4
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          To determine the repeatability and agreement using corneal tomography of a swept-source optical coherence tomographer (SS-OCT) compared to a rotating Scheimpflug camera (RSC) in normal eyes and keratoconus (KC) eyes.

          Methods

          This prospective repeatability analysis was performed at the Department of Ophthalmology of University Hospital Carl Gustav Carus, Dresden, Germany. Forty-three normal and 57 KC eyes were enrolled in the study. Three consecutive measurements were performed by the same operator on each device. Corneal parameters of anterior and posterior corneal surface, such as simulated keratometry (SimK), as well as central and thinnest corneal thickness were evaluated. Repeatability and agreement were assessed by using the coefficient of repeatability and Bland-Altman analysis.

          Results

          The repeatability of anterior corneal parameters was comparable between RSC and SS-OCT in normal eyes (repeatability < 0.5 D). Repeatability was increased in mild and moderate KC for all parameters using both devices. In moderate KC, repeatability of Kmax was 1.33 D and 0.78 D for RSC and SS-OCT, respectively. Repeatability of posterior corneal parameters was consistently better for SS-OCT. Significant offsets and wide ranges of limits of agreement were found between the devices for SimK and corneal thickness values.

          Conclusions

          SS-OCT showed highly repeatable measurements of anterior and posterior corneal parameters in normal and KC eyes. Compared to RSC, the SS-OCT had a better repeatability of anterior corneal parameters in mild and moderate KC as well as posterior corneal parameters in all groups. Both devices should not be used interchangeably in the diagnostic process of patients.

          Trial registration NCT04251143 at Clinicaltrials.gov, registered on 12 March 2018, https://clinicaltrials.gov/ct2/show/NCT04251143?cond=Keratoconus&cntry=DE&city=Dresden&draw=2&rank=1

          Supplementary Information

          The online version contains supplementary material available at 10.1186/s40662-022-00290-6.

          Related collections

          Most cited references43

          • Record: found
          • Abstract: not found
          • Article: not found

          Keratoconus

            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Global consensus on keratoconus and ectatic diseases.

            Despite extensive knowledge regarding the diagnosis and management of keratoconus and ectatic corneal diseases, many controversies still exist. For that reason, there is a need for current guidelines for the diagnosis and management of these conditions. This project aimed to reach consensus of ophthalmology experts from around the world regarding keratoconus and ectatic diseases, focusing on their definition, concepts, clinical management, and surgical treatments. The Delphi method was followed with 3 questionnaire rounds and was complemented with a face-to-face meeting. Thirty-six panelists were involved and allocated to 1 of 3 panels: definition/diagnosis, nonsurgical management, or surgical treatment. The level of agreement considered for consensus was two thirds. Numerous agreements were generated in definitions, methods of diagnosing, and management of keratoconus and other ectatic diseases. Nonsurgical and surgical treatments for these conditions, including the use of corneal cross-linking and corneal transplantations, were presented in a stepwise approach. A flowchart describing a logical management sequence for keratoconus was created. This project resulted in definitions, statements, and recommendations for the diagnosis and management of keratoconus and other ectatic diseases. It also provides an insight into the current worldwide treatment of these conditions.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Riboflavin/ultraviolet-a-induced collagen crosslinking for the treatment of keratoconus.

              In animal eyes, a significant increase in corneal biomechanical stiffness has been found after collagen crosslinking by combined riboflavin/ultraviolet-A (UVA) treatment. The aim of the present study was to evaluate the clinical usefulness of riboflavin/UVA-induced collagen crosslinking for bringing the progression of keratoconus to a halt. Prospective, nonrandomized clinical pilot study. Twenty-three eyes of 22 patients with moderate or advanced progressive keratoconus (maximum K value, 48-72 diopters) were included. After central corneal abrasion, photosensitizing riboflavin drops were applied and the eyes exposed to UVA (370 nm, 3 mW/cm(2)) in a 1-cm distance for 30 minutes. Postoperative examinations were performed in 6-month intervals, including visual acuity testing, corneal topography, slit-lamp examination, measurement of endothelial cell density, and photographic documentation. The follow-up time was between 3 months and 4 years. In all treated eyes, the progression of keratoconus was at least stopped. In 16 eyes (70%) regression with a reduction of the maximal keratometry readings by 2.01 diopters and of the refractive error by 1.14 diopters was found. Corneal and lens transparency, endothelial cell density, and intraocular pressure remained unchanged. Visual acuity improved slightly in 15 eyes (65%). Collagen crosslinking may be a new way for stopping the progression of keratectasia in patients with keratoconus. The need for penetrating keratoplasty might then be significantly reduced in keratoconus. Given the simplicity and minimal costs of the treatment, it might also be well-suited for developing countries. Long-term results are necessary to evaluate the duration of the stiffening effect and to exclude long term side-effects.
                Bookmark

                Author and article information

                Contributors
                robert.herber@uniklinikum-dresden.de
                Journal
                Eye Vis (Lond)
                Eye Vis (Lond)
                Eye and Vision
                BioMed Central (London )
                2326-0254
                23 May 2022
                23 May 2022
                2022
                : 9
                : 19
                Affiliations
                GRID grid.412282.f, ISNI 0000 0001 1091 2917, Department of Ophthalmology, , University Hospital Carl Gustav Carus, TU Dresden, ; Fetscherstraße 74, 01307 Dresden, Germany
                Author information
                http://orcid.org/0000-0003-0829-7577
                Article
                290
                10.1186/s40662-022-00290-6
                9128256
                35606839
                10b763f4-d535-4768-9feb-e2bf2f0006a4
                © The Author(s) 2022

                Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

                History
                : 30 November 2021
                : 25 April 2022
                Categories
                Research
                Custom metadata
                © The Author(s) 2022

                anterion,cornea,keratoconus,oct,pentacam,scheimpflug,swept-source,topography,tomography

                Comments

                Comment on this article