48
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: not found

      Partial volume correction strategies for quantitative FDG PET in oncology

      research-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose

          Quantitative accuracy of positron emission tomography (PET) is affected by partial volume effects resulting in increased underestimation of the standardized uptake value (SUV) with decreasing tumour volume. The purpose of the present study was to assess accuracy and precision of different partial volume correction (PVC) methods.

          Methods

          Three methods for PVC were evaluated: (1) inclusion of the point spread function (PSF) within the reconstruction, (2) iterative deconvolution of PET images and (3) calculation of spill-in and spill-out factors based on tumour masks. Simulations were based on a mathematical phantom with tumours of different sizes and shapes. Phantom experiments were performed in 2-D mode using the National Electrical Manufacturers Association (NEMA) NU2 image quality phantom containing six differently sized spheres. Clinical studies (2-D mode) included a test-retest study consisting of 10 patients with stage IIIB and IV non-small cell lung cancer and a response monitoring study consisting of 15 female breast cancer patients. In all studies tumour or sphere volumes of interest (VOI) were generated using VOI based on adaptive relative thresholds.

          Results

          Simulations and experiments provided similar results. All methods were able to accurately recover true SUV within 10% for spheres equal to and larger than 1 ml. Reconstruction-based recovery, however, provided up to twofold better precision than image-based methods. Clinical studies showed that PVC increased SUV by 5–80% depending on tumour size. Test-retest variability slightly worsened from 9.8 ± 6.5 without to 10.8 ± 7.9% with PVC. Finally, PVC resulted in slightly smaller SUV responses, i.e. from −30.5% without to −26.3% with PVC after the first cycle of treatment ( p < 0.01).

          Conclusion

          PVC improves accuracy of SUV without decreasing (clinical) test-retest variability significantly and it has a small, but significant effect on observed tumour responses. Reconstruction-based PVC outperforms image-based methods, but requires dedicated reconstruction software. Image-based methods are good alternatives because of their ease of implementation and their similar performance in clinical studies.

          Related collections

          Most cited references33

          • Record: found
          • Abstract: found
          • Article: not found

          Correction for partial volume effects in PET: principle and validation.

          The accuracy of PET for measuring regional radiotracer concentrations in the human brain is limited by the finite resolution capability of the scanner and the resulting partial volume effects (PVEs). We designed a new algorithm to correct for PVEs by characterizing the geometric interaction between the PET system and the brain activity distribution. The partial volume correction (PVC) algorithm uses high-resolution volumetric MR images correlated with the PET volume. We used a PET simulator to calculate recovery and cross-contamination factors of identified tissue components in the brain model. These geometry-dependent transfer coefficients form a matrix representing the fraction of true activity from each distinct brain region observed in any given set of regions of interest. This matrix can be inverted to correct for PVEs, independent of the tracer concentrations in each tissue component. A sphere phantom was used to validate the simulated point-spread function of the PET scanner. Accuracy and precision of the PVC method were assessed using a human basal ganglia phantom. A constant contrast experiment was performed to explore the recovery capability and statistic error propagation of PVC in various noise conditions. In addition, a dual-isotope experiment was used to evaluate the ability of the PVC algorithm to recover activity concentrations in small structures surrounded by background activity with a different radioactive half-life. This models the time-variable contrast between regions that is often seen in neuroreceptor studies. Data from the three-dimensional brain phantom demonstrated a full recovery capability of PVC with less than 10% root mean-square error in terms of absolute values, which decreased to less than 2% when results from four PET slices were averaged. Inaccuracy in the estimation of 18F tracer half-life in the presence of 11C background activity was in the range of 25%-50% before PVC and 0%-6% after PVC, for resolution varying from 6 to 14 mm FWHM. In terms of noise propagation, the degradation of the coefficient of variation after PVC was found to be easily predictable and typically on the order of 25%. The PVC algorithm allows the correction for PVEs simultaneously in all identified brain regions, independent of tracer levels.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study.

            Semiquantitative standard uptake values (SUVs) are used for tumor diagnosis and response monitoring. However, the accuracy of the SUV and the accuracy of relative change during treatment are not well documented. Therefore, an experimental and simulation study was performed to determine the effects of noise, image resolution, and region-of-interest (ROI) definition on the accuracy of SUVs. Experiments and simulations are based on thorax phantoms with tumors of 10-, 15-, 20-, and 30-mm diameter and background ratios (TBRs) of 2, 4, and 8. For the simulation study, sinograms were generated by forward projection of the phantoms. For each phantom, 50 sinograms were generated at 3 noise levels. All sinograms were reconstructed using ordered-subset expectation maximization (OSEM) with 2 iterations and 16 subsets, with or without a 6-mm gaussian filter. For each tumor, the maximum pixel value and the average of a 50%, a 70%, and an adaptive isocontour threshold ROI were derived as well as with an ROI of 15 x 15 mm. The accuracy of SUVs was assessed using the average of 50 ROI values. Treatment response was simulated by varying the tumor size or the TBR. For all situations, a strong correlation was found between maximum and isocontour-based ROI values resulting in similar dependencies on image resolution and noise of all studied SUV measures. A strong variation with tumor size of > or =50% was found for all SUV values. For nonsmoothed data with high noise levels this variation was primarily due to noise, whereas for smoothed data with low noise levels partial-volume effects were most important. In general, SUVs showed under- and overestimations of > or =50% and depended on all parameters studied. However, SUV ratios, used for response monitoring, were only slightly dependent of ROI definition but were still affected by noise and resolution. The poor accuracy of the SUV under various conditions may hamper its use for diagnosis, especially in multicenter trials. SUV ratios used to measure response to treatment, however, are less dependent on noise, image resolution, and ROI definition. Therefore, the SUV might be more suitable for response-monitoring purposes.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              A fuzzy locally adaptive Bayesian segmentation approach for volume determination in PET.

              Accurate volume estimation in positron emission tomography (PET) is crucial for different oncology applications. The objective of our study was to develop a new fuzzy locally adaptive Bayesian (FLAB) segmentation for automatic lesion volume delineation. FLAB was compared with a threshold approach as well as the previously proposed fuzzy hidden Markov chains (FHMC) and the fuzzy C-Means (FCM) algorithms. The performance of the algorithms was assessed on acquired datasets of the IEC phantom, covering a range of spherical lesion sizes (10-37 mm), contrast ratios (4:1 and 8:1), noise levels (1, 2, and 5 min acquisitions), and voxel sizes (8 and 64 mm(3)). In addition, the performance of the FLAB model was assessed on realistic nonuniform and nonspherical volumes simulated from patient lesions. Results show that FLAB performs better than the other methodologies, particularly for smaller objects. The volume error was 5%-15% for the different sphere sizes (down to 13 mm), contrast and image qualities considered, with a high reproducibility (variation < 4%). By comparison, the thresholding results were greatly dependent on image contrast and noise, whereas FCM results were less dependent on noise but consistently failed to segment lesions < 2 cm. In addition, FLAB performed consistently better for lesions < 2 cm in comparison to the FHMC algorithm. Finally the FLAB model provided errors less than 10% for nonspherical lesions with inhomogeneous activity distributions. Future developments will concentrate on an extension of FLAB in order to allow the segmentation of separate activity distribution regions within the same functional volume as well as a robustness study with respect to different scanners and reconstruction algorithms.
                Bookmark

                Author and article information

                Contributors
                +31-20-4445240 , +31-20-4443090 , n.hoetjes@vumc.nl
                Journal
                Eur J Nucl Med Mol Imaging
                European Journal of Nuclear Medicine and Molecular Imaging
                Springer-Verlag (Berlin/Heidelberg )
                1619-7070
                1619-7089
                27 April 2010
                27 April 2010
                August 2010
                : 37
                : 9
                : 1679-1687
                Affiliations
                [1 ]Department of Nuclear Medicine & PET Research, VU University Medical Center, P.O. Box 7057, 1007 MB Amsterdam, The Netherlands
                [2 ]Department of Nuclear Medicine, Jeroen Bosch Hospital, Den Bosch, The Netherlands
                [3 ]Department of Radiology, Erasmus Medical Center, Rotterdam, The Netherlands
                Article
                1472
                10.1007/s00259-010-1472-7
                2918791
                20422184
                1c936eab-cd39-4f98-919a-7005b6cbd7cd
                © The Author(s) 2010
                History
                : 25 January 2010
                : 6 April 2010
                Categories
                Original Article
                Custom metadata
                © Springer-Verlag 2010

                Radiology & Imaging
                oncology,partial volume correction,fdg,standardized uptake value (suv),positron emission tomography (pet)

                Comments

                Comment on this article