      Sub-arc collimator angle optimization based on the conformity index heatmap for VMAT planning of multiple brain metastases SRS treatments

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Objective

          The aim of this study was to investigate the impact of collimator angle optimization in single-isocenter coplanar volume modulated arc therapy (VMAT) stereotactic radiosurgery (SRS) for multiple metastases with respect to dosimetric quality and treatment delivery efficiency. In particular, this is achieved by a novel algorithm of sub-arc collimator angle optimization (SACAO).

          Methods

          Twenty patients with multiple brain metastases were retrospectively included in this study. A multi-leaf collimator (MLC) conformity index (MCI) that is defined as the ratio of the area of target projection in the beam’s eye view (BEV) to the related area fitted by MLC was applied. Accordingly, for each control point, 180 MCI values were calculated with a collimator angle interval of 1°. A two-dimensional heatmap of MCI as a function of control point and collimator angle for each full arc was generated. The optimal segmentation of sub-arcs was achieved by avoiding the worst MCI at each control point. Then, the optimal collimator angle for each sub-arc would be determined by maximizing the summation of MCI. Each patient was scheduled to undergo single-center coplanar VMAT SRS based on either the novel SACAO algorithm or the conventional VMAT with static collimator angle (ST-VMAT). The dosimetric parameters, field sizes, and the monitoring units (Mus) were evaluated.

          Results

          The mean dose-volumetric parameters for the target volume of SACAO were comparable to ST-VMAT, while the conformity index (CI), homogeneity index (HI), and gradient index (GI) were reduced by SACAO. Improved sparing of organs at risk (OARs) was also obtained by SACAO. In particular, the SACAO method significantly ( p < 0.01) reduced the field size (76.59 ± 32.55 vs. 131.95 ± 56.71 cm 2) and MUs (655.35 ± 71.99 vs. 729.85 ± 73.52) by 41.11%.

          Conclusions

          The SACAO method could be superior in improving the CI, HI, and GI of the targets as well as normal tissue sparing for multiple brain metastases SRS. In particular, SACAO has the potential of increasing treatment efficiency in terms of field size and MU.

          Related collections

          Most cited references24

          • Record: found
          • Abstract: found
          • Article: not found

          Volumetric modulated arc therapy: IMRT in a single gantry arc

          Karl Otto (2008)
          In this work a novel plan optimization platform is presented where treatment is delivered efficiently and accurately in a single dynamically modulated arc. Improvements in patient care achieved through image-guided positioning and plan adaptation have resulted in an increase in overall treatment times. Intensity-modulated radiation therapy (IMRT) has also increased treatment time by requiring a larger number of beam directions, increased monitor units (MU), and, in the case of tomotherapy, a slice-by-slice delivery. In order to maintain a similar level of patient throughput it will be necessary to increase the efficiency of treatment delivery. The solution proposed here is a novel aperture-based algorithm for treatment plan optimization where dose is delivered during a single gantry arc of up to 360 deg. The technique is similar to tomotherapy in that a full 360 deg of beam directions are available for optimization but is fundamentally different in that the entire dose volume is delivered in a single source rotation. The new technique is referred to as volumetric modulated arc therapy (VMAT). Multileaf collimator (MLC) leaf motion and number of MU per degree of gantry rotation is restricted during the optimization so that gantry rotation speed, leaf translation speed, and dose rate maxima do not excessively limit the delivery efficiency. During planning, investigators model continuous gantry motion by a coarse sampling of static gantry positions and fluence maps or MLC aperture shapes. The technique presented here is unique in that gantry and MLC position sampling is progressively increased throughout the optimization. Using the full gantry range will theoretically provide increased flexibility in generating highly conformal treatment plans. In practice, the additional flexibility is somewhat negated by the additional constraints placed on the amount of MLC leaf motion between gantry samples. A series of studies are performed that characterize the relationship between gantry and MLC sampling, dose modeling accuracy, and optimization time. Results show that gantry angle and MLC sample spacing as low as 1 deg and 0.5 cm, respectively, is desirable for accurate dose modeling. It is also shown that reducing the sample spacing dramatically reduces the ability of the optimization to arrive at a solution. The competing benefits of having small and large sample spacing are mutually realized using the progressive sampling technique described here. Preliminary results show that plans generated with VMAT optimization exhibit dose distributions equivalent or superior to static gantry IMRT. Timing studies have shown that the VMAT technique is well suited for on-line verification and adaptation with delivery times that are reduced to approximately 1.5-3 min for a 200 cGy fraction.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            A simple dose gradient measurement tool to complement the conformity index.

            A dose gradient index (GI) is proposed that can be used to compare treatment plans of equal conformity. The steep dose gradient outside the radiosurgical target is one of the factors that makes radiosurgery possible. It therefore makes sense to measure this variable and to use it to compare rival plans, explore optimal prescription isodoses, or compare treatment modalities. The GI is defined as the ratio of the volume of half the prescription isodose to the volume of the prescription isodose. For a plan normalized to the 50% isodose line, it is the ratio of the 25% isodose volume to that of the 50% isodose volume. The GI will differentiate between plans of similar conformity, but with different dose gradients, for example, where isocenters have been inappropriately centered on the edge of the target volume. In a retrospective series of 50 dose plans for the treatment of vestibular schwannoma, the optimal prescription isodose was assessed. A mean value of 40% (median 38%, range 30-61%) was calculated, not 50% as might be anticipated. The GI can show which of these prescription isodoses will give the steepest dose falloff outside the target. When planning a multiisocenter treatment, there may be a temptation to place some isocenters on the edge of the target. This has the apparent advantage of producing a plan of good conformity and a predictable prescription isodose; however, it risks creating a plan that has a low dose gradient outside the target. The quality of this dose gradient is quantified by the GI.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT: plan quality, delivery efficiency and accuracy.

              Helical tomotherapy (HT) and volumetric modulated arc therapy (VMAT) are arc-based approaches to IMRT delivery. The objective of this study is to compare VMAT to both HT and fixed field IMRT in terms of plan quality, delivery efficiency, and accuracy. Eighteen cases including six prostate, six head-and-neck, and six lung cases were selected for this study. IMRT plans were developed using direct machine parameter optimization in the Pinnacle3 treatment planning system. HT plans were developed using a Hi-Art II planning station. VMAT plans were generated using both the Pinnacle3 SmartArc IMRT module and a home-grown arc sequencing algorithm. VMAT and HT plans were delivered using Elekta's PreciseBeam VMAT linac control system (Elekta AB, Stockholm, Sweden) and a TomoTherapy Hi-Art II system (TomoTherapy Inc., Madison, WI), respectively. Treatment plan quality assurance (QA) for VMAT was performed using the IBA MatriXX system while an ion chamber and films were used for HT plan QA. The results demonstrate that both VMAT and HT are capable of providing more uniform target doses and improved normal tissue sparing as compared with fixed field IMRT. In terms of delivery efficiency, VMAT plan deliveries on average took 2.2 min for prostate and lung cases and 4.6 min for head-and-neck cases. These values increased to 4.7 and 7.0 min for HT plans. Both VMAT and HT plans can be delivered accurately based on their own QA standards. Overall, VMAT was able to provide approximately a 40% reduction in treatment time while maintaining comparable plan quality to that of HT.
                Bookmark

                Author and article information

                Contributors
                Journal
                Front Oncol
                Front Oncol
                Front. Oncol.
                Frontiers in Oncology
                Frontiers Media S.A.
                2234-943X
                06 September 2022
                2022
                : 12
                : 987971
                Affiliations
                [1] 1 Department of Radiation and Medical Oncology, Zhongnan Hospital of Wuhan University , Wuhan, China
                [2] 2 Hubei Radiotherapy Quality Control Center, Wuhan University , Wuhan, China
                [3] 3 National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital and Shenzhen Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College , Shenzhen, China
                Author notes

                Edited by: Jason W. Sohn, Allegheny Health Network, United States

                Reviewed by: Manju Sharma, University of California, San Francisco, United States; Maria F. Chan, Memorial Sloan Kettering Cancer Center, United States

                *Correspondence: Xiaoyong Wang, wangxiaoyong@ 123456znhospital.cn ; Zhitao Dai, daizt_sinap@ 123456163.com

                †These authors have contributed equally to this work

                This article was submitted to Radiation Oncology, a section of the journal Frontiers in Oncology

                Article
                10.3389/fonc.2022.987971
                9487306
                36147903
                7255a9ca-624b-4137-8767-a2dd057f1176
                Copyright © 2022 Shen, Dai, Yu, Yuan, Kang, Chen, Liu, Xie and Wang

                This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

                History
                : 06 July 2022
                : 11 August 2022
                Page count
                Figures: 5, Tables: 4, Equations: 5, References: 24, Pages: 10, Words: 4847
                Funding
                Funded by: Sanming Project of Medicine in Shenzhen , doi 10.13039/501100012151;
                Funded by: Basic and Applied Basic Research Foundation of Guangdong Province , doi 10.13039/501100021171;
                Categories
                Oncology
                Original Research

                Oncology & Radiotherapy
                collimator angle optimization,volumetric modulated arc therapy,stereotactic radiosurgery,multiple brain metastases,sub-arc

                Comments

                Comment on this article