      Failure of the rotating-hinge knee megaprosthesis.

      The Journal of Arthroplasty
      Arthroplasty, Replacement, Knee, adverse effects, Female, Humans, Knee Joint, surgery, Knee Prosthesis, Middle Aged, Prosthesis Design, Prosthesis Failure

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Recently, rotating hinge knee prostheses were applied more frequently due to improving modern implant designs. They are predominantly used in specific conditions with major bone defect or insufficiency of the collateral ligaments around the knee, often as salvage procedures. A case of rotating hinge knee megaprosthesis failure due to isolated tibial polyethylene stopper broken, which was never reported before, was investigated and treated in our institution. We suggested that rotating hinge knee prosthesis with incompetent medial collateral ligament is apt to failure due to the high valgus moment during gait. Sacrificing lateral collateral ligament or cutting the femur in slightly less than the normal 5° to 7° valgus may eliminate the risk of complication. Copyright © 2013 Elsevier Inc. All rights reserved.

          Related collections

          Author and article information

          Journal
          23146584
          10.1016/j.arth.2012.07.030

          Chemistry
          Arthroplasty, Replacement, Knee,adverse effects,Female,Humans,Knee Joint,surgery,Knee Prosthesis,Middle Aged,Prosthesis Design,Prosthesis Failure