      Abbott® ID NOW™ COVID-19 rapid molecular assay versus Hologic® Panther Aptima™ SARS-CoV-2 assay in nasopharyngeal specimens: results from 1-year retrospective study in an emergency department.

      brief-report

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          We compared ID Now™ and Hologic® Panther Aptima™ for the detection of SARS-COV-2. ID Now™ showed a positive and negative percent agreement of 86.9% and 99.7% respectively. This facilitates faster clinical decision-making, along with rapid implementation of infection control measures, and improvement of patient flow in the emergency department towards inpatient wards.

          Related collections

          Most cited references8

          • Record: found
          • Abstract: found
          • Article: found
          Is Open Access

          Clinical Evaluation of Three Sample-to-Answer Platforms for Detection of SARS-CoV-2

          Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has now spread across the globe. As part of the worldwide response, many molecular diagnostic platforms have been granted emergency use authorization (EUA) by the Food and Drug Administration (FDA) to identify SARS-CoV-2 positive patients. Our objective was to evaluate three sample-to-answer molecular diagnostic platforms (Cepheid Xpert Xpress SARS-CoV-2 [Xpert Xpress], Abbott ID NOW COVID-19 [ID NOW], and GenMark ePlex SARS-CoV-2 Test [ePlex]) to determine analytical sensitivity, clinical performance, and workflow for the detection of SARS-CoV-2 in nasopharyngeal swabs from 108 symptomatic patients.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Performance of Abbott ID Now COVID-19 Rapid Nucleic Acid Amplification Test Using Nasopharyngeal Swabs Transported in Viral Transport Media and Dry Nasal Swabs in a New York City Academic Institution

            The recent emergence of the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) pandemic has posed formidable challenges for clinical laboratories seeking reliable laboratory diagnostic confirmation. The swift advance of the crisis in the United States has led to Emergency Use Authorization (EUA) facilitating the availability of molecular diagnostic assays without the more rigorous examination to which tests are normally subjected prior to FDA approval. Our laboratory currently uses two real-time reverse transcription-PCR (RT-PCR) platforms, the Roche Cobas SARS-CoV2 and the Cepheid Xpert Xpress SARS-CoV-2.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: found
              Is Open Access

              Comparison of Abbott ID Now and Abbott m2000 Methods for the Detection of SARS-CoV-2 from Nasopharyngeal and Nasal Swabs from Symptomatic Patients

              LETTER The ID Now COVID-19 (IDNCOV) assay performed on the ID Now instrument (Abbott Diagnostics, Inc., Scarborough, ME) is a rapid diagnostic test that can be performed in a point-of-care setting equivalent to Clinical Laboratory Improvement Amendments (CLIA)-waived testing. The assay utilizes isothermal amplification and can reportedly deliver results in approximately 5 to 13 min. As this assay could provide significant improvements to workflow in our hospital system, we sought to compare the performance of this test with our current coronavirus disease 2019 (COVID-19) assay, the Abbott RealTime SARS-CoV-2 (severe acute respiratory syndrome coronavirus 2) (ACOV) assay performed on the Abbott m2000 system (Abbott Molecular Inc., Des Plaines, IL). We compared the results from 524 paired foam nasal swabs (NS) tested on IDNCOV with nasopharyngeal swabs (NPS) placed in viral transport media tested on ACOV collected consecutively from symptomatic patients meeting current criteria for a diagnosis of COVID-19 (1). Five locations were included in the evaluation including three emergency departments (ED) and two immediate care centers (IMCC). IMCC A and ED 2 were experienced users of the IDNow platform. The other sites were new users of the platform and received training specifically for the IDNCOV. All ACOV testing was performed by one central clinical laboratory, and all NPS were heat inactivated for 30 min at 60°C prior to testing. NS were tested directly on the IDNCOV from IMCCs, and the tests were performed on-site. NS from the EDs were transported to the clinical microbiology laboratory in sterile transport containers (urine cups or conical tubes) and tested by laboratory personnel at each separate location. Statistical analysis was performed using SPSS v.26. The overall positivity rate in this sample collection was 35%, ranging from 22% to 60% among the five sites. Overall agreement was 75% positive agreement (95% confidence interval [95% CI], 67.74%, 80.67%) and 99% negative agreement (95% CI, 97.64%, 99.89%) between IDNCOV and ACOV for all specimens tested. Agreement at individual sites varied (Table 1). Two subjects tested positive on IDNCOV that were initially negative on ACOV. In case one, a repeat sample was positive on ACOV (repeat IDNCOV was not performed), and the case was resolved as a true positive result. For case two, all repeat testing (both IDNCOV and ACOV) was negative and was resolved as a likely false-positive result. This sample was collected during the first day of testing and could have been operator error. TABLE 1 Agreement between ACOV and IDNCOV Site Total no. of samples tested No. of samples with the following result a : % Positivity Positive agreement (95% CI) Negative agreement (95% CI) Performance agreement (kappa) (95% CI) A+/IND+ A+/IND− A−/IND+ A−/IND− IMCC A 208 33 13 1 161 22 71.74 (56.32, 83.54) 99.38 (96.09, 99.97) 0.783 (0.779, 0.788) IMCC B 125 39 17 0 69 44 69.64 (55.74, 80.84) 100.0 (93.43, 100.0) 0.711 (0.706, 0.717) ED 1 105 26 11 0 68 35 70.27 (52.83, 83.56) 100.0 (93.33, 100.0) 0.751 (0.744, 0.757) ED 2 31 12 3 0 16 50 80.0 (51.37, 94.69) 100 (75.92, 100.0) 0.803 (0.792, 0.814) ED 3 55 29 3 1 22 60 90.63 (73.83, 97.55) 95.65 (76.03, 99.77) 0.852 (0.844, 0.861) Total 524 139 47 2 336 35 74.73 (67.74, 80.67) 99.41 (97.64, 99.89) a Positive (+) and negative (−) results by ACOV (A) and IDNCOV (IND) are shown. Fleiss kappa analysis comparing the performance at each of the sites demonstrated that strength of agreement between the sites (Table 1) was rated as good to very good with comparable standard errors. We interpret this to mean that a site’s ability to run the test (or lack of experience) did not necessarily contribute to the variability in positivity that was found in this evaluation. Compared to the ACOV cycle numbers (CN) (which are similar but not directly comparable to cycle thresholds from other reverse transcription-PCR [RT-PCR] assays due to the unique ACOV assay design), a significant proportion, but not all, discordant samples exhibited at higher cycle numbers (Fig. 1). The mean CN for concordant positive samples was 12.71 (95% CI, 11.76, 13.67), ranging from 2.99 to 31.01, with a standard deviation of 5.5. The mean CN for discordant samples (ACOV positive [ACOV+]/IDNCOV negative [IDNCOV−]) was 21.07 (95% CI, 19.55, 22.60), ranging from 6.79 to 30.63, with a standard deviation of 5.1. These differences are statistically different (P = 6.75e−16). The stated limit of detection in the published instructions for use is 100 copies/ml for ACOV (2) and approximately 3,225 copies/ml when calculated based on the published genomes/reaction for IDNCOV (3). Based on the distribution of cycle numbers seen in Fig. 1 and performance agreement among the sites, negative results on IDNCOV are likely related to both a higher limit of detection on IDNCOV and preanalytical sampling error. FIG 1 Boxplot of cycle numbers of concordant and discordant paired results. Distribution of cycle numbers from IDNCOV-positive/ACOV-positive samples (including a single data point [CN 31.01] outlier beyond the standard error) compared to INDCOV-negative/ACOV-positive samples. Overall, the ID Now COVID-19 assay demonstrated significantly different performance characteristics compared to the Abbott RealTime SARS-CoV-2 assay.
                Bookmark

                Author and article information

                Journal
                Diagn Microbiol Infect Dis
                Diagn Microbiol Infect Dis
                Diagnostic Microbiology and Infectious Disease
                Published by Elsevier Inc.
                0732-8893
                1879-0070
                12 June 2022
                12 June 2022
                : 115742
                Affiliations
                [1 ]Microbiology unit, Begin teaching military hospital, Saint-Mandé, France
                [2 ]Emergency unit, Begin teaching military hospital, Saint-Mandé, France
                [3 ]Department of Infectious Diseases and Tropical Medicine, Begin teaching military hospital, Saint-Mandé, France
                Author notes
                [* ]Corresponding author: Aurore Bousquet, HIA Begin, 69 avenue de Paris 94160 Saint Mandé, France
                Article
                S0732-8893(22)00108-0 115742
                10.1016/j.diagmicrobio.2022.115742
                9188679
                80b3166f-a79a-4e7b-9f12-149118a8f765
                © 2022 Published by Elsevier Inc.

                Since January 2020 Elsevier has created a COVID-19 resource centre with free information in English and Mandarin on the novel coronavirus COVID-19. The COVID-19 resource centre is hosted on Elsevier Connect, the company's public news and information website. Elsevier hereby grants permission to make all its COVID-19-related research that is available on the COVID-19 resource centre - including this research content - immediately available in PubMed Central and other publicly funded repositories, such as the WHO COVID database with rights for unrestricted research re-use and analyses in any form or by any means with acknowledgement of the original source. These permissions are granted for free by Elsevier for as long as the COVID-19 resource centre remains active.

                History
                : 7 February 2022
                : 6 June 2022
                Categories
                Technical Note

                Microbiology & Virology
                sars-cov-2,covid-19,id now,hologic panther,point of care device
                Microbiology & Virology
                sars-cov-2, covid-19, id now, hologic panther, point of care device

                Comments

                Comment on this article