      Comparative study of a novel proximal femoral bionic nail and three conventional cephalomedullary nails for reverse obliquity intertrochanteric fractures: a finite element analysis

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose

          Conventional cephalomedullary nails (CMNs) are commonly employed for internal fixation in the treatment of reverse obliquity intertrochanteric (ROI) fractures. However, the limited effectiveness of conventional CMNs in addressing ROI fractures results in significant implant-related complications. To address challenges associated with internal fixation, a novel Proximal Femoral Bionic Nail (PFBN) has been developed.

          Methods

          In this study, a finite element model was constructed using a normal femoral specimen, and biomechanical verification was conducted using the GOM non-contact optical strain measurement system. Four intramedullary fixation approaches—PFBN, Proximal Femoral Nail Antirotation InterTan nail (ITN), and Gamma nail (Gamma nail)—were employed to address three variations of ROI fractures (AO/OTA 31-A3). The biomechanical stability of the implant models was evaluated through the calculation of the von Mises stress contact pressure and displacement.

          Results

          Compared to conventional CMNs, the PFBN group demonstrated a 9.36%–59.32% reduction in the maximum VMS at the implant. The A3.3 ROI fracture (75% bone density) was the most unstable type of fracture. In comparison to conventional CMNs, PFBN demonstrated more stable data, including VMS values (implant: 506.33 MPa, proximal fracture fragment: 34.41 MPa), contact pressure (13.28 MPa), and displacement (17.59 mm).

          Conclusion

          Compared to the PFNA, ITN, and GN, the PFBN exhibits improvements in stress concentration, stress conduction, and overall model stability in ROI fractures. The double triangle structure aligns better with the tissue structure and biomechanical properties of the proximal femur. Consequently, the PFBN has significant potential as a new fixation strategy for the clinical treatment of ROI fractures.

          Related collections

          Most cited references32

          • Record: found
          • Abstract: not found
          • Article: not found

          Fracture and Dislocation Classification Compendium—2018

            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Reverse obliquity fractures of the intertrochanteric region of the femur.

            The reverse obliquity fracture of the proximal part of the femur is a distinct fracture pattern that is mechanically different from most intertrochanteric fractures. The purpose of this retrospective study was to determine the prevalence of these fractures and the results and complications of different types of internal fixation used in their treatment. Between 1988 and 1998, 2472 consecutive patients with a hip fracture were treated at our Level-One Trauma Center; 1035 of the fractures were classified as intertrochanteric or subtrochanteric. Clinical and radiographic records were retrospectively reviewed, and fifty-five fractures with a reverse obliquity pattern were identified. Forty-nine patients were followed until the fracture united or a revision operation was performed. The duration of clinical follow-up averaged eighteen months (range, three to sixty-seven months), and the duration of radiographic follow-up averaged fifteen months (range, three to sixty months). Fractures were classified with the Orthopaedic Trauma Association scheme. Results were analyzed according to the fracture pattern, type of implant, quality of the reduction, position of the implant, and use of bone graft at the index operation. Function was assessed on the basis of pain, living situation, need for walking aids, need for analgesics, and walking capacity. Thirty-two (68%) of forty-seven hips treated with internal fixation healed without an additional operation. Fifteen (32%) of the forty-seven failed to heal or had a failure of fixation. The failure rate was nine of sixteen for the sliding hip screws, two of fifteen for the blade-plates, three of ten for the dynamic condylar screws, one of three for the cephalomedullary nails, and zero of three for the intramedullary hip screws. Use of the fixed-angle devices (the blade-plate and the dynamic condylar screw) resulted in fewer failures than did use of the sliding hip screw (p = 0.023). Eleven (46%) of twenty-four nonanatomically reduced fractures and four (17%) of twenty-three anatomically reduced fractures had a failure of treatment (p = 0.060). Eleven (26%) of forty-two fractures with an ideally placed implant and four (80%) of five fractures with a non-ideally placed implant had a failure of treatment (p = 0.023). Of the fifteen fractures that failed to heal or had a failure of fixation, five were treated with revision to a calcar-replacement prosthesis, seven were treated with revision open reduction and internal fixation with bone-grafting, and one was treated with bone-grafting without revision of the fixation. Two patients refused additional surgery because they had limited functional demands. The two-year mortality rate was 33%. Functional results were poor, with many patients requiring walking aids and losing the capacity for independent walking and self-care. In this series, reverse obliquity fractures accounted for 2% of all hip fractures and 5% of all intertrochanteric and subtrochanteric fractures. Ninety-five-degree fixed-angle internal fixation devices performed significantly better than did sliding hip screws. Results were also worse for fractures with poor reduction and those with a poorly placed implant.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Treatment of unstable trochanteric fractures. Randomised comparison of the gamma nail and the proximal femoral nail.

              The proximal femoral nail (PFN) is a recently introduced intramedullary system, designed to improve treatment of unstable trochanteric fractures of the hip. In a multicentre prospective clinical study, the intra-operative use, complications and outcome of treatment using the PFN (n = 211) were compared with those using the gamma nail (GN) (n = 213). The intra-operative blood loss was lower with the PFN (220 ml v 287 ml, p = 0.001). Post-operatively, more lateral protrusion of the hip screws of the PFN (7.6%) was documented, compared with the gamma nail (1.6%, p = 0.02). Most local complications were related to suboptimal reduction of the fracture and/or positioning of the implant. Functional outcome and consolidation were equal for both implants. Generally, the results of treatment of unstable trochanteric fractures were comparable for the PFN and GN. The pitfalls and complications were similar, and mainly surgeon- or fracture-related, rather than implant-related.
                Bookmark

                Author and article information

                Contributors
                Role:
                Role: Role:
                Role: Role:
                URI : https://loop.frontiersin.org/people/1871316/overviewRole:
                URI : https://loop.frontiersin.org/people/1831072/overviewRole:
                Role:
                URI : https://loop.frontiersin.org/people/1923835/overviewRole:
                URI : https://loop.frontiersin.org/people/2598155/overviewRole:
                Journal
                Front Bioeng Biotechnol
                Front Bioeng Biotechnol
                Front. Bioeng. Biotechnol.
                Frontiers in Bioengineering and Biotechnology
                Frontiers Media S.A.
                2296-4185
                13 June 2024
                2024
                : 12
                : 1393154
                Affiliations
                [1] 1 Trauma Emergency Center , Third Hospital of Hebei Medical University , Shijiazhuang, China
                [2] 2 Orthopaedic Research Institute of Hebei Province , Shijiazhuang, China
                [3] 3 Key Laboratory of Biomechanics of Hebei Province , Shijiazhuang, China
                [4] 4 NHC Key Laboratory of Intelligent Orthopaedic Equipment , Shijiazhuang, China
                [5] 5 Hebei Orthopaedic Clinical Research Center , Shijiazhuang, China
                Author notes

                Edited by: Ridha Hambli, Polytech Orléans, France

                Reviewed by: Wenxin Niu, Tongji University, China

                Zikai Hua, Orthotek Laboratory, China

                *Correspondence: Qi Zhang, drzhangqi01@ 123456gmail.com ; Yingze Zhang, yzzhangdr@ 123456163.com
                [ † ]

                These authors have contributed equally to this work

                Article
                1393154
                10.3389/fbioe.2024.1393154
                11208680
                38938983
                2c89d28b-555f-4cf4-ac63-9fba92cb4ceb
                Copyright © 2024 Yang, Tong, Cheng, Zhu, Chen, Cui, Zhang and Zhang.

                This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

                History
                : 28 February 2024
                : 27 May 2024
                Funding
                The author(s) declare that financial support was received for the research, authorship, and/or publication of this article. This study was supported by “National Natural Science Foundation of China” (Grant No. 82272578) and “Health Commission of Hebei Province” (Grant No. 20242277).
                Categories
                Bioengineering and Biotechnology
                Original Research
                Custom metadata
                Biomechanics

                biomechanics,cephalomedullary nail,finite element analysis,proximal femoral bionic nail,reverse obliquity intertrochanteric fracture

                Comments

                Comment on this article