      Assessment of aortic diameter in Marfan patients: intraindividual comparison of 3D-Dixon and 2D-SSFP magnetic resonance imaging

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Objectives

          To compare the accuracy and precision of 3D-Dixon and 2D-SSFP MR-imaging for assessment of aortic diameter in Marfan patients.

          Methods

          This prospective single-center study investigated respiratory-gated 3D-Dixon and breath-hold 2D-SSFP non-contrast MR-imaging at 3 T in 47 Marfan patients (36.0 ± 13.2 years, 28♀,19♂). Two radiologists performed individual diameter measurements at five levels of the thoracic aorta and evaluated image quality on a four-grade scale (1 = poor, 4 = excellent) and artifacts (1 = severe, 4 = none). Aortic root diameters acquired by echocardiography served as a reference standard. Intraclass correlation coefficient, Bland-Altman analyses, F-test, t-test, and regression analyses were used to assess agreement between observers and methods.

          Results

          Greatest aortic diameters were observed at the level of the sinuses of Valsalva (SOV) for 3D-Dixon (38.2 ± 6.8 mm) and 2D-SSFP (38.3 ± 7.1 mm) ( p = 0.53). Intra- and interobserver correlation of diameter measurements was excellent at all aortic levels for both 3D-Dixon ( r = 0.94–0.99 and r = 0.94–0.98) and 2D-SSFP ( r = 0.96–1.00 and r = 0.95–0.99). 3D-Dixon-derived and 2D-SSFP-derived diameter measurements at the level of the SOV revealed a strong correlation with echocardiographic measurements ( r = 0.92, p < 0.001 and r = 0.93, p < 0.001, respectively). The estimated mean image quality at the level of SOV was higher for 2D-SSFP compared to that for 3D-Dixon (3.3 (95%-CI: 3.1–3.5) vs. 2.9 (95%-CI: 2.7–3.1)) ( p < 0.001). Imaging artifacts were less at all aortic levels for 3D-Dixon compared to 2D-SSFP (3.4–3.8 vs. 2.8–3.1) (all p < 0.002).

          Conclusion

          Respiratory-gated 3D-Dixon and breath-hold 2D-SSFP MR-imaging provide accurate and precise aortic diameter measurements. We recommend 3D-Dixon imaging for monitoring of aortic diameter in Marfan patients due to fewer imaging artifacts and the possibility of orthogonal multiplanar reformations of the aortic root.

          Key Points

          • Respiratory-gated 3D-Dixon and breath-hold 2D-SSFP imaging provide accurate and precise aortic diameter measurements in patients suffering from Marfan syndrome.

          • Imaging artifacts are stronger in 2D-SFFP imaging than in 3D-Dixon imaging.

          • We recommend 3D-Dixon imaging for monitoring of aortic diameter in Marfan patients due to fewer imaging artifacts and the possibility of orthogonal multiplanar reformations.

          Related collections

          Most cited references28

          • Record: found
          • Abstract: found
          • Article: not found

          The ImageJ ecosystem: An open platform for biomedical image analysis.

          Technology in microscopy advances rapidly, enabling increasingly affordable, faster, and more precise quantitative biomedical imaging, which necessitates correspondingly more-advanced image processing and analysis techniques. A wide range of software is available-from commercial to academic, special-purpose to Swiss army knife, small to large-but a key characteristic of software that is suitable for scientific inquiry is its accessibility. Open-source software is ideal for scientific endeavors because it can be freely inspected, modified, and redistributed; in particular, the open-software platform ImageJ has had a huge impact on the life sciences, and continues to do so. From its inception, ImageJ has grown significantly due largely to being freely available and its vibrant and helpful user community. Scientists as diverse as interested hobbyists, technical assistants, students, scientific staff, and advanced biology researchers use ImageJ on a daily basis, and exchange knowledge via its dedicated mailing list. Uses of ImageJ range from data visualization and teaching to advanced image processing and statistical analysis. The software's extensibility continues to attract biologists at all career stages as well as computer scientists who wish to effectively implement specific image-processing algorithms. In this review, we use the ImageJ project as a case study of how open-source software fosters its suites of software tools, making multitudes of image-analysis technology easily accessible to the scientific community. We specifically explore what makes ImageJ so popular, how it impacts the life sciences, how it inspires other projects, and how it is self-influenced by coevolving projects within the ImageJ ecosystem.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            The revised Ghent nosology for the Marfan syndrome.

            The diagnosis of Marfan syndrome (MFS) relies on defined clinical criteria (Ghent nosology), outlined by international expert opinion to facilitate accurate recognition of this genetic aneurysm syndrome and to improve patient management and counselling. These Ghent criteria, comprising a set of major and minor manifestations in different body systems, have proven to work well since with improving molecular techniques, confirmation of the diagnosis is possible in over 95% of patients. However, concerns with the current nosology are that some of the diagnostic criteria have not been sufficiently validated, are not applicable in children or necessitate expensive and specialised investigations. The recognition of variable clinical expression and the recently extended differential diagnosis further confound accurate diagnostic decision making. Moreover, the diagnosis of MFS--whether or not established correctly--can be stigmatising, hamper career aspirations, restrict life insurance opportunities, and cause psychosocial burden. An international expert panel has established a revised Ghent nosology, which puts more weight on the cardiovascular manifestations and in which aortic root aneurysm and ectopia lentis are the cardinal clinical features. In the absence of any family history, the presence of these two manifestations is sufficient for the unequivocal diagnosis of MFS. In absence of either of these two, the presence of a bonafide FBN1 mutation or a combination of systemic manifestations is required. For the latter a new scoring system has been designed. In this revised nosology, FBN1 testing, although not mandatory, has greater weight in the diagnostic assessment. Special considerations are given to the diagnosis of MFS in children and alternative diagnoses in adults. We anticipate that these new guidelines may delay a definitive diagnosis of MFS but will decrease the risk of premature or misdiagnosis and facilitate worldwide discussion of risk and follow-up/management guidelines.
              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              Multimodality imaging of diseases of the thoracic aorta in adults: from the American Society of Echocardiography and the European Association of Cardiovascular Imaging: endorsed by the Society of Cardiovascular Computed Tomography and Society for Cardiovascular Magnetic Resonance.

                Bookmark

                Author and article information

                Contributors
                f.wright@uke.de
                Journal
                Eur Radiol
                Eur Radiol
                European Radiology
                Springer Berlin Heidelberg (Berlin/Heidelberg )
                0938-7994
                1432-1084
                21 October 2022
                21 October 2022
                2023
                : 33
                : 3
                : 1687-1697
                Affiliations
                [1 ]GRID grid.13648.38, ISNI 0000 0001 2180 3484, Department of Diagnostic and Interventional Radiology and Nuclear Medicine, , University Medical Center Hamburg-Eppendorf, ; Martinistraße 52, 20246 Hamburg, Germany
                [2 ]GRID grid.418621.8, ISNI 0000 0004 0373 4886, Philips GmbH Market DACH, ; Hamburg, Germany
                [3 ]GRID grid.13648.38, ISNI 0000 0001 2180 3484, Department of Vascular Medicine, , University Heart and Vascular Center Hamburg, ; Hamburg, Germany
                [4 ]GRID grid.13648.38, ISNI 0000 0001 2180 3484, Institute of Medical Biometry and Epidemiology, , University Medical Center Hamburg-Eppendorf, ; Hamburg, Germany
                Author information
                http://orcid.org/0000-0002-0687-8063
                Article
                9162
                10.1007/s00330-022-09162-y
                9935710
                36269370
                27f0d678-dbb6-494e-8ec1-c88673c545be
                © The Author(s) 2022

                Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

                History
                : 18 May 2022
                : 16 September 2022
                : 20 September 2022
                Funding
                Funded by: Universitätsklinikum Hamburg-Eppendorf (UKE) (5411)
                Categories
                Magnetic Resonance
                Custom metadata
                © European Society of Radiology 2023

                Radiology & Imaging
                marfan syndrome,magnetic resonance angiography,artifacts,echocardiography,thoracic aorta

                Comments

                Comment on this article