      Evaluation of a combined respiratory‐gating system comprising the TrueBeam linear accelerator and a new real‐time tumor‐tracking radiotherapy system: a preliminary study

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          A combined system comprising the TrueBeam linear accelerator and a new real‐time, tumor‐tracking radiotherapy system, SyncTraX, was installed in our institution. The goals of this study were to assess the capability of SyncTraX in measuring the position of a fiducial marker using color fluoroscopic images, and to evaluate the dosimetric and geometric accuracy of respiratory‐gated radiotherapy using this combined system for the simple geometry. For the fundamental evaluation of respiratory‐gated radiotherapy using SyncTraX, the following were performed: 1) determination of dosimetric and positional characteristics of sinusoidal patterns using a motor‐driven base for several gating windows; 2) measurement of time delay using an oscilloscope; 3) positional verification of sinusoidal patterns and the pattern in the case of a lung cancer patient; 4) measurement of the half‐value layer (HVL in mm AL), effective kVp, and air kerma, using a solid‐state detector for each fluoroscopic condition, to determine the patient dose. The dose profile in a moving phantom with gated radiotherapy having a gating window 4mm was in good agreement with that under static conditions for each photon beam. The total time delay between TrueBeam and SyncTraX was <227ms for each photon beam. The mean of the positional tracking error was <0.4mm for sinusoidal patterns and for the pattern in the case of a lung cancer patient. The air‐kerma rates from one fluoroscopy direction were 1.93±0.01, 2.86±0.01, 3.92±0.04, 5.28±0.03, and 6.60±0.05mGy/min for 70, 80, 90, 100, and 110 kV X‐ray beams at 80 mA, respectively. The combined system comprising TrueBeam and SyncTraX could track the motion of the fiducial marker and control radiation delivery with reasonable accuracy; therefore, this system provides significant dosimetric improvement. However, patient exposure dose from fluoroscopy was not clinically negligible.

          PACS number(s): 87.53.Bn, 87.55.km, 87.55.Qr

          Related collections

          Most cited references25

          • Record: found
          • Abstract: found
          • Article: not found

          Task Group 142 report: quality assurance of medical accelerators.

          The task group (TG) for quality assurance of medical accelerators was constituted by the American Association of Physicists in Medicine's Science Council under the direction of the Radiation Therapy Committee and the Quality Assurance and Outcome Improvement Subcommittee. The task group (TG-142) had two main charges. First to update, as needed, recommendations of Table II of the AAPM TG-40 report on quality assurance and second, to add recommendations for asymmetric jaws, multileaf collimation (MLC), and dynamic/virtual wedges. The TG accomplished the update to TG-40, specifying new test and tolerances, and has added recommendations for not only the new ancillary delivery technologies but also for imaging devices that are part of the linear accelerator. The imaging devices include x-ray imaging, photon portal imaging, and cone-beam CT. The TG report was designed to account for the types of treatments delivered with the particular machine. For example, machines that are used for radiosurgery treatments or intensity-modulated radiotherapy (IMRT) require different tests and/or tolerances. There are specific recommendations for MLC quality assurance for machines performing IMRT. The report also gives recommendations as to action levels for the physicists to implement particular actions, whether they are inspection, scheduled action, or immediate and corrective action. The report is geared to be flexible for the physicist to customize the QA program depending on clinical utility. There are specific tables according to daily, monthly, and annual reviews, along with unique tables for wedge systems, MLC, and imaging checks. The report also gives specific recommendations regarding setup of a QA program by the physicist in regards to building a QA team, establishing procedures, training of personnel, documentation, and end-to-end system checks. The tabulated items of this report have been considerably expanded as compared with the original TG-40 report and the recommended tolerances accommodate differences in the intended use of the machine functionality (non-IMRT, IMRT, and stereotactic delivery).
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Organ motion and its management.

            To compile and review data on the topic of organ motion and its management. Data were classified into three categories: (a) patient position-related organ motion, (b) interfraction organ motion, and (c) intrafraction organ motion. Data on interfraction motion of gynecological tumors, the prostate, bladder, and rectum are reviewed. Literature pertaining to the intrafraction movement of the liver, diaphragm, kidneys, pancreas, lung tumors, and prostate is compiled. Methods for managing interfraction and intrafraction organ motion in radiation therapy are also reviewed.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Physical aspects of a real-time tumor-tracking system for gated radiotherapy.

              To reduce uncertainty due to setup error and organ motion during radiotherapy of tumors in or near the lung, by means of real-time tumor tracking and gating of a linear accelerator. The real-time tumor-tracking system consists of four sets of diagnostic X-ray television systems (two of which offer an unobstructed view of the patient at any time), an image processor unit, a gating control unit, and an image display unit. The system recognizes the position of a 2.0-mm gold marker in the human body 30 times per second using two X-ray television systems. The marker is inserted in or near the tumor using image guided implantation. The linear accelerator is gated to irradiate the tumor only when the marker is within a given tolerance from its planned coordinates relative to the isocenter. The accuracy of the system and the additional dose due to the diagnostic X-ray were examined in a phantom, and the geometric performance of the system was evaluated in 4 patients. The phantom experiment demonstrated that the geometric accuracy of the tumor-tracking system is better than 1.5 mm for moving targets up to a speed of 40 mm/s. The dose due to the diagnostic X-ray monitoring ranged from 0.01% to 1% of the target dose for a 2.0-Gy irradiation of a chest phantom. In 4 patients with lung cancer, the range of the coordinates of the tumor marker during irradiation was 2.5-5.3 mm, which would have been 9.6-38.4 mm without tracking. We successfully implemented and applied a tumor-tracking and gating system. The system significantly improves the accuracy of irradiation of targets in motion at the expense of an acceptable amount of diagnostic X-ray exposure.
                Bookmark

                Author and article information

                Contributors
                shiinoki@yamaguchi-u.ac.jp
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                08 July 2016
                July 2016
                : 17
                : 4 ( doiID: 10.1002/acm2.2016.17.issue-4 )
                : 202-213
                Affiliations
                [ 1 ] Department of Radiation Oncology Graduate School of Medicine, Yamaguchi University Yamaguchi Japan
                [ 2 ] Department of Radiological Technology Yamaguchi University Hospital Yamaguchi Japan
                Author notes
                [*] [* ]Corresponding author: Takehiro Shiinoki, Department of Radiation Oncology, Yamaguchi University, Graduate school of Medicine, 1–1‐1 Minamikogushi, Ube, Yamaguchi 755–8505, Japan; phone: (0836) 22 2966; fax: (0836) 22 2961; email: shiinoki@ 123456yamaguchi-u.ac.jp
                Article
                ACM20202
                10.1120/jacmp.v17i4.6114
                5690064
                27455483
                4d71158c-9be9-406e-b58b-7cf5d7257ad9
                © 2016 The Authors.

                This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 06 October 2015
                : 10 March 2016
                Page count
                Figures: 7, Tables: 3, References: 27, Pages: 12, Words: 5269
                Funding
                Funded by: Japan Society for the Promotion of Science (JSPS)
                Award ID: 15K21194
                Categories
                Radiation Oncology Physics
                Radiation Oncology Physics
                Custom metadata
                2.0
                acm20202
                July 2016
                Converter:WILEY_ML3GV2_TO_NLMPMC version:5.2.5 mode:remove_FC converted:16.11.2017

                respiratory motion,real‐time tumor‐tracking,respiratory‐gated radiotherapy

                Comments

                Comment on this article

                scite_
                0
                0
                0
                0
                Smart Citations
                0
                0
                0
                0
                Citing PublicationsSupportingMentioningContrasting
                View Citations

                See how this article has been cited at scite.ai

                scite shows how a scientific paper has been cited by providing the context of the citation, a classification describing whether it supports, mentions, or contrasts the cited claim, and a label indicating in which section the citation was made.

                Similar content176

                Cited by12

                Most referenced authors301