      Lumen-apposing metal stent-induced perforation during endoscopic ultrasound-directed transgastric retrograde cholangiopancreaticography managed with an endoscopic suturing device

      research-article
      , Dr., , MD, PhD
      Endoscopy
      Georg Thieme Verlag KG

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          A 37-year-old woman was referred to our unit because of typical biliary colic with transient cholestasis. Her relevant medical history included a cholecystectomy and a sleeve gastrectomy, which was subsequently converted to a gastric bypass. After multidisciplinary team (MDT) discussion, we opted for an endoscopic ultrasound-directed transgastric endoscopic retrograde cholangiopancreatography (ERCP; EDGE) using a 20/10 lumen-apposing metal stent (LAMS). Owing to the patient’s previous sleeve gastrectomy, the LAMS had to be placed between the jejunum and the excluded antrum. As per recently published international expert recommendations, the ERCP was scheduled 2 weeks post-LAMS placement to allow for tract maturation 1 . Nevertheless, the proximal flange of the LAMS dislocated during duodenoscope introduction and ended up in the peritoneal cavity ( Fig. 1 a,b ; Video 1 ). A decompressing needle was placed and the proximal flange was endoscopically retrieved from the peritoneal cavity under fluoroscopic guidance. After an unsuccessful attempt had been made to reposition the proximal flange in the jejunum ( Fig. 1 c ), a guidewire was placed through the LAMS lumen into the excluded antrum. A bridging attempt using a 22-mm × 10-cm fully covered enteral stent through the LAMS lumen also failed ( Fig. 1 d ). Subsequently, intraperitoneal endoscopic closure of the perforation in the excluded antrum was performed with the X-tack system (Boston Scientific, Massachusetts, USA) ( Fig. 1 e ). Next, we removed the LAMS and the enteral stent from the peritoneal cavity and completely sealed the fistula tract from the jejunal side with X-tack endosutures and clips ( Fig. 1 f,g ). Complete closure of the jejunal perforation was confirmed with contrast enterography ( Fig. 1 h ). Fig. 1 Endoscopic images showing: a the duodenoscope being introduced through the lumen-apposing metal stent (LAMS); b migration of the LAMS into the peritoneal cavity; c an unsuccessful attempt to reposition the proximal flange of the LAMS into the jejunum; d an unsuccessful attempt to bridge the excluded stomach and the alimentary jejunal limb with a covered enteral stent; e intraperitoneal endoscopic closure of the perforation in the excluded antrum using an endoscopic suturing system and one clip; f removal of the enteral stent and the LAMS; g closure of the jejunal perforation with the endoscopic suturing system and clips; h fluoroscopic image during contrast enterography confirming complete sealing of the jejunal perforation. Iatrogenic perforation resulting from a lumen-apposing metal stent that had migrated intraperitoneally during endoscopic ultrasound-directed transgastric endoscopic retrograde cholangiopancreatography is managed with a novel dedicated endoscopic suturing device. Video 1 The patient was put on antibiotics and was able to start oral feeding the day after the procedure. Other than moderate abdominal pain for the first 2 days after the procedure and a transient elevation of the inflammatory markers, her further clinical course was uneventful. After repeat MDT discussion, she was scheduled for a laparoscopy-assisted ERCP, during which her biliary disease was treated successfully. This case illustrates the usefulness of endoscopic suturing devices as minimally invasive alternatives to surgery for management of large iatrogenic perforations. Endoscopy_UCTN_Code_CPL_1AL_2AD

          Related collections

          Most cited references1

          • Record: found
          • Abstract: found
          • Article: not found

          International Consensus Recommendations for Safe Use of LAMS for On- and Off-Label Indications Using a Modified Delphi Process.

          The study aimed to develop international consensus recommendations on the safe use of lumen-apposing metal stents (LAMSs) for on- and off-label indications.
            Bookmark

            Author and article information

            Journal
            Endoscopy
            Endoscopy
            10.1055/s-00000012
            Endoscopy
            Endoscopy
            Georg Thieme Verlag KG (Rüdigerstraße 14, 70469 Stuttgart, Germany )
            0013-726X
            1438-8812
            07 August 2024
            December 2024
            1 August 2024
            : 56
            : Suppl 1
            : E679-E680
            Affiliations
            [1 ]Ringgold 60200, Gastroenterology, University Hospital Ghent, Ghent, Belgium;
            Author notes
            Correspondence Pieter Hindryckx, MD, PhD Department of Gastroenterology, University Hospital of Ghent Corneel Heymanslaan 10GhentBelgium pieter.hindryckx@ 123456uzgent.be
            Author information
            http://orcid.org/0000-0002-5949-2607
            Article
            E-Videos-2024-03-4912-EV
            10.1055/a-2362-0806
            11305860
            39111752
            727a9c2f-f0d5-4bb8-9649-182123254a66
            The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/).

            This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

            History
            Categories
            E-Videos

            Comments

            Comment on this article

            scite_
            0
            0
            0
            0
            Smart Citations
            0
            0
            0
            0
            Citing PublicationsSupportingMentioningContrasting
            View Citations

            See how this article has been cited at scite.ai

            scite shows how a scientific paper has been cited by providing the context of the citation, a classification describing whether it supports, mentions, or contrasts the cited claim, and a label indicating in which section the citation was made.

            Similar content2,127

            Most referenced authors22