30
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Infant Cardiac CT Angiography with 64-Slice and 256-Slice CT: Comparison of Radiation Dose and Image Quality Using a Pediatric Phantom

      research-article
      1 , 2 , 3 , 4 , 1 , *
      PLoS ONE
      Public Library of Science

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          The aims of this study were to investigate the image quality and radiation exposure of pediatric protocols for cardiac CT angiography (CTA) in infants under one year of age.

          Methodology/Principal Findings

          Cardiac CTA examinations were performed using an anthropomorphic phantom representing a 1-year-old child scanned with non-electrocardiogram-gated (NG), retrospectively electrocardiogram-gated helical (RGH) and prospectively electrocardiogram-gated axial (PGA) techniques in 64-slice and 256-slice CT scanners. The thermoluminescent dosimeters (TLD) were used for direct organ dose measurement, while dose-length product and effective mAs were also used to estimate the patient dose. For image quality, noise and signal-to-noise-ratio (SNR) were assessed based on regions-of-interest drawn on the reconstructed CT images, and were compared with the proposed cardiac image quantum index (CIQI). Estimated dose results were in accordant to the measured doses. The NG scan showed the best image quality in terms of noise and SNR. The PGA scan had better image quality than the RGH scan with 83.70% dose reduction. Noise and SNR were also corresponded to the proposed CIQI.

          Conclusions/Significance

          The PGA scan protocol was a good choice in balancing radiation exposure and image quality for infant cardiac CTA. We also suggested that the effective mAs and the CIQI were suitable in assessing the tradeoffs between radiation dose and image quality for cardiac CTA in infants. These results are useful for future implementation of dose reduction strategies in pediatric cardiac CTA protocols.

          Related collections

          Most cited references27

          • Record: found
          • Abstract: found
          • Article: not found

          Estimated risks of radiation-induced fatal cancer from pediatric CT.

          In light of the rapidly increasing frequency of pediatric CT examinations, the purpose of our study was to assess the lifetime cancer mortality risks attributable to radiation from pediatric CT. Organ doses as a function of age-at-diagnosis were estimated for common CT examinations, and estimated attributable lifetime cancer mortality risks (per unit dose) for different organ sites were applied. Standard models that assume a linear extrapolation of risks from intermediate to low doses were applied. On the basis of current standard practice, the same exposures (milliampere-seconds) were assumed, independent of age. The larger doses and increased lifetime radiation risks in children produce a sharp increase, relative to adults, in estimated risk from CT. Estimated lifetime cancer mortality risks attributable to the radiation exposure from a CT in a 1-year-old are 0.18% (abdominal) and 0.07% (head)-an order of magnitude higher than for adults-although those figures still represent a small increase in cancer mortality over the natrual background rate. In the United States, of approximately 600,000 abdominal and head CT examinations annually performed in children under the age of 15 years, a rough estimate is that 500 of these individuals might ultimately die from cancer attributable to the CT radiation. The best available risk estimates suggest that pediatric CT will result in significantly increased lifetime radiation risk over adult CT, both because of the increased dose per milliampere-second, and the increased lifetime risk per unit dose. Lower milliampere-second settings can be used for children without significant loss of information. Although the risk-benefit balance is still strongly tilted toward benefit, because the frequency of pediatric CT examinations is rapidly increasing, estimates that quantitative lifetime radiation risks for children undergoing CT are not negligible may stimulate more active reduction of CT exposure settings in pediatric patients.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            ACCF/SCCT/ACR/AHA/ASE/ASNC/NASCI/SCAI/SCMR 2010 Appropriate Use Criteria for Cardiac Computed Tomography. A Report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, the Society of Cardiovascular Computed Tomography, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the American Society of Nuclear Cardiology, the North American Society for Cardiovascular Imaging, the Society for Cardiovascular Angiography and Interventions, and the Society for Cardiovascular Magnetic Resonance.

            The American College of Cardiology Foundation, along with key specialty and subspecialty societies, conducted an appropriate use review of common clinical scenarios where cardiac computed tomography (CCT) is frequently considered. The present document is an update to the original CCT/cardiac magnetic resonance appropriateness criteria published in 2006, written to reflect changes in test utilization, to incorporate new clinical data, and to clarify CCT use where omissions or lack of clarity existed in the original criteria. The indications for this review were drawn from common applications or anticipated uses, as well as from current clinical practice guidelines. Ninety-three clinical scenarios were developed by a writing group and scored by a separate technical panel on a scale of 1 to 9 to designate appropriate use, inappropriate use, or uncertain use. In general, use of CCT angiography for diagnosis and risk assessment in patients with low or intermediate risk or pretest probability for coronary artery disease was viewed favorably, whereas testing in high-risk patients, routine repeat testing, and general screening in certain clinical scenarios were viewed less favorably. Use of noncontrast computed tomography for calcium scoring was rated as appropriate within intermediate- and selected low-risk patients. Appropriate applications of CCT are also within the category of cardiac structural and functional evaluation. It is anticipated that these results will have an impact on physician decision making, performance, and reimbursement policy, and that they will help guide future research.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Radiation-related cancer risks at low doses among atomic bomb survivors.

              To clarify the information in the Radiation Effects Research Foundation data regarding cancer risks of low radiation doses, we focus on survivors with doses less than 0.5 Sv. For reasons indicated, we also restrict attention mainly to survivors within 3, 000 m of the hypocenter of the bombs. Analysis is of solid cancer incidence from 1958-1994, involving 7,000 cancer cases among 50,000 survivors in that dose and distance range. The results provide useful risk estimates for doses as low as 0.05-0.1 Sv, which are not overestimated by linear risk estimates computed from the wider dose ranges 0-2 Sv or 0-4 Sv. There is a statistically significant risk in the range 0-0.1 Sv, and an upper confidence limit on any possible threshold is computed as 0.06 Sv. It is indicated that modification of the neutron dose estimates currently under consideration would not markedly change the conclusions.
                Bookmark

                Author and article information

                Contributors
                Role: Editor
                Journal
                PLoS One
                PLoS ONE
                plos
                plosone
                PLoS ONE
                Public Library of Science (San Francisco, USA )
                1932-6203
                2012
                21 November 2012
                : 7
                : 11
                : e49609
                Affiliations
                [1 ]Department of Biomedical Imaging and Radiological Sciences, National Yang Ming University, Taipei, Taiwan
                [2 ]Department of Radiology, Kaohsiung Chang Gung Memorial Hospital and Chang Gung University College of Medicine, Kaohsiung, Taiwan
                [3 ]Department of Radiological Technology, Tzu Chi college of Technology, Hualien, Taiwan
                [4 ]Department of Electrical and Electronics Engineering, Faculty of Science and Technology, University of Macau, Macau, China
                The University of Tennessee Health Science Center, United States of America
                Author notes

                Competing Interests: The authors have declared that no competing interests exist.

                Conceived and designed the experiments: YL TW. Performed the experiments: YL CY. Analyzed the data: YL CY TW. Contributed reagents/materials/analysis tools: CY GM TW. Wrote the paper: YL CY GM TW.

                Article
                PONE-D-12-26196
                10.1371/journal.pone.0049609
                3504147
                23185380
                200f3417-e2a3-441b-a91b-86c27f3331b9
                Copyright @ 2012

                This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

                History
                : 29 August 2012
                : 11 October 2012
                Page count
                Pages: 9
                Funding
                The study was financially supported by National Science Council of Taiwan (NSC99-2314-B-010 -043-MY3). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
                Categories
                Research Article
                Biology
                Anatomy and Physiology
                Cardiovascular System
                Radiobiology
                Computer Science
                Computerized Simulations
                Computing Systems
                Quantum Computing
                Medicine
                Anatomy and Physiology
                Cardiovascular System
                Cardiovascular
                Cardiovascular Imaging
                Pediatric Cardiology
                Pediatrics
                Pediatric Cardiology
                Radiology
                Diagnostic Radiology
                Computed Tomography
                Medical Physics

                Uncategorized
                Uncategorized

                Comments

                Comment on this article