      Comparison of an imaging heel quantitative ultrasound device (DTU-one) with densitometric and ultrasonic measurements.

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The purpose of this study was to evaluate a new imaging ultrasound scanner for the heel, the DTU-one (Osteometer MediTech, Denmark), by comparing quantitative ultrasound (QUS) results with bone mineral density (BMD) of the heel and femur from dual X-ray absorptiometry (DXA), and by comparing the DTU-one with another QUS device, the UBA 575+. The regions of interest in the DXA heel scan were matched with the regions evaluated by the two QUS devices. 134 healthy and 16 osteoporotic women aged 30-84 years old were enrolled in the study. In vivo short-term precision of the DTU-one for broadband ultrasound attenuation (BUA) and speed of sound (SOS) was 2.9% and 0.1%, respectively, and long-term precision was 3.8% and 0.2%, respectively. Highest correlations (r) between QUS and BMD measurements were achieved when comparing DTU-one results with BMD in matched regions of the DXA heel scan. Correlation coefficients (r) were 0.81 for BUA and SOS. Highest correlations with the UBA 575+ were 0.68 and 0.72, respectively. The comparison of BMD in different femoral sites with BUA and SOS (DTU-one) varied from 0.62 to 0.69 when including the entire study population. The correlation between BMD values within different sites of the femur tended to be higher (from r = 0.81 to 0.93). When comparing BUA with BUA and SOS with SOS on the two QUS devices, the absolute QUS values differed significantly. However, correlations were relatively high, with 0.76 for BUA and 0.82 for SOS. In conclusion, the results of the new quantitative ultrasound device, the DTU-one, are highly correlated (r = 0.8) with results obtained using the UBA 575+ and with BMD in the heel. The precision of the DTU-one is comparable to other QUS devices for BUA and is high for SOS.

          Related collections

          Author and article information

          Journal
          Br J Radiol
          The British journal of radiology
          British Institute of Radiology
          0007-1285
          0007-1285
          Jan 2000
          : 73
          : 865
          Affiliations
          [1 ] Osteoporosis and Arthritis Research Group, University of California at San Francisco 94117, USA.
          Article
          10.1259/bjr.73.865.10721316
          10721316
          501ca41c-ad97-476c-8852-7c5d9591c1a2
          History

          Non-programmatic
          Non-programmatic

          Comments

          Comment on this article