          INTRODUCTION: Reports of fatigue failure of the femoral component of a total knee arthroplasty (TKA) is scanty in the literature. As a result, there are no clearly defined risk factors to aid us in predicting fatigue failure of an implant. Furthermore, these patients may present with non-specific knee pain, which may or may not be well tolerated, depending on the stability of the implant. We report a case of fatigue failure of a poorly cemented femoral component of a TKA in a 72-year-old female, approximately seven years after the initial surgery CASE REPORT: A 72-year-old female presented to our tertiary level arthroplasty unit with new-onset knee pain approximately seven years after undergoing a TKA at our unit. She reported hearing a crack six months earlier, while standing up from a seated position. She had initially presented to her local clinic, but the pathology was missed. She received revision surgery at our institution and was doing well at early follow-up. DISCUSSION: We reviewed the literature on fatigue failure of femoral components in TKA in an attempt to define risk factors. We also summarised all cases of femoral component fatigue failure in the English literature. CONCLUSION: Although femoral component fatigue failure in TKA is rare, the majority of cases have attributed the failure to poor surgical technique. Despite this, certain implants have been failing more often than others, and proposed mechanisms for this exist. Orthopaedic surgeons need to be aware of which implant designs are prone to failure, as well as how meticulous surgical technique can reduce the chances of fatigue failure. Level of evidence: Level 5

          Related collections

          Most cited references15

          • Record: found
          • Abstract: found
          • Article: not found

          Rationale of the Knee Society clinical rating system.

          A new total knee rating system has been developed by The Knee Society to provide an up-to-date more stringent evaluation form. The system is subdivided into a knee score that rates only the knee joint itself and a functional score that rates the patient's ability to walk and climb stairs. The dual rating system eliminates the problem of declining knee scores associated with patient infirmity.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Cementing the femoral component in total knee arthroplasty: which technique is the best?

            Although several techniques exist for cementing the femoral component in TKA, no data are available on which is the best one to use. We therefore compared four cementing techniques in an anatomical open pore sawbone model (n=20), in order to investigate the influence of cementation technique on overall cement penetration as well as length of the cement mantle over the different cuts. The technique which included cement application onto the anterior and distal bone surfaces, as well as the posterior flanges of the prosthesis, was statistically superior to the other techniques. We therefore advocate this technique as the standard for cementing the femoral component.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Fracture of the femoral component associated with polyethylene wear and osteolysis after total knee arthroplasty.

              Fracture of the femoral component associated with polyethylene wear and osteolysis after total knee arthroplasty (TKA) has not been well reported before. A 63-year-old man with osteoarthritis of the right knee underwent TKA with a New Jersey LCS Knee, with cementing on the tibia and patella but not on the femoral component. After 42 months, in addition to wearing of polyethylene of the tibia and patella, severe osteonecrosis of the medial femoral condyle was noted. Osteonecrosis caused loss of osseous support of the medial flange of the femoral component, and the bone ingrowth of the central and lateral flange to the distal femur was so good that it overcame the yield stress of the metal of the femoral component and caused fracture of the femoral component. The osteolytic area was filled with autogenous iliac bone, and a new femoral component was inserted and cemented. The patient's condition became satisfactory with relief of pain. Although uncommon, fracture of the femoral component does occur associated with polyethylene wear and osteolysis.
                Bookmark

                Author and article information

                Journal
                saoj
                SA Orthopaedic Journal
                SA orthop. j.
                Medpharm Publications (Centurion, Gauteng, South Africa )
                1681-150X
                2309-8309
                December 2020
                : 19
                : 4
                : 235-238
                Affiliations
                [08] Johannesburg orgnameUniversity of the Witwatersrand South Africa
                [03] orgnameChris Hani Baragwanath Academic Hospital orgdiv1Department of Orthopaedic Surgery orgdiv2Consultant, Arthroplasty Unit
                [01] orgnameChris Hani Baragwanath Academic Hospital orgdiv1Department of Orthopaedic Surgery
                [05] orgnameChris Hani Baragwanath Academic Hospital orgdiv1Department of Orthopaedic Surgery orgdiv2Head of Arthroplasty Unit
                [07] orgnameChris Hani Baragwanath Academic Hospital orgdiv1Department of Orthopaedic Surgery orgdiv2Consultant, Arthroplasty Unit
                [02] Johannesburg orgnameUniversity of the Witwatersrand South Africa
                [04] Johannesburg orgnameUniversity of the Witwatersrand South Africa
                [06] Johannesburg orgnameUniversity of the Witwatersrand South Africa
                Article
                S1681-150X2020000400008 S1681-150X(20)01900400008
                10.17159/2309-8309/2020/v19n4a7
                361006a4-5fe3-4c66-abb2-388a6c5c4863

                This work is licensed under a Creative Commons Attribution 4.0 International License.

                History
                : June 2020
                : June 2019
                Page count
                Figures: 0, Tables: 0, Equations: 0, References: 15, Pages: 4
                Product

                SciELO South Africa

                Categories
                Arthoplasty

                fatigue failure,stress fracture,total knee arthroplasty,femoral component

                Comments

                Comment on this article