      Collision risk mitigation of Varian TrueBeam linear accelerator with supplemental live-view cameras

      , , ,
      Practical Radiation Oncology
      Elsevier BV

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Noncoplanar radiation therapy techniques such as 4π have potential dosimetric advantages but introduce complexities in treatment delivery that increase the risk for collision. Direct or remote visual confirmation of clearance is a safeguard against collisions of the gantry, couch, and patient. With our institution's Varian TrueBeam system, we identified configurations that cannot be visualized with the included closed-circuit television cameras. At our practice, electronic, portal imaging device (EPID) collision risk also exists because of the routine deployment to capture exit-dose images for treatment quality assurance. We propose a simple, cost-effective solution using network cameras to help eliminate blind spots that permits safe, noncoplanar arrangements with an EPID-acquired exit dose.