      ARTEMIS, THE 14C AMS FACILITY OF THE LMC14 NATIONAL LABORATORY: A STATUS REPORT ON QUALITY CONTROL AND MICROSAMPLE PROCEDURES

      Read this article at

      ScienceOpenPublisher
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          ABSTRACT

          Quality control procedures have been developed at the Laboratoire de Mesure du Carbone 14 (LMC14) national laboratory throughout the years of operation. Routine procedures are applied to sample preparation depending on their composition and their size. The tuning of the ARTEMIS AMS facility, hosted by the LMC14 laboratory, uses an accurate procedure. A batch of unknown samples is measured with accompanying samples (primary and secondary standards and blanks), which give a powerful set of data to control the quality of each measurement. A homemade database has been created to store the sample information and study the evolution of the accompanying samples. The LMC14 laboratory participated in the Sixth International Radiocarbon Intercomparison, SIRI. The results are presented here, with statistical tests to assess the quality of the preparations and measurements done at the LMC14 national laboratory. To obtain a reliable radiocarbon ( 14C) age by AMS, 1 mg of sample is required in routine analysis. Recently, the LMC14 developed a new procedure dedicated to microsamples, allowing the size of samples to be reduced and contributing to opening 14C dating to materials that were previously unreachable. This new procedure has been successfully tested on valuable Cultural Heritage samples: lead white mural paintings.

          Related collections

          Most cited references2

          • Record: found
          • Abstract: found
          • Article: not found

          Artemis, the New 14C AMS at LMC14 in Saclay, France

          The new facility Artemis was installed in 2003 in Saclay, France. This 3MV NEC Pelletron is dedicated to high-precision radiocarbon measurements for French 14 C laboratories. We will present information on Artemis along with our sample preparation methods. Results from measurements on some intercalibration samples will be given along with the values of measured blanks. Finally, we report on some problems we have encountered when measuring sputter cathodes with high CH − outputs.
            Bookmark
            • Record: found
            • Abstract: not found
            • Book: not found

            Artists’ Pigments: a handbook of their history and characteristics

              Bookmark

              Author and article information

              Contributors
              (View ORCID Profile)
              (View ORCID Profile)
              Journal
              Radiocarbon
              Radiocarbon
              Cambridge University Press (CUP)
              0033-8222
              1945-5755
              December 2020
              August 25 2020
              December 2020
              : 62
              : 6
              : 1755-1770
              Article
              10.1017/RDC.2020.73
              47a042bc-0832-48e3-90a6-c6b2f7bd8030
              © 2020

              https://www.cambridge.org/core/terms

              History

              Comments

              Comment on this article