Mindray UA-5600 demonstrates robust detection abilities for both BLD and LEU, and its results for PRO and GLU align closely with those obtained from the chemistry analyzer. The NB model has a good screening ability and shows promise as an effective screening tool.

          Related collections

          Most cited references24

          • Record: found
          • Abstract: found
          • Article: not found

          Receiver Operating Characteristic (ROC) Curve Analysis for Medical Diagnostic Test Evaluation.

          This review provides the basic principle and rational for ROC analysis of rating and continuous diagnostic test results versus a gold standard. Derived indexes of accuracy, in particular area under the curve (AUC) has a meaningful interpretation for disease classification from healthy subjects. The methods of estimate of AUC and its testing in single diagnostic test and also comparative studies, the advantage of ROC curve to determine the optimal cut off values and the issues of bias and confounding have been discussed.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Foundation models for generalist medical artificial intelligence

              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              Machine learning in medical applications: A review of state-of-the-art methods

                Bookmark

                Author and article information

                Journal
                Transl Androl Urol
                Transl Androl Urol
                TAU
                Translational Andrology and Urology
                AME Publishing Company
                2223-4683
                2223-4691
                27 June 2024
                30 June 2024
                : 13
                : 6
                : 1024-1036
                Affiliations
                [1 ]deptDepartment of Laboratory Medicine, The First Affiliated Hospital , Sun Yat-sen University , Guangzhou, China;
                [2 ]Department of Pathology and Laboratory Medicine, Emory University School of Medicine, Grady Health System , Atlanta, GA, USA
                Author notes

                Contributions: (I) Conception and design: S Zhang; (II) Administrative support: S Zhang; (III) Provision of study materials or patients: Y Liu, X Zheng; (IV) Collection and assembly of data: Y Liu, X Zheng; (V) Data analysis and interpretation: Y Liu, X Zheng; (VI) Manuscript writing: All authors; (VII) Final approval of manuscript: All authors.

                [#]

                These authors contributed equally to this work as co-first authors.

                Correspondence to: Shihong Zhang, BMed. Department of Laboratory Medicine, The First Affiliated Hospital, Sun Yat-sen University, No. 58 Zhongshan Second Road, Guangzhou 510080, China. Email: zhshih@ 123456mail.sysu.edu.cn
                Article
                tau-13-06-1024
                10.21037/tau-24-189
                11228674
                908b568d-9fd3-4f82-b4de-8111c7374036
                2024 Translational Andrology and Urology. All rights reserved.

                Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0.

                History
                : 18 April 2024
                : 12 June 2024
                Funding
                Funded by: the Wu Jieping Medical Foundation Clinical Research Special Grant Fund
                Award ID: No. 320. 6750. 2021- 06- 3
                Categories
                Original Article

                automatic dry chemistry urine analyzer,ua-5600,ml models,renal disease

                Comments

                Comment on this article