      Acute aortic dissection as a late and fatal complication of transcatheter persistent ductus arteriosus occlusion: a case report

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background 

          Persistent ductus arteriosus (PDA) is a common congenital heart defect, of which, the preferred treatment in children and adults is transcatheter occlusion. This report describes the first reported case of acute aortic dissection as a late complication of transcatheter PDA occlusion.

          Case summary 

          A 66-year-old healthy woman, with a history of transcatheter PDA closure 12 years previously, died suddenly at home. Post-mortem revealed pericardial tamponade complicating an acute aortic dissection, with a large intimal tear identified adjacent to the PDA occluder.

          Discussion 

          Transcatheter occlusion is accepted as a safe and effective treatment in older children and adults, important in preventing haemodynamic sequelae such as heart failure and pulmonary hypertension. Complications are uncommon and mostly acute, including device embolization and haemolysis. Aortic dissection has been reported as an acute complication but only in the setting of incorrect device deployment. This is the first description of aortic dissection as a late complication of transcatheter PDA occlusion and although rare has important ramifications given it is a potentially fatal outcome of a common procedure.

          Related collections

          Most cited references6

          • Record: found
          • Abstract: found
          • Article: not found

          The incidence of congenital heart disease.

          This study was designed to determine the reasons for the variability of the incidence of congenital heart disease (CHD), estimate its true value and provide data about the incidence of specific major forms of CHD. The incidence of CHD in different studies varies from about 4/1,000 to 50/1,000 live births. The relative frequency of different major forms of CHD also differs greatly from study to study. In addition, another 20/1,000 live births have bicuspid aortic valves, isolated anomalous lobar pulmonary veins or a silent patent ductus arteriosus. The incidences reported in 62 studies published after 1955 were examined. Attention was paid to the ways in which the studies were conducted, with special reference to the increased use of echocardiography in the neonatal nursery. The total incidence of CHD was related to the relative frequency of ventricular septal defects (VSDs), the most common type of CHD. The incidences of individual major forms of CHD were determined from 44 studies. The incidence of CHD depends primarily on the number of small VSDs included in the series, and this number in turn depends upon how early the diagnosis is made. If major forms of CHD are stratified into trivial, moderate and severe categories, the variation in incidence depends mainly on the number of trivial lesions included. The incidence of moderate and severe forms of CHD is about 6/1,000 live births (19/1,000 live births if the potentially serious bicuspid aortic valve is included), and of all forms increases to 75/1,000 live births if tiny muscular VSDs present at birth and other trivial lesions are included. Given the causes of variation, there is no evidence for differences in incidence in different countries or times.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Multicenter USA Amplatzer patent ductus arteriosus occlusion device trial: initial and one-year results.

            We sought to review and report initial and one-year efficacy and safety results of the multicenter USA Amplatzer ductal occluder (ADO) device trial. Transcatheter closure of a moderate to large patent ductus arteriosus (PDA) using conventional techniques is challenging. The ADO can close a PDA up to 12 mm in diameter. From September 1999 to June 2002, 484 patients were enrolled in 25 U.S. centers. Forty-five (9%) of 484 patients did not have ADO implantation, because the PDA was too small or because of elevated pulmonary resistance. The median age of the patients at catheterization was 1.8 years (range 0.2 to 70.7 years), and weight was 11 kg (range 4.5 to 164.5 kg). The median PDA minimal diameter was 2.6 mm (range 0.9 to 11.2 mm); 76 (17%) of 439 were larger than 4.0 mm. Median pulmonary artery mean pressure was 20 mm Hg (range 7 to 80 mm Hg). The ADO was implanted successfully in 435 (99%) of 439 patients, with a median fluoroscopy time of 7.1 min (range 2.9 to 138.4 min). Angiographic demonstration of occlusion was seen in 329 (76%) of 435. This increased to 384 (89%) of 433 on post-catheterization day 1, with occlusion documented in 359 (99.7%) of 360 at one year. At the last evaluation in all patients at any time, PDA closure was documented in 428 (98%) of 435 patients. There have been two cases of partial left pulmonary artery occlusion after ADO implantation and no cases of significant aortic obstruction. Moderate to large PDAs can be effectively and safely closed using the ADO device, with excellent initial and one-year results. This device should obviate the need for multiple coils or surgical intervention for these defects.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Trial occlusion to assess the risk of persistent pulmonary arterial hypertension after closure of a large patent ductus arteriosus in adolescents and adults with elevated pulmonary artery pressure.

              No method is available to predict whether patients with patent ductus arteriosus (PDA) and severe pulmonary arterial hypertension (PAH) will show persistent postprocedural PAH (PP-PAH) after PDA closure. This study evaluated the usefulness of trial occlusion for predicting PP-PAH after transcatheter PDA closure in patients with severe PAH.
                Bookmark

                Author and article information

                Contributors
                Role: Handling Editor
                Role: Editor
                Role: Editor
                Role: Editor
                Role: Editor
                Journal
                Eur Heart J Case Rep
                Eur Heart J Case Rep
                ehjcr
                European Heart Journal: Case Reports
                Oxford University Press
                2514-2119
                January 2021
                11 December 2020
                11 December 2020
                : 5
                : 1
                : ytaa484
                Affiliations
                [1 ] Department of Cardiology, Royal Prince Alfred Hospital , Missenden Road, Camperdown, Sydney, NSW, Australia
                [2 ] Sydney Medical School, University of Sydney , Sydney, NSW, Australia
                Author notes
                Corresponding author. Tel: +612 9515 7110, Fax: +612 9550 6262, Email: David.Celermajer@ 123456Sydney.edu.au
                Author information
                http://orcid.org/0000-0001-7349-7554
                Article
                ytaa484
                10.1093/ehjcr/ytaa484
                7850608
                4676f230-3d7f-486f-a4e4-432e027ecc65
                © The Author(s) 2020. Published by Oxford University Press on behalf of the European Society of Cardiology.

                This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License ( http://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com

                History
                : 23 June 2020
                : 24 August 2020
                : 12 November 2020
                Page count
                Pages: 3
                Categories
                Case Report
                AcademicSubjects/MED00200

                persistent ductus arteriosus,device occlusion,complication,aortic dissection,case report

                Comments

                Comment on this article